Wpływ zakłóceń elektromagnetycznych na funkcje wszczepialnych urządzeń kontrolujących rytm serca
|
|
- Sabina Nowak
- 9 lat temu
- Przeglądów:
Transkrypt
1 PRACA POGLĄDOWA Folia Cardiol. 2005, tom 12, nr 4, Copyright 2005 Via Medica ISSN Wpływ zakłóceń elektromagnetycznych na funkcje wszczepialnych urządzeń kontrolujących rytm serca Effect of electromagnetic interferences on implantable cardiac devices Adam Sokal, Oskar Kowalski, Radosław Lenarczyk, Patrycja Pruszkowska-Skrzep, Janusz Prokopczuk i Zbigniew Kalarus I Katedra i Oddział Kliniczny Kardiologii Śląskiej Akademii Medycznej, Śląskie Centrum Chorób Serca w Zabrzu Abstract Simultaneously with the development of medical technology, the number of implanted rhythm controlling devices gradually rises. Additionally, increasing complication level of implanted devices together with growing number of potential sources of electromagnetic fields creates problem of potential interferences every time more important. Among different reactions of implanted device on interfering electromagnetic signal the most common are: inhibition of single stimulus or generation of additional stimulus, complete inhibition of stimulation, switch to emergency stimulation and increase or change of stimulation frequency. Possible sources of interferences can be divided into in- and out-hospital and occupational. First group comprise magnetic nuclear resonance (NMR), diathermy devices and radiotherapy. In out-hospital environment the most frequent interferences are mobile phones and electronic surveillance devices. This review present level of knowledge about possible interferences and their clinical consequences. (Folia Cardiol. 2005; 12: ) implantable devices, pacemakers, implantable cardioverters defibrillators, electromagnetic interferences Wstęp Większość wytwórców aktualnie dostępnych kardiostymulatorów i wszczepialnych kardiowerterów-defibrylatorów (ICD, implantable cardioverter defibrillator) w fazie projektowej dąży do zminimalizowania ryzyka związanego z wpływem zewnętrznych pól elektromagnetycznych na czynność im- Adres do korespondencji: Dr biol., lek. med. Adam Sokal I Katedra i Oddział Kliniczny Kardiologii Śl. AM Śląskie Centrum Chorób Serca ul. Szpitalna 2, Zabrze tel. (0 32) , faks (0 32) karzab@infomed.slam.katowice.pl Nadesłano: r. Przyjęto do druku: r. plantowanych urządzeń. Jednakże sama ich natura nie pozwala na zupełne wyeliminowanie zakłóceń powstałych w wyniku oddziaływania zewnętrznych pól elektromagnetycznych, poprzez odpowiednie ekranowanie czy też wykorzystanie zabezpieczeń filtrujących. Urządzenia te muszą posiadać możliwość detekcji powstających wewnątrz serca impulsów elektromagnetycznych i modulacji własnej czynności zgodnie z aktualnym zapotrzebowaniem organizmu. Jednocześnie elektroda przewodząca sygnały z generatora i do niego stanowi idealną antenę odbierającą sygnały elektromagnetyczne z otoczenia. Obecnie stosowane generatory urządzeń wszczepialnych nie stwarzają ryzyka interakcji zarówno z polem elektrycznym, jak i magnetycznym w zakresie natężeń spotykanych w codziennym 261
2 Folia Cardiol. 2005, tom 12, nr 4 życiu (włączając typową ekspozycję zawodową) lub środowisku szpitalnym. Jedynie natężenia promieniowania elektromagnetycznego stosowanego w radioterapii mogą, poprzez interakcję z warstwami metalicznymi, uszkadzać układy elektroniczne generatora. Mało prawdopodobne jest również przeprogramowanie wszczepionego urządzenia poprzez przypadkowy impuls ze źródeł promieniowania elektromagnetycznego, gdyż na algorytmy wykonywania tego typu procedur, poza sygnałem modulującym, składają się sygnały autoryzacji i potwierdzenia, uniemożliwiające w praktyce przypadkowe przeprogramowanie. Typy odpowiedzi stymulatora na działanie zakłócających pól elektromagnetycznych Zewnętrzne pola elektromagnetyczne mogą jednak oddziaływać na układ stymulujący, wytwarzając impulsy o częstotliwości Hz, zbliżone do zjawisk fizjologicznych [1]. Efekt tego typu interferencji jest trudny do przewidzenia i zależy od charakteru sygnały (źródło i jego lokalizacja w stosunku do układu stymulującego), czasu trwania, typu i nastawień generatora oraz indywidualnych cech chorego. Ustawienie układu sterującego na tryb dwubiegunowy znacznie zmniejsza ryzyko ewentualnych zakłóceń. Reakcje układu stymulującego na działanie zewnętrznych pól elektromagnetycznych mogą przyjmować następującą postać [2]: wyhamowanie pojedynczego pobudzenia; zupełne wyhamowanie stymulacji (generator interpretuje sygnał zakłócający jako prawidłową czynność serca), prowadzące do znacznego zwolnienia lub zatrzymania czynności serca; przejście do trybu stymulacji awaryjnej (noise reversion mode), w którym generator przestaje być sterowany sygnałami zewnętrznymi i kontynuuje stymulację według nastawionych fabrycznie parametrów, najczęściej w trybie VOO; zwiększenie częstości rytmu; nieprawidłowa częstość rytmu. Wypadnięcie pojedynczego pobudzenia, poza uczuciem dyskomfortu i ewentualnego lęku ze strony chorego, nie wiąże się z istotnymi konsekwencjami klinicznymi. Potencjalnie znacznie poważniejsze jest trwałe zahamowanie stymulacji, szczególnie u chorych zależnych od stymulatora. W zależności od czasu trwania zahamowanie może być przyczyną zasłabnięcia, omdlenia lub nawet zgonu. W implantowanych współcześnie kardiostymulatorach istnieją specjalne algorytmy zapobiegające przedłużającej się inhibicji. Jednocześnie niewielu chorych ze wszczepionym rozrusznikiem jest całkowicie zależnych od stymulatora. Z tego powodu nie opisano dotąd żadnego zgonu pacjenta ze stymulatorem, którego przyczyną byłaby długotrwała interferencja stymulatora z zewnętrznym polem elektromagnetycznym, prowadząca do całkowitego zahamowania stymulacji. Chorzy z ICD, wymagający trwałej stymulacji z powodu bradykardii, są potencjalnie bardziej wrażliwi na tego typu zakłócenia. Automatyczne obniżanie progu czułości (nawet do 0,2 0,3 mv), którego celem jest detekcja epizodów niskonapięciowego migotania komór, może prowadzić do zwiększenia wpływu zewnątrzsercowych pól elektromagnetycznych na układ sterowania defibrylatora [3, 4]. Współczesne stymulatory są wyposażone w tryb stymulacji wyzwalanej zakłóceniami (noise reversion mode) [5]. W większości algorytmów tryb ten jest aktywowany w przypadku detekcji pobudzenia bezpośrednio po okresie bezwzględnej refrakcji (ventricular blanking, atrial blanking) i skutkuje włączeniem asynchronicznej stymulacji niekiedy poprzedzonej jedno- lub dwukrotnym wydłużeniem czasu refrakcji bezwzględnej [6]. Asynchroniczna stymulacja, mimo iż zabezpiecza chorego przed skutkami przedłużającego się zahamowania, może prowadzić do uczucia niemiarowej pracy serca. Potencjalnie istnieje też możliwość wywołania tachyarytmii komorowej, jeśli bodziec stymulatora trafi na okres nadwrażliwości tuż po szczycie załamka T. W badaniach Seeda i wsp. [7] wykazano, że w grupie osób ze zwiększonym ryzykiem tachyarytmii komorowych stymulacja asynchroniczna prowadziła do powstania monomorficznego częstoskurczu komorowego (VT, ventricular tachycardia) u 5% (13 na 268) badanych. W większości dostępnych współcześnie ICD włączenie trybu stymulacji wywołanej zakłóceniami ma charakter opcjonalny i może być aktywowane w trakcie programowania. Dodatkowo, ze względu na konieczność detekcji tachyarytmii, okres wrażliwy na zakłócenia jest stosunkowo krótki i wynosi 40 ms. Dlatego, jak już wspomniano, zahamowanie stymulacji przez sygnał zakłócający może wiązać się z potencjalnie poważniejszymi następstwami. Drugim typem odpowiedzi stymulatora na działanie zakłócających pól elektromagnetycznych jest wyzwalanie nieadekwatnych pobudzeń. U chorych ze stymulatorami sterowanymi z przedsionka (DDD, VDD) czułość elektrody przedsionkowej jest z reguły wyższa niż komorowej, dlatego też bodziec zewnętrzny może pobudzić kanał przedsionkowy, prowadząc w efekcie do przedwczesnej stymulacji komory. Podobnie jak w przypadku zahamowania, w większości sytuacji pojedyncze pobudzenie, poza dyskomfortem odczuwanym przez pacjenta, nie 262
3 A. Sokal i wsp., Zakłócenia pracy urządzeń wszczepialnych wiąże się z poważnymi konsekwencjami klinicznymi. Istnieje oczywiście potencjalna możliwość wyzwolenia w mechanizmie R na T tachyarytmii komorowej, ale wobec niezbędnego dla takiego efektu współistnienia niezależnych czynników, takich jak podatność na przetrwałe komorowe zaburzenia rytmu i odpowiednia synchronizacja impulsu zakłócającego z własnym rytmem, jest ona bardzo mało prawdopodobna. Znacznie poważniejsza z klinicznego punktu widzenia jest przetrwała, sterowana sygnałem zakłócającym tachyarytmia. Górny pułap częstości rytmu jest wprawdzie ograniczony przez nastawienia stymulatora, ale utrzymująca się asynchroniczna tachykardia może objawiać się jako kołatanie serca, a u podatnych pacjentów może prowadzić do objawów niestabilności hemodynamicznej (duszność, spadek ciśnienia) lub dławicy. Inną możliwością generowania tachykardii przez sygnał zakłócający może być interferencja z mechanizmem dostosowania częstości rytmu stymulatora do fizjologicznych zapotrzebowań organizmu w stymulatorach typu DDDR i VVIR. W rozrusznikach sterowanych długością odstępu QT zakłócenia w obrębie okresu detekcji QT mogą prowadzić do przyspieszenia czynności serca. Podobnie w stymulatorach wykorzystujących wielkość wentylacji minutowej do sterowania częstością rytmu zewnętrzny bodziec elektromagnetyczny może spowodować przyspieszenie częstości rytmu [9, 10]. Potencjalnie najbardziej niebezpieczna dla chorego jest szybka stymulacja przez nadprogowe sygnały generowane bezpośrednio w elektrodzie przez silne pole elektromagnetyczne o odpowiedniej częstotliwości. W badaniach przeprowadzonych na zwierzętach wykazano, że pole o częstotliwości 64 MHz i mocy wymaganej do wykonania skanu magnetycznego rezonansu jądrowego może wytwarzać szybką stymulację o częstości Hz, zdolną do indukcji migotania komór [11]. Tego typu stymulację o częstości ok. 100/min obserwowano u chorego z rozrusznikiem VVI poddanego badaniu magnetycznego rezonansu jądrowego [12]. Źródła sygnałów mogących zakłócać pracę wszczepialnych urządzeń kontrolujących rytm serca Z praktycznego punktu widzenia istotna jest znajomość źródeł impulsów elektromagnetycznych mogących zakłócać pracę kardiostymulatorów oraz ICD. Ważna jest również znajomość najczęściej spotykanych aberracji w działaniu urządzeń wszczepialnych, wywołanych przez konkretne źródło. Z reguły klasyfikuje się je na podstawie danych w środowisku stwarzającym ryzyko ekspozycji. Źródła impulsów elektromagnetycznych interferujących z urządzeniami wszczepialnymi można podzielić na: wewnątrzszpitalne; środowiskowe lub wewnątrzszpitalne; zawodowe. W środowisku wewnątrzszpitalnym możliwość zakłóceń pracy stymulatorów i ICD istnieje w trakcie badania MRI, diatermii i elektrokoagulacji, radioterapii, ablacji prądem RF, defibrylacji i kardiowersji, a także zewnątrzustrojowej litotrypsji (ESWL, extracorporeal shock wave lithotripsy) oraz przezskórnej elektrycznej stymulacji nerwów (TENS, transcutaneous electrical nerve stimulation). W przypadku ekspozycji środowiskowej praktyczne znaczenie mają jedynie telefony komórkowe i bramki zabezpieczające (antykradzieżowe) w sklepach, bibliotekach itp. miejscach (EAS, electronic article surveillance). Inne przyrządy elektryczne, takie jak: golarki [13], kosiarki elektryczne, bojlery [14], a nawet zabawki [15, 16], mogą być źródłem interferencji, jeżeli stosowane urządzenie jest wadliwe lub uszkodzone. Niekiedy uszkodzenie układu stymulującego przyczynia się do zwiększenia wrażliwości na działanie impulsów elektromagnetycznych wytwarzanych przez urządzenia elektryczne codziennego użytku. Opisano przypadek chorego, u którego występowały nieadekwatne wyładowania ICD przy stosowaniu elektrycznej golarki. Wykazano, że przyczyną tego zjawiska było uszkodzenie izolacji pierścienia komorowego [13]. W środowisku zawodowym potencjalne niebezpieczeństwo istnieje w sytuacji, gdy pacjent pracuje w bezpośredniej bliskości stacji transformatorowych, linii wysokiego napięcia [17], pieców indukcyjnych wysokonapięciowych urządzeń spawających lub odgazowujących [18]. Stosowanie typowych narzędzi, działających w zakresie napięć ogólnodostępnej sieci elektrycznej, w tym urządzeń spawających działających w zakresie natężeń A, nie stwarza niebezpieczeństwa interferencji ze stymulatorem lub ICD [18]. Bezpośredni wpływ pola wytwarzanego przez linie wysokiego napięcia na czynność stymulatorów i ICD jest w praktyce mało prawdopodobny. Natężenie pola generowanego w odległości 40 m od 400 kv linii wysokiego napięcia wynosi 1 kv/m. Aby zaistniała możliwość interferencji, powinno ono wynosić więcej niż 5 kv/m [18]. W incydentalnych przypadkach bezpośredniego kontaktu z nieizolowanym przewodem elektrycznym, w warunkach umożliwiających przepływ prądu przez ciało chorego (mokry grunt lub mokra podłoga, np. w łazience), może dojść do nieadekwatnej reakcji wszczepionego urządzenia [19]. 263
4 Folia Cardiol. 2005, tom 12, nr 4 Najczęstsze źródła interferencji elektromagnetycznych w środowisku szpitalnym Magnetyczny rezonans jądrowy Zgodnie z aktualnymi poglądami obecność wszczepionego kardiostymulatora lub ICD stanowi całkowite przeciwwskazanie do przeprowadzenia badania magnetycznego rezonansu jądrowego (NMR, nuclear magnetic resonance). Tego typu postępowanie podyktowane jest opisami zgonów, do których doszło po przeprowadzeniu badania u chorych z wszczepionym kardiostymulatorem. Na stronie internetowej FDA (Food and Drug Administration) znajdują się informacje o dwóch przypadkach zgonów osób z wszczepionym kardiostymulatorem, które nastąpiły po badaniu NMR. Na uwagę zasługuje fakt, że ostatni miał miejsce w grudniu 1992 r. Bezwzględne przeciwwskazanie do przeprowadzania badań NMR u pacjentów z implantowanym stymulatorem lub ICD pozbawia relatywnie dużą grupę chorych możliwości przeprowadzenia tego badania diagnostycznego. Zdaniem autorów japońskich u 17% z 1567 pacjentów z implantowanym stymulatorem serca istniały wskazania do wykonania badania NMR [20]. Ze względu na mniejszą dostępność omawianego badania w Polsce odsetek ten jest na pewno mniejszy, ale można się spodziewać, że z czasem będzie wzrastał. W trakcie badania NMR wszczepione urządzenie poddawane jest działaniu stałego pola magnetycznego, gradientowego pola magnetycznego i promieniowania elektromagnetycznego o częstotliwości radiowej. Elektrody w urządzeniach wszczepialnych są wykonane ze stopu niklu, kobaltu, chromu i molibdenu (MP35N), który nie będąc ferromagnetykiem, nie oddziałuje ze stałym polem magnetycznym. Zbudowana ze stali puszka stymulatora w stałym polu magnetycznym będzie podlegać sile magnetycznej równoległej do linii sił pola. Siła ta może spowodować przesunięcie puszki oraz wprowadzać ją w ruch drgający, co wywołuje u pacjenta uczucie dyskomfortu w loży stymulatora, ale nie wiąże się z poważnymi następstwami. W przypadku stymulatorów wartość siły pola magnetycznego wynosi 0,05 3,60 N i jest niższa od siły ciężkości. W przypadku ICD, będących urządzeniami o większych rozmiarach, uzyskane wartości były nieco wyższe i wynosiły 1,03 5,85 N [21]. Pole magnetyczne wytwarzane w tomografach NMR może również aktywować kontaktron (reed switch), który standardowo pozwala na wyzwolenie stymulacji asynchronicznej pod wpływem pola magnetycznego. Istnieje również potencjalna możliwość trwałego uszkodzenia kontaktronu prowadzącego do jego permanentnej aktywacji lub dezaktywacji. W badaniach Luechingera i wsp. [22] wykazano, że zachowanie kontaktronu zależy od orientacji i natężenia pola elektromagnetycznego. Aktywacja przełącznika kontraktonu występowała przy natężeniu równoległej składowej pola magnetycznego równym 1 ± 0,2 mt. Przełącznik pozostawał zamknięty przy natężeniach pola o wartości mniejszej niż 50 mt. Przy natężeniach przekraczających 200 mt w 50% badanych orientacji kontaktronu w polu magnetycznym dochodziło do jego otwarcia i wyłączenia trybu stymulacji asynchronicznej. Kontaktrony wbudowane do stymulatorów zachowywały się podobnie jak wyizolowane urządzenia. Zatem w polu magnetycznym wytwarzanym w czasie badania NMR, które wynosi 0,5; 1,0 i 1,5 T, nie można przewidzieć, czy stymulator pozostanie w trybie stymulacji asynchronicznej (przełącznik kontaktronu zamknięty), czy będzie odpowiadał na sygnały sterujące (przełącznik otwarty) [22]. Pulsujące z częstotliwością radiową zmienne gradientowe pole magnetyczne, oddziałując z elektrodą stymulatora, która w tej sytuacji przyjmuje funkcję anteny, może wytwarzać napięcie w elektrodzie, zdolne do bezpośredniej stymulacji lub interferujące z układem sterującym. Przypadek tego typu stymulacji u chorego z dwujamowym stymulatorem przełączonym w tryb OOO opisali Fontaine i wsp. [25]. Promieniowanie elektromagnetyczne o częstotliwości radiowej niezbędne do stworzenia obrazu NMR może, oddziałując z obwodami stymulatora lub ICD, prowadzić do szybkiej stymulacji. Sygnały zakłócające generowane przez promieniowanie elektromagnetyczne w kanale przedsionkowym, oddziałując z układem sterującym, będą wywoływać stymulację komór z maksymalną zaprogramowaną częstością. W przypadku ICD takie zakłócenie może być przez urządzenie błędnie interpretowane jako migotanie komór lub tachykardia, co może prowadzić do niewłaściwej reakcji. Bezpośrednia interferencja promieniowania RF z elementami elektronicznymi stymulatora może powodować niekontrolowaną stymulację o częstotliwości zsynchronizowanej z częstotliwością promieniowania elektromagnetycznego. W badaniach przeprowadzonych na psach obserwowano przypadki wywołanej w tym mechanizmie stymulacji z częstotliwością 300/min [23, 24]. Energia promieniowania radiowego o częstotliwości radiowej wytwarzana w tomografach NMR jest wystarczająca do podgrzania elementów układu stymulującego w stopniu mogącym prowadzić do destrukcji otaczającej tkanki. W badaniach in vitro, oceniających zdolność promieniowania RF wykorzystywanego w badaniach NMR do rozgrzewania elek- 264
5 A. Sokal i wsp., Zakłócenia pracy urządzeń wszczepialnych trod stymulatorów, obserwowano niezależny od typu elektrody wzrost temperatury o około 20 C [27, 36]. Istnieje więc ryzyko, że zwiększenie temperatury elektrody może być czynnikiem wystarczającym, by doprowadzić do martwicy przylegającej do niej tkanki, a w konsekwencji do istotnego wzrostu progu stymulacji. Wprawdzie w przeprowadzonych dotychczas badaniach nie dostrzeżono takiego efektu bezpośrednio po badaniu NMR, ale spowodowane podwyższeniem temperatury minimalne uszkodzenie tkanki może zaowocować aktywacją włóknienia wokół końcówki elektrody z powolnym zwiększeniem progu stymulacji prowadzącym w efekcie do zupełnego bloku wyjścia (exit block) [37]. Mimo ryzyka podejmuje się próby optymalizacji techniki i oceny bezpieczeństwa badania NMR u chorych z implantowanym kardiostymulatorem. W analizie przeprowadzonej przez Gimbela i wsp. [26], obejmującej 5 pacjentów poddanych badaniu NMR, tylko w jednym przypadku, u zależnego od stymulacji chorego ze stymulatorem pracującym w czasie badania w trybie DOO wystąpiła 2-sekundowa przerwa. Martin i wsp. [27] poddali badaniu NMR w 1,5 T tomografie 55 osób, nie stwierdzając u żadnej z nich istotnych lub trwałych konsekwencji w pracy układu stymulującego. Podobne wyniki uzyskała grupa Sommera i wsp. [28], która oceniała bezpieczeństwo badania 0,5 T tomografem NMR u 44 chorych z implantowanym kardiostymulatorem. U żadnego z nich nie wystąpiła istotna dysfunkcja kardiostymulatora ani nie odnotowano żadnych poważnych konsekwencji zdrowotnych. Wyniki cytowanych prac oraz opisy pojedynczych przypadków i doniesienia dotyczące małych grup badanych, u których bezpiecznie przeprowadzono badanie NMR [29 35], stwarzają nadzieję na weryfikację wcześniejszego poglądu, traktującego wszczepiony kardiostymulator jako całkowite przeciwwskazanie do wykonania NMR. Jednak w przypadku omawianej grupy wskazania do przeprowadzenia tego badania należy rozważyć niezwykle starannie. Trzeba pamiętać, aby u pacjentów, którzy nie są zależni od stymulatora, był on wyłączony na czas badania, zaś u chorych, którzy są zależni od stymulatora przeprowadzany w tryb stymulacji asynchronicznej (XOO). Diatermia elektryczna Diatermia elektryczna to leczenie prądem zmiennym o wysokiej częstotliwości i napięciu, który w zależności od zastosowanych parametrów rozgrzewa lub zwęgla, niszcząc miejscowo tkankę. Rozróżniamy diatermię krótkofalową i diatermię chirurgiczną. Pierwsza, będąca jedną z technik rehabilitacyjnych, jest całkowicie przeciwwskazana u chorych z kardiostymulatorami lub ICD. Terminem diatermii chirurgicznej, nazywanej również elektrokoagulacją, określa się szereg technik chirurgicznych wykorzystujących wysoką temperaturę do lokalnego niszczenia tkanek. Zaletą technik elektrochirurgicznych jest dobra kontrola hemostazy, dlatego w wielu rodzajach zabiegów są one nie do zastąpienia. Techniki diatermii chirurgicznej dzieli się na jedno- i dwubiegunowe. Techniki dwubiegunowe wykorzystuje się do lokalnej kontroli hemostazy. W ich przypadku prąd płynie pomiędzy ramionami pęsety, koagulując znajdującą się między nimi tkankę. Ponieważ przepływ prądu jest zlokalizowany i w praktyce w żadnym przypadku nie przebiega poprzez okolice ciała, w których znajduje się układ stymulujący, więc niebezpieczeństwo jego uszkodzenia lub istotnej interferencji z jego czynnością w praktyce nie istnieje [38]. W przypadku diatermii jednobiegunowej, czyli popularnego noża elektrycznego, ciało chorego jest przewodnikiem dla prądu, dlatego układ stymulujący jest narażony na oddziaływanie przepływającego prądu. W trakcie zabiegów z użyciem noża elektrycznego przepływający prąd o wysokiej częstotliwości poprzez interakcję z układem sterującym stymulatora lub kardiowertera może doprowadzić do wyzwolenia nieadekwatnej stymulacji, zmiany trybu stymulacji (mode switch), włączenia systemu tłumienia zakłóceń (noise reversion) lub fałszywej detekcji tachyarytmii [39]. W związku z tym stosowanie noża elektrycznego u pacjentów z włączonym implantowanym ICD obarczone jest ryzykiem sprowokowania wyładowania kardiowertera. W stymulatorach z opcją adaptacji rytmu, sterowanych wentylacją minutową, może ponadto dojść do przyspieszenia częstości stymulacji do górnego limitu [40]. Najpoważniejszą konsekwencją stosowania jednobiegunowej elektrokoagulacji jest przypadkowe przeprogramowanie, zresetowanie lub całkowite uszkodzenie wszczepionego urządzenia, co u zależnego od stymulatora chorego może mieć negatywne skutki. W starszych typach stymulatorów do zresetowania dochodziło w ok. 21% przypadków zabiegów z użyciem noża elektrycznego. Nowsze stymulatory, nieco mniej wrażliwe, mogą zostać zresetowane w ok. 7% przypadków, przy czym prawdopodobieństwo zajścia tego zjawiska jest tym większe, im bliżej układu stymulującego znajduje się nóż elektryczny [41, 42]. W sytuacji pracy nożem elektrycznym w bezpośrednim sąsiedztwie stymulatora opisywano przykłady trwałego uszkodzenia urządzenia, które może przybrać formę przetrwałej szybkiej stymulacji (runway syndrom) [43] lub trwałego braku skutecznej stymulacji [44, 45]. 265
6 Folia Cardiol. 2005, tom 12, nr 4 Techniki elektrochirurgiczne są aktualnie na tyle rozpowszechnione, że ich stosowanie u chorych z implantowanym stymulatorem jest nie do uniknięcia. Każdy zabieg powinien być przeprowadzany w szpitalu, najlepiej w asyście lekarza zajmującego się kontrolą i implantacją stymulatorów. Jeżeli zabiegowi wymagającemu stosowania elektrochirurgii poddawana jest osoba z ICD, urządzenie to powinno być przed rozpoczęciem wyłączone lub inaktywowane na czas zabiegu poprzez przyklejenie ponad kieszenią defibrylatora magnesu. Jeśli pacjent nie jest zależny od stymulatora, zalecane jest jego wyłączenie przed zabiegiem. W przypadku chorych, którzy są zależni od stymulatora, aby zapobiec wywołanemu przez nóż elektryczny zahamowaniu funkcji stymulatora, sugeruje się niekiedy umieszczenie na czas zabiegu magnesu ponad lożą stymulatora w celu przeprowadzenia go w tryb stymulacji asynchronicznej. Sposób ten uważa się jednak za kontrowersyjny [46]. Zasady opieki nad osobą z implantowanym stymulatorem, która ma zostać poddana zabiegowi z wykorzystaniem technik elektrochirurgicznych, zostały przedstawione przez Hayes i Strathmore podsumowane w tabeli 1. Radioterapia Pierwsze, stosunkowo prymitywne stymulatory wytwarzane w latach 70. były odporne na promieniowanie przenikliwe stosowane w radioterapii. Nowoczesne stymulatory i ICD, zbudowane z elementów elektronicznych opartych na półprzewodnikach wykorzystujących tlenki metali, są znacznie bardziej wrażliwe na działanie promieniowania jonizującego [47]. Efekty wpływu promieniowania jonizującego na czynność urządzeń wszczepialnych mogą przybierać postać nieistotnych odchyleń od prawidłowej czynności (inhibicja lub wyzwalanie pojedynczych pobudzeń, drobne zmiany w charakterystyce impulsu lub wielkości zaprogramowanych parametrów) lub poważnych, potencjalnie groźnych dla chorego zaburzeń. Nieistotna z klinicznego punktu widzenia dysfunkcja urządzenia wszczepialnego może wystąpić już przy względnie małych, (sięgających 2 Gy) dawkach promieniowania jonizującego [48]. Istotne zaburzenia w pracy stymulatorów odnotowano przy dawkach Gy [49 51]. Należy jednak pamiętać, że powyższe wielkości dawek uzyskano w badaniach, w których dawkę sumaryczną osiągano poprzez naświetlanie stymulatora in vitro krótkimi pulsami promieniowania o stosunkowo dużej gęstości dawki. Lepiej odpowiadający rzeczywistości klinicznej system naświetlania, w którym pojedyncza dawka nie przekraczała 3 Gy, pozwolił na uzyskanie łącznej dawki 60 Gy bez dysfunkcji badanych stymulatorów [52]. Wrażliwość urządzeń wszczepialnych na dzia- Tabela 1. Zasady przeprowadzania zabiegu chirurgicznego z wykorzystaniem elektrokoagulacji u chorych z implantowanym kardiostymulatorem lub kardiowerterem-defibrylatorem (wg [1]) Table 1. Principles of surgical procedures with usage of electrocoagulation in patients with implanted pacemaker or ICD (based on reference [1]) Czynności, które należy wykonać przed zabiegiem: identyfikacja stymulatora i określenie trybu stymulacji po zresetowaniu (reset mode) kontrola parametrów stymulacji: progi, stan baterii, opór elektrody dezaktywacja zbędnych funkcji (rate respose, vario itp.) wydrukowanie pełnej informacji o stanie stymulatora Zasady, których należy przestrzegać w czasie zabiegu: umieszczenie biernej elektrody tak daleko od stymulatora jak to możliwe ciągłe monitorowanie chorego za pomocą pulsoksymetru (zapis EKG może być nieczytelny z powodu zakłóceń) zabezpieczenie łatwego dostępu do programatora jeżeli to możliwe, zastosowanie koagulacji dwubiegunowej nieużywanie noża elektrycznego blisko stymulatora stosowanie krótkich koagulacji w krótkich aplikacjach jeśli stymulacja po zresetowaniu powoduje niestabilność hemodynamiczną, reprogramowanie stymulatora tak szybko jak to możliwe rozważenie włączenia na czas zabiegu trybu VVT Czynności, które należy wykonać po zabiegu: ponowne dokładne sprawdzenie stymulatora jeżeli to niezbędne, przeprogramowanie 266
7 A. Sokal i wsp., Zakłócenia pracy urządzeń wszczepialnych łanie promieniowania jonizującego zwiększała się wraz ze wzrostem grubości warstwy półprzewodników metalowo-nadtlenkowych (CMOS, complementary metal-oxide semiconductor) [53]. Co ciekawe, w tej samej pracy wykazano, że większość ICD reagowała na napromieniowanie łączną dawką zbliżoną do 50 Gy istotnym wydłużeniem czasu detekcji i ładowania. Stosowane w radioterapii akceleratory liniowe ze względu na wytwarzane pole elektryczne mogą być źródłem interferencji niezależnych od wielkości dawki promieniowania. Ich wpływ na czynność stymulatorów może być różnorodny i nieprzewidywalny: od kompletnego zahamowania funkcji stymulatora, po zupełny brak jakiejkolwiek reakcji [54, 55]. Przeważnie zakłócenia w pracy urządzeń wszczepialnych pojawiają się w trakcie uruchamiania i wyłączania akceleratora [56]. Zalecenia przedstawiające zasady radioterapii u chorych ze stymulatorem lub ICD ujęto w tabeli 2. Inne szpitalne źródła interferencji elektromagnetycznych Poza dokładniej omówionymi powyżej szpitalnymi źródłami interferencji elektromagnetycznych istnieją również inne procedury lub urządzenia, które mogą w szczególnych sytuacjach powodować sporadycznie podobne efekty. Należy do nich: litotrypsja ultradźwiękowa, zewnętrzna kardiowersja i defibrylacja oraz ablacja prądem RF. Występujące w trakcie tych procedur ewentualne zaburzenia funkcji urządzeń wszczepialnych mają charakter incydentalny i nie wiążą się z potencjalnymi konsekwencjami dla chorego. Ponadto w trakcie procedur, takich jak ablacja prądem RF lub kardiowersja wykonujący je personel jest z reguły bardzo dobrze przygotowany do przeciwdziałania ewentualnym niekorzystnym dla pacjenta zaburzeniom w funkcji stymulatora lub ICD. Pozaszpitalne źródła interferencji elektromagnetycznych Telefony komórkowe Mimo masowego stosowania telefonów komórkowych do tej pory pojawiło się tylko jedno doniesienie o uszkodzeniu stymulatora wywołanym przez ich używanie [57]. Wpływowi telefonów komórkowych na pracę urządzeń wszczepialnych poświęcono wiele badań [58 65]. Znaczna rozbieżność wyników wiąże się faktem, że istnieją różnorodne systemy telefonii komórkowej różniące się zarówno częstotliwością pracy, jak i sposobem kodowania informacji. Moc promieniowania elektromagnetycznego wytwarzanego przez telefony zależy od odległości od stacji bazowej i liczby aktywnych telefonów w sieci i może zmieniać się w trakcie rozmowy. Różnorodność systemów telefonii komórkowej oraz zależność mocy sygnału od stanu sieci i odległości od stacji bazowej powodują, że wyniki wielu badań różnią się znacznie miedzy sobą. W dużym wieloośrodkowym badaniu, do którego zakwalifikowano 980 chorych z implantowanym kardiostymulatorem, częstość różnych, lecz z reguły nieistotnych incydentów wynosiła 20%, w tym 7,2% klinicznie istotnych (odczuwanych przez pacjentów). Po wyeliminowaniu z analizy jednego modelu telefonu zmniejszyła się ona do odpowiednio 13,1% i 2,8%. Wynik ten uzyskano jednak w warunkach nietypowej ekspozycji, w której telefon przykładano bezpośrednio do loży stymulatora. Liczba incydentów, szczególnie tych odczuwanych przez pacjentów, jeśli telefon znajdował się przy uchu, była bardzo mała. Zwiększała się istotnie w przypadku stymulatorów dwujamowych (55,8%) w porównaniu z jednojamowymi (28,9%). Ponadto w nowoczesnych stymulatorach wyposażonych w filtr systemu sterowania dochodziło do znikomej liczby incydentów: 0,4 0,8% w porównianiu z 28,9% i 55,8% Tabela 2. Zasady prowadzenia bezpiecznej radioterapii u chorego ze stymulatorem (wg [47]) Table 2. Rules of safe radiotherapy in patients with pacemaker (based on reference [47]) Stan pacjenta i stymulatora powinien ocenić wykwalifikowany personel przed każdą sesją radioterapii i po niej Radioterapia betatronem jest ściśle przeciwwskazana u chorych ze stymulatorem lub ICD Stymulator lub ICD powinny znajdować się poza skolimowaną wiązką promieniowania jonizującego Pacjenta należy uważnie obserwować i monitorować w trakcie ekspozycji, aby w porę dostrzec ewentualną dysfunkcję urządzenia Dawka, którą otrzymał stymulator w trakcie całego leczenia, powinna być oszacowana i odnotowana w dokumentacji chorego. Nie powinna ona przekraczać 2 Gy Jeżeli nie jest możliwe utrzymanie sumarycznej bezpiecznej dawki w podanej powyżej wielkości, można rozważyć usunięcie stymulatora na czas naświetlań lub jego reimplantację w inne miejsce 267
8 Folia Cardiol. 2005, tom 12, nr 4 w starszych typach [61]. W innym badaniu, w którym oceniono wpływ różnego typu telefonów komórkowych na pracę stymulatorów z filtrowanym układem sterowania w grupie 100 chorych z wszczepionym stymulatorem, nie stwierdzono żadnego incydentu, niezależnie od położenia telefonu względem stymulatora [60]. Zagadnieniu interferencji telefonów komórkowych z czynnością ICD poświęcono w ostatnim okresie również wiele publikacji. Podobnie jak w przypadku stymulatorów prace z połowy lat 90. sugerują możliwość zaistnienia istotnych klinicznie interferencji, z nieadekwatną interwencją włącznie [65]. W nowszych pracach, w których oceniano kardiowertery III generacji, nie wykazano istotnych zakłóceń czynności badanych ICD. Mimo znacznej redukcji incydentów zakłóceń pracy kardiostymulatorów i ICD spowodowanej zastosowaniem filtrów układu sterowania i nowych konstrukcji tych ostatnich niebezpieczeństwo, że któryś z nowo wprowadzonych systemów łączności bezprzewodowej może być przyczyną istotnych klinicznie interferencji, wymaga ostrożnego podchodzenia do wszelkich nowości w tej dziedzinie i przeprowadzania regularnych badań testujących nowe technologie pod względem ich kompatybilności z urządzeniami wszczepialnymi. Ponieważ we wszystkich przeprowadzonych badaniach wystąpienie istotnych incydentów wiązało się z przyłożeniem telefonu bezpośrednio do loży stymulatora, należy zalecać chorym trzymanie i używanie telefonów komórkowych w odległości większej niż 10 cm od stymulatora. Systemy zabezpieczenia antykradzieżowego Elektroniczne systemy antykradzieżowe stanowią aktualnie stałe wyposażenie wielkich domów towarowych i mniejszych sklepów samoobsługowych. Zasada działania takiego urządzenia opiera się na wymianie sygnałów elektromagnetycznych pomiędzy przymocowanym do towaru czujnikiem a elektroniczną bramką znajdującą się przy wyjściu ze sklepu. Bramka ta jest więc źródłem stosunkowo silnego pola elektromagnetycznego o modulowanej częstotliwości, którego działaniu poddawani są klienci wychodzący ze sklepu. Istnieją doniesienia o nieadekwatnych interwencjach ICD pod wpływem elektronicznych systemów antywłamaniowych [66 68]. Systematyczne badania ewentualnych interferencji pomiędzy urządzeniami wszczepialnymi a bramkami systemów antykradzieżowych pozwoliły na ocenę rzeczywistej skali potencjalnego zagrożenia dla chorego. Przegląd dostępnej literatury pozwala na stwierdzenie, że oddziaływania pomiędzy wszczepionymi kardiostymulatorami a systemami zapobiegającymi kradzieżom są przyczyną stosunkowo częstych zaburzeń w pracy kardiostymulatorów, rzadko jednak klinicznie istotnych. W badaniu przeprowadzonym przez Mugica i wsp., do którego włączono łącznie 204 pacjentów, zaburzenia w pracy kardiostymulatora obserwowano u 17% badanych, przy czym ponad 2-krotnie częściej do zaburzeń dochodziło w przypadku systemu typu akustomagnetycznego niż systemu audiomagnetycznego. Stwierdzone zaburzenia były najczęściej związane z upośledzeniem sensingu. Kilkakrotnie odnotowano pauzy spowodowane wypadnięciem pojedynczych pobudzeń. Zaburzenia występowały jedynie w momencie bezpośredniego kontaktu chorego z polem magnetycznym bramki [71]. Podobne wyniki uzyskano w mniejszych badaniach in vitro [72] i in vivo [73]. W żadnym przypadku nie doszło do przeprogramowania stymulatora lub innych trwałych następstw. Mimo wstępnych kazuistycznych doniesień analiza wpływu systemów antykradzieżowych na czynność ICD również nie wykazała istotnych klinicznie odchyleń od prawidłowej funkcji tych urządzeń. W badaniu SPICED-TEAS 25 ochotników z ICD i 50 ze stymulatorami zostało poddanych działaniu pola magnetycznego wytwarzanego przez 6 różnych systemów antykradzieżowych. Upośledzenie czynności stymulatora polegające na stymulacji asynchronicznej, tachykardii wywołanej zakłóceniami w kanale przedsionkowym, wypadnięciu jednego lub kilku pobudzeń odnotowano w 48/50 przypadkach systemów akustomagnetycznych i 2/50 przypadkach systemów audiomagnetycznych. Objawy kliniczne zgłaszane przez poddanych badaniu chorych ograniczały się do uczucia słabości lub szybkiego bicia serca. Ustępowały bezpośrednio po opuszczeniu przez pacjentów bramki systemu. Nie odnotowano zaburzeń w funkcji ICD [70]. Dokonana przez Groha i wsp. [69] ocena funkcji ICD u 169 chorych poddanych działaniu pola wytwarzanego przez 3 typowe systemy antykradzieżowe wykazała brak jakichkolwiek interferencji w trakcie przechodzenia przez bramkę. U 7 spośród 169 pacjentów zatrzymanych w bramce wystąpiły interferencje mogące zaowocować nieadekwatną interwencją. Zatem systemy zapobiegające kradzieżom nie stwarzają istotnego zagrożenia dla osoby ze stymulatorem lub ICD, jeśli zachowuje się on w sposób typowy i przechodzi przez bramkę. Chory musi być jednak poinstruowany, aby zachował ostrożność i nie zatrzymywał się w obrębie bramki lub w jej sąsiedztwie. 268
9 A. Sokal i wsp., Zakłócenia pracy urządzeń wszczepialnych Inne pozaszpitalne źródła interferencji elektromagnetycznych Poza potencjalnie możliwą ekspozycją zawodową inne środowiskowe źródła ewentualnych interferencji elektromagnetycznych, w tym urządzenia gospodarstwa domowego, wykrywacze metali i urządzenia ogrodnicze, nie stwarzają zagrożenia związanego z zaburzeniem czynności urządzeń wszczepialnych. Opisane dotychczas kazuistyczne przypadki takich interakcji wynikały z wadliwego działania stymulatora lub ICD czy urządzenia będącego źródłem zakłócenia. Należy jednak pamiętać, że szybki rozwój technologii wiąże się z pojawianiem się na rynku coraz to nowszych urządzeń mogących być przyczyną różnorakich interakcji ze stymulatorami lub ICD i chory powinien być poinformowany o konieczności zachowania szczególnej ostrożności przy korzystaniu z urządzeń dotychczas nieznanych. Streszczenie Wraz z rozwojem techniki medycznej co roku rośnie liczba wszczepialnych urządzeń kontrolujących rytm serca. Zwiększa się również stopień komplikacji implantowanych urządzeń, co w połączeniu z pojawianiem się w codziennym życiu oraz środowisku zawodowym i szpitalnym coraz to nowych emiterów zakłóceń elektromagnetycznych sprawia, że problem ewentualnych interferencji między tymi urządzeniami a środowiskiem zewnętrznym jest coraz bardziej istotny. Do najczęstszych reakcji urządzenia wszczepialnego na zewnętrzne zakłócenia elektromagnetyczne należą: wyhamowanie lub wytworzenie pojedynczego pobudzenia, całkowite zahamowanie stymulacji, przejście do trybu stymulacji awaryjnej oraz zwiększenie częstości rytmu lub nieprawidłowa częstość rytmu. Możliwe źródła zakłóceń elektromagnetycznych można podzielić na: wewnątrzszpitalne, środowiskowe i zawodowe. W pierwszej grupie do najczęściej spotykanych urządzeń medycznych emitujących zakłócenia elektromagnetyczne należą: magnetyczny rezonans jądrowy oraz urządzenia do diatermii i radioterapii. W środowisku zewnątrzszpitalnym najczęściej spotykanym źródłami interferencji elektromagnetycznych są telefony komórkowe oraz systemy zabezpieczenia przed kradzieżą. W niniejszej pracy omówiono aktualny stan wiedzy na temat możliwych interferencji oraz ich konsekwencji klinicznych. (Folia Cardiol. 2005; 12: ) urządzenia wszczepialne, rozruszniki, defibrylatory kardiowertery, zakłócenia elektromagnetyczne Piśmiennictwo 1. Hayes D.L., Strathmore N.F. Electromagnetic interference with implantable devices. W: Ellenbogen K.A., Kay G.N., Wilkoff B.L. red. Clinical cardiac pacing and defibrillation. Wyd. II. W.B. Saunders Company, Philadelphia 2000; Piński S.L., Trohman R.G. Interference in implanted cardiac devices. Cz. I. Pacing Clin. Electrophysiol. 2002; 25: Jones G.K., Brady G.H. Considerations for ventricular fibrillation detection by implantable cardioverter defibrillators. Am. Heart J. 1994; 127 (4 cz. 2): Brumwell D.A., Kroll K., Lehman M.H. The amplifier: Sensing the depolarization. W: Kroll M.W., Lehman M.H. red. Implantable cardioverter-defibrillator therapy: engineering-clinical interface. Norwell, Massachusetts, Kluwer Academical Publishers 1966; Strathmore N.F. Interference in cardiac pacemakers. W: Ellenbogen K.A., Kay G.N., Wilkoff B.L. red. Clinical cardiac pacing. W.B. Saunders, Philadelphia 1995; Barold S.S. Timing cycles and operative haracteristics of pacemakers. W: Ellenbogen K.A., Kay G.N., Wilkoff B.L. red. Clinical cardiac pacing. W.B. Saunders, Philadelphia 1995; Bilitch M., Cosby R.S., Cafferky E.A. Ventricular fibrillation and competitive pacing. N. Engl. J. Med. 1967; 276: Saeed M., Link M., Mahapatra S. i wsp. Analysis of intracardiac electrograms showing monomorphic 269
10 Folia Cardiol. 2005, tom 12, nr 4 ventricular tachycardia in patients with implantable cardioverter-defibrillators. Am. J. Cardiol. 2000; 85: Chew E.W., Troughear R.H., Kuchar D.L., Thorburn C.W. Inappropriate rate change in minute ventilation rate responsive pacemakers due to interference by cardiac monitors. Pacing Clin. Electrophysiol. 1997; 20 (2 cz. 1): Southorn P.A., Kamath G.S., Vasdev G.M., Hayes D.L. Monitoring equipment induced tachycardia in patients with minute ventilation rate-responsive pacemakers. Br. J. Anaesth. 2000; 84: McIvor M.E., Reddinger J., Floden E., Sheppard R.C. Study of Pacemaker and Implantable Cardioverter Defibrillator Triggering by Electronic Article Surveillance Devices (SPICED TEAS) Pacing Clin. Electrophysiol. 1998; 21: Fontaine J.M., Mohamed F.B., Gottlieb C., Callans D.J., Marchlinski F.E. Rapid ventricular pacing in a pacemaker patient undergoing magnetic resonance imaging. Pacing Clin. Electrophysiol. 1998; 21: Seifert T., Block M., Borggrefe M., Breithardt G. Erroneous discharge of an implantable cardioverter defibrillator caused by an electric razor. Pacing Clin. Electrophysiol. 1995; 18: Manolis A.G., Katsivas A.G., Vassilopoulos C.V., Louvros N.E. Implantable cardioverter defibrillatoran unusual case of inappropriate discharge during showering. Implantable cardioverter defibrillator-an unusual case of inappropriate discharge during showering. J. Interv. Card. Electrophysiol. 2000; 4: Man K.C., Davidson T., Langberg J.J., Morady F., Kalbfleisch S.J. Interference from a hand held radiofrequency remote control causing discharge of an implantable defibrillator. Pacing Clin. Electrophysiol. 1993; 16: Kolb C., Zrenner B., Schmitt C. Incidence of electromagnetic interference in implantable cardioverter defibrillators. Pacing Clin. Electrophysiol. 2001; 24 (4 cz. 1): Marco D., Eisinger G., Hayes D.L. Testing of work environnments for electromagnetic interference. Pacing Clin. Electrophysiol. 1992; 15 (11 cz. 19.2): Butrous G.S., Bexton R.S., Barton D.G., Male J.C., Camm A.J. Interference with the pacemakers of two workers at electricity substations. Br. J. Ind. Med. 1983; 40: Mehdirad A., Love C., Nelson S., Schaal S., Collins J., Huffman K. Alternating current electrocution detection and termination by an implantable cardioverter defibrillator. Pacing Clin. Electrophysiol. 1997; 20: Sakakibara Y., Mitsui T. Concerns about sources of electromagnetic interference in patients with pacemakers. Jpn. Heart J. 1999; 40: Luechinger R., Duru F., Scheidegger M.B., Boesiger P., Candinas R. Force and torque effects of a 1.5-Tesla MRI scanner on cardiac pacemakers and ICDs. Pacing Clin. Electrophysiol. 2001; 24: Luechinger R., Duru F., Zeijlemaker V.A., Scheidegger M.B., Boesiger P., Candinas R. Pacemaker reed switch behavior in 0.5, 1.5, and 3.0 Tesla magnetic resonance imaging units: are reed switches always closed in strong magnetic fields? Pacing Clin. Electrophysiol. 2002; 25: Holmes D.R. Jr, Hayes D.L., Gray J.E., Merideth J. The effects of magnetic resonance imaging on implantable pulse generators. Pacing Clin. Electrophysiol. 1986; 9: Hayes D.L., Holmes D.R. Jr, Gray J.E. Effect of 1.5 tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakers. J. Am. Coll. Cardiol. 1987; 10: Fontaine J.M., Mohamed F.B., Gottlieb C., Callans D.J., Marchlinski F.E. Rapid ventricular pacing in a pacemaker patient undergoing magnetic resonance imaging. Pacing Clin. Electrophysiol. 1998; 21: Gimbel J.R., Johnson D., Levine P.A., Wilkoff B.L. Safe performance of magnetic resonance imaging on five patients with permanent cardiac pacemakers. Pacing Clin. Electrophysiol. 1996; 19: Martin E.T., Coman J.A., Shellock F.G., Pulling C.C., Fair R., Jenkins K. Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla. J. Am. Coll. Cardiol. 2004; 43: Sommer T., Vahlhaus C., Lauck G. i wsp. MR imaging and cardiac pacemakers: in-vitro evaluation and in-vivo studies in 51 patients at 0.5 T. Radiology 2000; 215: Iberer F., Justich E., Stenzel W. i wsp. Nuclear magnetic resonance imaging of a patient with implanted transvenous pacemaker. Herzschrittmacher 1987; 7: Alagona P., Toole J., Maniscalco B. i wsp. Nuclear magnetic resonance imaging in a patient with a DDD pacemaker (letter). Pace 1989; 12: Inbar S., Larson J., Burt T. i wsp. Nuclear magnetic resonance imaging in a patient with a pacemaker. Am. J. Med. Sci. 1993; 305: Achenbach S., Moshage W., Kuhn I., Schibgilla V., Bachmann K. Fallvorstellung: Kernspintomographie bei einem Patienten mit Zweikammer-Schrittmachersystem. Z. Kardiol. 1995; 84 (supl.): Garcia-Bolao I., Albaladejo V., Benito A., Alegria E., Zubieta J.L. Magnetic resonance imaging in a patient 270
11 A. Sokal i wsp., Zakłócenia pracy urządzeń wszczepialnych with a dual-chamber pacemaker. Acta Cardiol. 1998; 53: Vahlhaus C., Sommer T., Lewalter T. i wsp. Interference with cardiac pacemakers by magnetic resonance imaging: are there irreversible changes at 0.5 Tesla? Pacing Clin. Electrophysiol. 2001; 24 (4 cz. 1): Pennell D.J. Cardiac magnetic resonance with a pacemaker in situ: can it be done. J. Cardiovasc. Mag. Res. 1999; 1: 72 (streszczenie). 36. Achenbach S., Moshage W., Diem B., Bieberle T., Schibgilla V., Bachmann K. Effects of magnetic resonance imaging on cardiac pacemakers and electrodes. Am. Heart J. 1997; 134: Gimbel J.R., Kanal E. Can patients with implantable pacemakers safely undergo magnetic resonance imaging? J. Am. Coll. Cardiol. 2004; 43: Batra Y.K., Bali I.M. Effect of coagulating and cutting current on a demand pacemaker during transurethral resection of the prostate. A case report. Can. Anaesth. Soc. J. 1978; 25: Casavant D., Haffajee C., Stevens S., Pacetti P. Aborted implantable cardioverter defibrillator shock during facial electrosurgery. Pacing Clin. Electrophysiol. 1998; 21: Wong D.T., Middleton W. Electrocautery-induced tachycardia in a rate-responsive pacemaker. Anesthesiology 2001; 94: Lamas G.A., Antman E.M., Gold J.P., Braunwald N.S., Collins J.J. Pacemaker backup-mode reversion and injury during cardiac surgery. Ann. Thorac. Surg. 1986; 41: Roman-Gonzalez J., Hyberger L.K., Hayes D.L. Is electrocautery still a clinical significant problem with contemporary technology? PACE 2001; 24: Heller L.I. Surgical electrocautery and the runaway pacemaker syndrome. Pacing Clin. Electrophysiol. 1990; 13: Peters R.W., Gold M.R. Reversible prolonged pacemaker failure due to electrocautery. J. Interv. Card. Electrophysiol. 1998; 2: Nercessian O.A., Wu H., Nazarian D., Mahmud F. Intraoperative pacemaker dysfunction caused by the use of electrocautery during a total hip arthroplasty. J. Arthroplasty 1998; 13: Atlee J.L., Bernstein A.D. Cardiac rhythm management devices (part I): indications, device selection, and function. Anesthesiology 2001; 95: Last A. Radiotherapy in patients with cardiac pacemakers. Br. J. Radiol. 1998; 71: Marbach J.R., Sontag M.R., Van Dyk J., Wolbarst A.B. Management of radiation oncology patients with implanted cardiac pacemakers: report of AAPM Task Group No. 34. American Association of Physicists in Medicine. Med. Phys. 1994; 21: Katzenberg C.A., Marcus F.I., Heusinkveld R.S., Mammana R.B. Pacemaker failure due to radiation therapy. Pacing Clin. Electrophysiol. 1982; 5: Pourhamidi A.H. Radiation effect on implanted pacemakers. Chest 1983; 84: Lewin A.A., Serago C.F., Schwade J.G., Abitbol A.A., Margolis S.C. Radiation induced failures of complementary metal oxide semiconductor containing pacemakers: a potentially lethal complication. Int. J. Radiat. Oncol. Biol. Phys. 1984; 10: Maxted K.J. The effect of therapeutic x-radiation on a sample of pacemaker generators. Phys. Med. Biol. 1984; 29: Rodriguez F., Filimonov A., Henning A., Coughlin C., Greenberg M. Radiation-induced effects in multiprogrammable pacemakers and implantable defibrillators. Pacing Clin. Electrophysiol. 1991; 14: Venselaar J.L. The effects of ionizing radiation on eight cardiac pacemakers and the influence of electromagnetic interference from two linear accelerators. Radiother. Oncol. 1985; 3: Ngu S.L., O Meley P., Johnson N., Collins C. Pacemaker function during irradiation: in vivo and in vitro effect. Australas. Radiol. 1993; 37: Salmi J., Eskola H.J., Pitkanen M.A., Malmivuo J.A. The influence of electromagnetic interference and ionizing radiation on cardiac pacemakers. Strahlenther. Onkol. 1990; 166: Yesil M., Bayata S., Postaci N., Aydin C. Pacemaker inhibition and asystole in a pacemaker dependent patient. Pacing Clin. Electrophysiol. 1995; 18: Hayes D.L., Carrillo R.G., Findlay G.K., Embrey M. State of the science: pacemaker and defibrillator interference from wireless communication devices. Pacing Clin. Electrophysiol. 1996; 19: Irnich W., Batz L., Muller R., Tobisch R. Electromagnetic interference of pacemakers by mobile phones. Pacing Clin. Electrophysiol. 1996; 19: Hekmat K., Salemink B., Lauterbach G. i wsp. Interference by cellular phones with permanent implanted pacemakers: an update. Europace 2004; 6: Hayes D.L., Wang P.J., Reynolds D.W. i wsp. Interference with cardiac pacemakers by cellular telephones. N. Engl. J. Med. 1997; 336: Naegeli B., Osswald S., Deola M., Burkart F. Intermittent pacemaker dysfunction caused by digital mobile telephones. J. Am. Coll. Cardiol. 1996; 27: Elshershari H., Celiker A., Ozer S., Ozme S. Influence of D-net (EUROPEAN GSM-standard) cellular telephones on implanted pacemakers in children. Pacing Clin. Electrophysiol. 2002; 25: Chiladakis J.A., Davlouros P., Agelopoulos G., Manolis A.S. In-vivo testing of digital cellular telephones in patients with implantable cardioverter- -defibrillators. Eur. Heart J. 2001; 22:
12 Folia Cardiol. 2005, tom 12, nr Bassen H.I., Moore H.J., Ruggera P.S. Cellular phone interference testing of implantable cardiac defibrillators in vitro. Pacing Clin. Electrophysiol. 1998; 21: McIvor M.E. Environmental electromagnetic interference from electronic article surveillance devices: interactions with an ICD. Pacing Clin. Electrophysiol. 1995; 18 (12 cz. 1): Mathew P., Lewis C., Neglia J., Krol R.B., Saksena S. Interaction between electronic article surveillance systems and implantable defibrillators: insights from a fourth generation ICD. Pacing Clin. Electrophysiol. 1997; 20: Santucci P.A., Haw J., Trohman R.G., Pinski S.L. Interference with an implantable defibrillator by an electronic antitheft-surveillance device. N. Engl. J. Med. 1998; 339: Groh W.J., Boschee S.A., Engelstein E.D. i wsp. Interactions between electronic article surveillance systems and implantable cardioverter-defibrillators. Circulation 1999; 100: McIvor M.E., Reddinger J., Floden E., Sheppard R.C. Study of Pacemaker and Implantable Cardioverter Defibrillator Triggering by Electronic Article Surveillance Devices (SPICED TEAS). Pacing Clin. Electrophysiol. 1998; 21: Mugica J., Henry L., Podeur H. Study of interactions between permanent pacemakers and electronic antitheft surveillance systems. Pacing Clin. Electrophysiol. 2000; 23: Lucas E.H., Johnson D., McElroy B.P. The effects of electronic article surveillance systems on permanent cardiac pacemakers: an in vitro study. Pacing Clin. Electrophysiol. 1994; 17 (11 cz. 2): Dodinot B., Godenir J.P., Costa A.B. Electronic article surveillance: a possible danger for pacemaker patients. Pacing Clin. Electrophysiol. 1993; 16 (1 cz. 1):
Rezonans magnetyczny u pacjentów z implantowanym układem stymulującym lub ICD
Rezonans magnetyczny u pacjentów z implantowanym układem stymulującym lub ICD Zalecenia ESC 2013 dr med. Artur Oręziak Klinika Zaburzeń Rytmu Serca Instytut Kardiologii, Warszawa Potencjalne zagrożenia
Pacemaker. Maria Kawczyńska
Pacemaker Maria Kawczyńska Management of Radiation Oncology Patients with Implanted Cardiac Pacemakers 45 AAPM RPT Gotowy do publikacji 1992 r Opublikowany 1994 Co roku 47% pacjentów leczonych z powodu
Małgorzata Łodyga Klinika Zaburzeń Rytmu Serca Instytut Kardiologii
Małgorzata Łodyga Klinika Zaburzeń Rytmu Serca Instytut Kardiologii Serce zdrowej osoby bije średnio od 60 do 90 razy na minutę. Różne czynniki zewnętrzne takie jak: wysiłek fizyczny, stres, stany gorączkowe
Interferencje elektromagnetyczne u osób ze stymulatorem serca
Folia Cardiologica 1999, tom 1, Suplement I Interferencje elektromagnetyczne u osób ze stymulatorem serca Rodzaje energii mogące oddziaływać niekorzystnie na układ stymulujący Interferencje elektromagnetyczne
PAKIET I-poz.1 Oddział Kardiologii Stymulator jednojamowy SSIR z elektrodami (Podstawowy) Producent: Nazwa/numer katalogowy: Kraj pochodzenia:
PAKIET I-poz.1 Oddział Kardiologii Stymulator jednojamowy SSIR z elektrodami (Podstawowy) Kraj pochodzenia: 1 Żywotność stymulatora min 8 lat (nastawy nominalne) 2 Waga max. 30 [g] Do 30 g 10 pkt powyżej
Ewa Lewicka Klinika Kardiologii i Elektroterapii Serca Gdański Uniwersytet Medyczny
Ewa Lewicka Klinika Kardiologii i Elektroterapii Serca Gdański Uniwersytet Medyczny II Konferencja Edukacyjna Czasopisma Anestezjologia Intensywna Terapia Sopot, 4-5.04.2014 Stymulator serca Kardiowerter-defibrylator
Wpływ telefonu komórkowego na pracę rozrusznika serca
PRACA ORYGINALNA Folia Cardiol. 2002, tom 9, nr 2, 117 122 Copyright 2002 Via Medica ISSN 1507 4145 Wpływ telefonu komórkowego na pracę rozrusznika serca Adam Curyło, Grzegorz Bilo, Mariusz Ślosarczyk
Przyczyny nieadekwatnych interwencji kardiowertera-defibrylatora
Przyczyny nieadekwatnych interwencji kardiowertera-defibrylatora Dr n. med. Aleksander Maciąg Pracownia Elektrofizjologii Klinicznej II Kliniki Choroby Wieńcowej Instytutu Kardiologii w Warszawie 1 Deklaracja
Anestezjologia Ratownictwo Nauka Praktyka / Anaesthesiology Rescue Medicine Science Practice
204 Anestezjologia i Ratownictwo 2012; 6: 204-211 ARTYKUŁ POGLĄDOWY/REVIEW PAPER Otrzymano/Submitted: 24.04.2012 Zaakceptowano/Accepted:25.05.2012 Akademia Medycyny Wpływ pola elektromagnetycznego związanego
PRZEDMIOT ZAMÓWIENIA - PARAMETRY JAKOŚCIOWE
Strona 1 z 7 PRZEDMIOT ZAMÓWIENIA - PARAMETRY JAKOŚCIOWE Załącznik Nr 2 do siwz Część 1 - Defibrylator - kardiowerter ICD-VR jednojamowy z elektrodami 1 Nazwa, nr katalogowy, producent 2 Rok produkcji
Załącznik Nr 3 do siwz OPIS PRZEDMIOTU ZAMÓWIENIA - PARAMETRY JAKOŚCIOWE. Część 1 - Defibrylator - kardiowerter ICD-VR jednojamowy z elektrodami
Strona 1 z 7 OPIS PRZEDMIOTU ZAMÓWIENIA - PARAMETRY JAKOŚCIOWE Załącznik Nr 3 do siwz Część 1 - Defibrylator - kardiowerter ICD-VR jednojamowy z elektrodami 1 Rok produkcji min. 201r. 2 Waga poniżej 80
Stymulator jednojamowy typu SSIR z elektrodami
Pakiet 1 Załącznik nr 2 SIWZ Formularz cenowy wraz z parametrami techniczno użytkowymi jednostko wa Stymulator jednojamowy typu SSIR z elektrodami zużycie na 12 m-cy oferowanego towaru, która będzie używana
Strategia bezpiecznego przeprowadzania badania metodą rezonansu magnetycznego u pacjenta z wszczepionym kardiowerterem-defibrylatorem
PRACA KAZUISTYCZNA Folia Cardiologica Excerpta 2006, tom 1, nr 8, 442 447 Copyright 2006 Via Medica ISSN 1896 2475 Strategia bezpiecznego przeprowadzania badania metodą rezonansu magnetycznego u pacjenta
Wysoki DFT jak postępować w przypadku jego wystąpienia.
Wysoki DFT jak postępować w przypadku jego wystąpienia. dr Jarosław Blicharz, Oddział Kardiologii, Szpital Wojewódzki im.św.łukasza SP ZOZ w Tarnowie DEFINICJE Wysoki próg defibrylacji (Defibrillation
Zastosowanie radioterapii u chorych ze stymulatorem serca
Współczesna Onkologia (2008) vol. 12; 9 (400 405) Od kilku lat na całym świecie obserwuje się ciągły wzrost liczby pacjentów ze stymulatorami serca wymagających radioterapii. W dostępnym piśmiennictwie
Przywrócenie rytmu zatokowego i jego utrzymanie
Przywrócenie rytmu zatokowego i jego utrzymanie Jak wspomniano we wcześniejszych artykułach cyklu, strategia postępowania w migotaniu przedsionków (AF) polega albo na kontroli częstości rytmu komór i zapobieganiu
Zastosowanie radioterapii u chorych ze stymulatorem serca
Współczesna Onkologia (2008) vol. 12; 9 (395 399) W ciągu ostatnich lat obserwuje się wzrost zachorowalności na choroby serca i tym samym rośnie liczba pacjentów, którzy wymagają wszczepienia stymulatorów
Opis przedmiotu zamówienia. Zestawienie wymaganych warunków i parametrów technicznych
Nr sprawy 105/PNE/SW/2011 Załącznik nr 1 do SIWZ Część 1 Opis przedmiotu zamówienia Zestawienie wymaganych warunków i parametrów technicznych Stymulator jednojamowy SSIR z elektrodami i zestawem do kaniulacji
Specjalistyczny Szpital im. dra Alfreda Sokołowskiego ul. A. Sokołowskiego Wałbrzych
ISO 9001 Specjalistyczny Szpital im. dra Alfreda Sokołowskiego ul. A. Sokołowskiego 4 58-309 Wałbrzych tel. 74/64 89 600 fax 74/ 64 89 746 www.zdrowie.walbrzych.pl szpitalsokolowski@zdrowie.walbrzych.pl
Elektrostymulacja i defibrylacja serca. Elektroniczna aparatura medyczna 1 Wykład - 8
Elektrostymulacja i defibrylacja serca Elektroniczna aparatura medyczna 1 Wykład - 8 1 Zaburzenia pracy serca Zbudowany z wyspecjalizowanych włókien mięśniowych, inicjowanie i koordynacja skurczów komór
POLSKIE TOWARZYSTWO RADIOTERAPII ONKOLOGICZNEJ
z 7. Wstęp Wzrost zachorowalności na nowotwory złośliwe powoduje konieczność szerszego stosowania radioterapii, również u pacjentów obciążonych kardiologicznie, którzy wymagają implantacji układu stymulującego
Shock on burst skuteczna, nieinwazyjna metoda indukcji migotania komór u chorych z kardiowerterem-defibrylatorem serca
PRC KZUISTYCZN Folia Cardiol. 2003, tom 10, nr 6, 823 827 Copyright 2003 Via Medica ISSN 1507 4145 Shock on burst skuteczna, nieinwazyjna metoda indukcji migotania komór u chorych z kardiowerterem-defibrylatorem
PAKIET IV poz. 1. STYMULATORY SERCA AUTOMATYCZNE PROSTE JEDNOJAMOWE SSIR w komplecie z elektrodą 2014 Parametry wymagane pkt 1-21 i oceniane pkt 19-22
PAKIET IV poz. 1. STYMULATORY SERCA AUTOMATYCZNE PROSTE JEDNOJAMOWE SSIR w komplecie z elektrodą 2014 Parametry wymagane pkt 1-21 i oceniane pkt 19-22 L.p. Opis parametru I Stymulator SSIR dla monitorowania
SPECYFIKACJA TECHNICZNA I PARAMETRY OCENIANE MATERIAŁÓW WSZCZEPIALNYCH
Załącznik nr 3A do SIWZ SPECYFIKACJA TECHNICZNA I Y OCENIANE MATERIAŁÓW WSZCZEPIALNYCH Część nr 1 : Stymulatory Pozycja nr 1: Stymulator dwujamowy DDDR z algorytmami zapobiegającymi napadom FA z elektrodami
Przy przyjęciu w EKG AFL z czynnością komór 120/min. Bezpośrednio przed zabiegiem, na sali elektrofizjologicznej,
Opis przypadku Częstoskurcz z szerokimi zespołami QRS u pacjenta po zawale serca czy zawsze VT? Wide QRS complex tachycardia in a patient after myocardial infarction: is it always ventricular tachycardia?
1- Wszczepialne kardiowertery- defibrylatory jednojamowe z elektrodą - 50 szt
ZAŁĄCZNIK NR 6 OPIS PRZEDMIOTU ZAMÓWIENIA ZESTAWIENIE WYMAGANYCH PARAMETRÓW Pakiet I Kardiowertery 1- Wszczepialne kardiowertery- defibrylatory jednojamowe z elektrodą - 50 szt Odp. Oferenta - Min. dwie
Anna PŁAWIAK-MOWNA 1, Andrzej KRAWCZYK 2. 1. Wprowadzenie
Anna PŁAWIAK-MOWNA 1, Andrzej KRAWCZYK 2 1 Uniwersytet Zielonogórski, Instytut Informatyki i Elektroniki 2 Centralny Instytut Ochrony Pracy Państwowy Instytut Badawczy Zaburzenia pracy wszczepialnych kardiowerterów
Opis przedmiotu zamówienia
Część Opis przedmiotu zamówienia Zestawienie wymaganych warunków i parametrów technicznych Kardiowertery-defibrylatory jednojamowe w standardzie Df/Df do wyboru Kardiowertery-defibrylatory jednojamowe
że korzystanie z telefonu komórkowego na zapis EKG rejestrowany metodą Holtera
PRACA ORYGINALNA Folia Cardiol. 2000, tom 7, nr 4, 321 326 Copyright 2000 Via Medica ISSN 1507 4145 Wpływ korzystania z telefonu komórkowego na zapis EKG rejestrowany metodą Holtera Magdalena Kumor, Rafał
Wojewódzki Szpital Specjalistyczny im. Janusza Korczaka w Słupsku
Wojewódzki Szpital Specjalistyczny im. Janusza Korczaka w Słupsku EN ISO 9001:2008 ul. Prof. Lotha 26 76-200 Słupsk tel. 059 8424867, fax 059 8428143 dyrekcja@szpital.slupsk.pl www.szpital.slupsk.pl WSS/
FORMULARZ CENOWY- ZADANIE 1 Stymulatory SSIR i DDDR pro MRI
FORMULARZ CENOWY- ZADANIE 1 Stymulatory SSIR i DDDR pro MRI Lp. Nazwa asortymentu J.m. Ilość Cena jedn. netto Wartość netto (poz. 4x5) Podatek VAT % Kwota Wartość brutto (poz. 6+8) Producent Nazwa Handlowa
Prowadzenie chorych z implantowanym kardiowerterem defibrylatorem serca
Prowadzenie chorych z implantowanym kardiowerterem defibrylatorem serca Medical care of patients post cardioverter defibrillator implantation Maciej Kempa, Klinika Kardiologii i Elektroterapii Serca Gdańskiego
Bigeminia komorowa jako przyczyna nieskuteczności terapii resynchronizującej i nieadekwatnych interwencji kardiowertera-defibrylatora
Elektrofizjologia inwazyjna/invasive electrophysiology Bigeminia komorowa jako przyczyna nieskuteczności terapii resynchronizującej i nieadekwatnych interwencji kardiowertera-defibrylatora Ventricular
Testy wysiłkowe w wadach serca
XX Konferencja Szkoleniowa i XVI Międzynarodowa Konferencja Wspólna SENiT oraz ISHNE 5-8 marca 2014 roku, Kościelisko Testy wysiłkowe w wadach serca Sławomira Borowicz-Bieńkowska Katedra Rehabilitacji
Dotyczy: postępowania o udzielenie zamówienia publicznego w trybie przetargu nieograniczonego na dostawę urządzeń do elektroterapii.
SAMODZIELNY PUBLICZNY SPECJALISTYCZNY SZPITAL ZACHODNI im. JANA PAWŁA II 05-825 Grodzisk Mazowiecki ul. Daleka 11 PN-EN ISO 9001:2009 AQAP 2120:2006 EIC ISO 27001:2005 CMJ HACCP PN-EN ISO 14001:2005 Dział
Wskazania do elektrostymulacji u chorych z omdleniami w świetle ostatnich wytycznych ESC. Piotr Kułakowski Klinika Kardiologii CMKP, Warszawa
Wskazania do elektrostymulacji u chorych z omdleniami w świetle ostatnich wytycznych ESC Piotr Kułakowski Klinika Kardiologii CMKP, Warszawa Deklaracja potencjalnego konfliktu interesów (w stosunku do
Nieoczekiwane zwolnienia rytmu w stymulatorach dwujamowych spowodowane obecnością arytmii nadkomorowych
PRACE KAZUISTYCZNE Nieoczekiwane zwolnienia rytmu w stymulatorach dwujamowych spowodowane obecnością arytmii nadkomorowych Krystian Stanek, Marek Prech, Przemysław Mitkowski, Marek Grygier, Franciszek
Różnorodne oblicza telemedycyny
Sympozjum Comarch e Zdrowie. Perspektywy jutra 2012 11 22, Warszawa Różnorodne oblicza telemedycyny Piotr Madej, Dyrektor Departamentu Rozwoju Systemów Telemedycznych, imed24 (Comarch Group) DROGA OD WIZJI
Wszyscy Wykonawcy. Grudziądz, dnia r. ZP / 16
Grudziądz, dnia 04.03.2016 r. Wszyscy Wykonawcy ZP - 453 / 16 Dotyczy przetargu nieograniczonego na dostawy stymulatorów oraz kardiowerterówdefibrylatorów serca wraz z wyposażeniem dla potrzeb Oddziału
Elektrostymulacja i defibrylacja serca. Elektroniczna aparatura medyczna 1 Wykład - 7
Elektrostymulacja i defibrylacja serca Elektroniczna aparatura medyczna 1 Wykład - 7 1 Z historii 1950 - John Hopps - pierwszy zewnętrzny rozrusznik serca Stymulatory w latach 50-tych były: ogromne zasilane
Cochlear Implants. Procedury medyczne. do systemów CI / ABI firmy MED EL. AW33315_6.0 (Polish)
Cochlear Implants Procedury medyczne do systemów CI / ABI firmy MED EL AW33315_6.0 (Polish) Niniejszy podręcznik zawiera ważne instrukcje i informacje dotyczące bezpieczeństwa skierowane do użytkowników
Postępowanie z chorym przed i po implantacji leczonym doustnymi lekami p-zakrzepowymi
Postępowanie z chorym przed i po implantacji leczonym doustnymi lekami p-zakrzepowymi Dr hab.n.med.barbara Małecka Krakowski Szpital Specjalistyczny im.jana Pawła II 1 1. Leczenie przeciwzakrzepowe wiąże
Zaburzenia stymulacji w codziennej praktyce pracowni 24 h monitorowania EKG ocena częstości, przykłady zapisów EKG
Folia Cardiol. 2001, tom 8, supl. A, A21 A29 Copyright 2001 Via Medica ISSN 1507 4145 Zaburzenia stymulacji w codziennej praktyce pracowni 24 h monitorowania EKG ocena częstości, przykłady zapisów EKG
Ocena stymulacji typu DDD u chorego z komorowymi zaburzeniami rytmu serca
PRACE KAZUISTYCZNE Ocena stymulacji typu DDD u chorego z komorowymi zaburzeniami rytmu serca Grażyna Kübler 1, Joanna Moszczyńska-Stulin 2 1 Poradnia Elektrostymulacji i Zaburzeń Rytmu Serca Szpitala Klinicznego
Nieadekwatne interwencje kardiowerteradefibrylatora.
Nieadekwatne interwencje kardiowerteradefibrylatora. Zapobieganie Dr n.med. Aleksander Maciąg Pracownia Elektrofizjologii Klinicznej II Kliniki Choroby Wieńcowej Instytutu Kardiologii w Warszawie 1 Deklaracja
Załącznik nr 2. Program zdrowotny Szkolenie elektrofizjologów inwazyjnych. Okres realizacji programu: 2008 rok.
Załącznik nr 2 Program zdrowotny Szkolenie elektrofizjologów inwazyjnych Okres realizacji programu: 2008 rok. Podstawa prawna realizacji programu Ustawa z dnia 27 sierpnia 2004 roku o świadczeniach opieki
Migotanie przedsionków Aleksandra Jarecka Studenckie Koło Naukowe przy Klinice Chorób Wewnętrznych i Kardiologii Szpitala Klinicznego Dzieciątka Jezus w Warszawie Kierownik Kliniki - prof. dr hab. Piotr
HRS 2014 LATE BREAKING
HRS 2014 LATE BREAKING DFT SIMPLE Michał Chudzik, Anna Nowek 1 Czy wyniki badania SIMPLE mogą wpłynąć na NIE wykonywanie rutynowego DFT? 2 Wyniki badnia SIMPLE pokazały, że wykonywanie DFT nie wpływa na
Rola monitorowania EKG przez telefon w diagnostyce zaburzeń czynności rozrusznika serca
Folia Cardiol. 2001, tom 8, supl. A, A35 A40 Copyright 2001 Via Medica ISSN 1507 4145 Rola monitorowania EKG przez telefon w diagnostyce zaburzeń czynności rozrusznika serca Dariusz Wojciechowski 1, Marek
Burza elektryczna - aktualne zasady postępowania Część 3 rokowanie, profilaktyka
Burza elektryczna - aktualne zasady postępowania Część 3 rokowanie, profilaktyka Maciej Kempa, Szymon Budrejko Klinika Kardiologii i Elektroterapii Serca, Gdański Uniwersytet Medyczny 1 Konsekwencje burzy
Rehabilitacja u chorych z wszczepionym stymulatorem i kardiowerterem - defibrylatorem II KATEDRA KARDIOLOGII CM UMK
Rehabilitacja u chorych z wszczepionym stymulatorem i kardiowerterem - defibrylatorem II KATEDRA KARDIOLOGII CM UMK 2014 Rehabilitacja pacjentów ze stymulatorem serca Prawidłowo działający stymulator serca
Unikanie nieadekwatnych wyładowań propozycje programowania ICD zgodnie z zaleceniami producentów dr n. med. Artur Oręziak
Unikanie nieadekwatnych wyładowań propozycje programowania ICD zgodnie z zaleceniami producentów dr n. med. Artur Oręziak KLINIKA ZABURZEŃ RYTMU SERCA, INSTYTUT KARDIOLOGII, WARSZAWA PARAMETRY DYSKRYMINACJI
Cochlear Implants. Procedury medyczne. dla systemów implantów MED EL. AW33315_1.0 (Polish)
Cochlear Implants Procedury medyczne dla systemów implantów MED EL AW33315_1.0 (Polish) Niniejsza instrukcja zawiera ważne wskazówki i informacje dotyczące bezpieczeństwa dla użytkowników systemu implantów
Znaczenie telemonitoringu u chorych z urządzeniami stymulującymi
Znaczenie telemonitoringu u chorych z urządzeniami stymulującymi Kraków, 11.12.2014 r. Adam Konka Prezes Zarządu Kardio-Med Silesia Park Technologii Medycznych KARDIO-MED SILESIA Nowoczesny ośrodek naukowo
Badania rezonansem magnetycznym u pacjentów z implantowanymi urządzeniami do elektroterapii serca
Kurs zgodny z konsensusem ekspertów HRS 2017 Kurs dla kardiologów, radiologów i specjalistów chorób wewnętrznych Dr n. med. Ewa Jędrzejczyk-Patej Badania rezonansem magnetycznym u pacjentów z implantowanymi
Układ bodźcoprzewodzący
ZABURZENIA RYTMU I PRZEWODZENIA II KATEDRA II KATEDRA KARDIOLOGII CM UMK 2014 2014 Układ bodźcoprzewodzący Węzeł zatokowo-przedsionkowy Węzeł przedsionkowo-komorowy Pęczek Hisa lewa i prawa odnoga Włókna
BADANIA INTERFERENCJI ELEKTROMAGNETYCZNYCH WYSTĘPUJĄCYCH W OBSZARZE KARDIOSTYMULATORA
II Konferencja Naukowa KNWS'05 "Informatyka- sztuka czy rzemios o" 15-18 czerwca 2005, Z otniki Luba skie BADANIA INTERFERENCJI ELEKTROMAGNETYCZNYCH WYSTĘPUJĄCYCH W OBSZARZE KARDIOSTYMULATORA Anna Pławiak-Mowna
Zadanie 1 ZESTAWIENIE PARAMETRÓW TECHNICZNYCH. nazwa handlowa,... producent...
Załącznik nr 2 do SIWZ Zadanie 1 ZESTAWIENIE PARAMETRÓW TECHNICZNYCH znak sprawy: Z/20/PN/15 Przedmiot zamówienia: Kardiowerter defibrylator serca jednojamowy wraz z wyposażeniem nazwa handlowa,... producent...
Interferencje elektromagnetyczne i inne praktyczne problemy u pacjenta ze stymulatorem serca
Praca poglądowa Interferencje elektromagnetyczne i inne praktyczne problemy u pacjenta ze stymulatorem serca Andrzej Kutarski, Piotr Ruciński Katedra i Klinika Kardiologii Akademii Medycznej w Lublinie
PAKIET II poz. 1. STYMULATORY SERCA AUTOMATYCZNE PROSTE JEDNOJAMOWE SSIR w komplecie z elektrodą Parametry wymagane pkt 1-21 i oceniane pkt 19-22
Sprawa nr 40/2016 Dostawa kardiowerterów i stymulatorów dla Szpitala Powiatowego w Chrzanowie AIT 2016 PAKIET II poz. 1. STYMULATORY SERCA AUTOMATYCZNE PROSTE JEDNOJAMOWE SSIR w komplecie z elektrodą Parametry
SURESCAN. Instrukcja techniczna
SURESCAN Informacje o procedurach związanych z badaniem metodą rezonansu magnetycznego urządzenie EnRhythm MRI SureScan EMDRO1 oraz elektrody SureScan warunkowo bezpieczne w badaniach MR Instrukcja techniczna
Współczesne sensory w monitorowaniu niewydolności serca
Współczesne sensory w monitorowaniu niewydolności serca Lek. Ewa Jędrzejczyk-Patej Katedra Kardiologii, Wrodzonych Wad Serca i Elektroterapii, Oddział Kliniczny Kardiologii, Śląskie Centrum Chorób Serca
Najczęstsze przyczyny nieadekwatnych interwencji kardiowertera-defibrylatora. Analiza zapisów wewnątrzsercowych
Najczęstsze przyczyny nieadekwatnych interwencji kardiowertera-defibrylatora. Analiza zapisów wewnątrzsercowych Dr n.med. Aleksander Maciąg Pracownia Elektrofizjologii Klinicznej II Kliniki Choroby Wieńcowej
RADOMSKI SZPITAL SPECJALISTYCZNY
RADOMSKI SZPITAL SPECJALISTYCZNY im. dr Tytusa Chałubińskiego RADOM ul. Lekarska 4 Dział Zamówień Publicznych, Funduszy Strukturalnych i Zaopatrzenia www.szpital.radom.pl; zampubl@rszs.regiony.pl NIP 796-00-12-187
Analiza czynności stymulatora w badaniu EKG metodą Holtera
ATYKUŁ POGLĄDOWY Folia Cardiol. 2002, tom 9, nr 3, 203 208 Copyright 2002 Via Medica IN 1507 4145 Analiza czynności stymulatora w badaniu EKG metodą Holtera omuald Ochotny, Hanna Wachowiak-Baszyńska i
Holter. odprowadzeń CM5, CS2, IS.
Norman Jefferis Jeff (1.1.1914-21.7.1983) amerykański biofizyk skonstruował urządzenie rejestrujące EKG przez 24 godziny, tzw. EKG. W zależności od typu aparatu sygnał EKG zapisywany jest z 2, 3, rzadziej
10. Rozrusznik serca. Tomasz Mroczek
10. Rozrusznik serca Tomasz Mroczek Prawidłowo funkcjonujące serce naprzemiennie kurczy się i rozkurcza, powodując tzw. bicie serca. Skurcz serca następuje w odpowiedzi na impuls elektryczny wywodzący
Pola elektromagnetyczne
Materiały szkoleniowe Krzysztof Gryz, Jolanta Karpowicz Pracownia Zagrożeń Elektromagnetycznych CIOP PIB, Warszawa krgry@ciop.pl, jokar@ciop.pl +22 623 46 50 1. Czym są pola elektromagnetyczne? tzw. fizyczny
Układy stymulujące Advisa MRI SureScan i Ensura MRI SureScan
Układy stymulujące Advisa MRI SureScan i Ensura MRI SureScan Jedno- i dwujamowe układy stymulujące Advisa MRI SureScan i Ensura MRI SureScan, warunkowo bezpieczne w badaniach MR, z elektrodami MRI SureScan
Terapia resynchronizująca u chorych z niewydolnością serca
PRACA POGLĄDOWA Folia Cardiologica Excerpta 2012, tom 7, nr 1, 41 45 Copyright 2012 Via Medica ISSN 1896 2475 Terapia resynchronizująca u chorych z niewydolnością serca Anna Hrynkiewicz-Szymańska 1, Marek
Wpływ pola elektromagnetycznego na { zdrowie }
Wpływ pola elektromagnetycznego na { zdrowie } Czym jest w ogóle promieniowane? Jest to zjawisko polegające na wysyłaniu i przekazywaniu energii na odległość. Energia ta może być wysyłana w postaci cząstek,
Co lekarz praktyk powinien wiedzieć o automatycznym kardiowerterze-defibrylatorze?
Choroby Serca i Naczyń 2014, tom 11, nr 3, 128 137 z a b u r z e n i a r y t m u s e r c a Redaktor działu: prof. dr hab. n. med. Zbigniew Kalarus Co lekarz praktyk powinien wiedzieć o automatycznym kardiowerterze-defibrylatorze?
PRĄDY WYSOKIEJ CZĘSTOTLIWOŚCI JOANNA GRABSKA -CHRZĄSTOWSKA
PRĄDY WYSOKIEJ CZĘSTOTLIWOŚCI JOANNA GRABSKA -CHRZĄSTOWSKA Drgania wysokiej częstotliwości Arsonwalizacja Arsonwalizacja HF08 aparat do arsonwalizacji Zastosowanie: Jest to urządzenie elektroniczne, działa
Badanie pilotażowe TCares 1; TCares 2
Badanie pilotażowe 1; 2 E-Zdrowie - opieka medyczna i niezależnośd Lek. med. Korneliusz Fil Celem nadrzędnym projektu jest zachęcenie do stosowania opieki telemedycznej oraz udoskonalenie działających
Podsumowanie danych o bezpieczeństwie stosowania produktu leczniczego Demezon
VI.2 VI.2.1 Podsumowanie danych o bezpieczeństwie stosowania produktu leczniczego Demezon Omówienie rozpowszechnienia choroby Deksametazonu sodu fosforan w postaci roztworu do wstrzykiwań stosowany jest
Pakiet nr 1 Cena zamówienia brutto. Słownie: w tym podatek VAT wynosi Cena zamówienia netto. słownie:.. ZGODNE Z PONIŻSZYM FORMULARZEM CENOWYM
Pakiet nr 1 słownie:. 1 Stymulator dwujamowy o przedłużonej żywotności 40,00 1.A Elektrody zgodne z zapisami w załączniku nr 4 do SIWZ 80,00 Pakiet nr 2 słownie:. 1 Stymulator jednojamowy o przedłużonej
Retrospektywna analiza przyczyn niepowodzeń wdrożenia stymulacji typu DDD u pacjentów operowanych w latach
PRACA BADAWCZA PRZEDRUK Folia Cardiologica Excerpta 2007, tom 2, nr 6, 230 235 Copyright 2007 Via Medica ISSN 1896 2475 Retrospektywna analiza przyczyn niepowodzeń wdrożenia stymulacji typu DDD u pacjentów
Anesthetic management of patients with pacemakers or implantable defibrillators
A R T Y K U Ł P O G L Ą D O W Y Wpłynęło: 13.03.2008 Poprawiono: 20.03.2008 Zaakceptowano: 20.03.2008 Postępowanie okołooperacyjne u pacjentów z rozrusznikami serca i wewnętrznymi defibrylatorami/kardiowerterami
BEZPRZEWODOWE WYJŚCIE CYFROWE (2-KANAŁOWE, KOMPAKTOWE) AS70DOC002
BEZPRZEWODOWE WYJŚCIE CYFROWE (2-KANAŁOWE, KOMPAKTOWE) INSTRUKCJA OBSŁUGI ASTOR SP. Z O.O. ul. Smoleńsk 29 31-112 Kraków tel. 12 428 63 00 info@comodis.pl comodis.pl DO CZEGO SŁUŻY? Bezprzewodowe wyjścia
Ocena czynności stymulatora serca w badaniu EKG metodą Holtera
Folia Cardiol. 2001, tom 8, supl. A, A11 A15 Copyright 2001 Via Medica ISSN 1507 4145 Ocena czynności stymulatora serca w badaniu EKG metodą Holtera Romuald Ochotny I Klinika Kardiologii Instytutu Kardiologii
MONITOROWANIE EKG, ZABURZENIA RYTMU SERCA RC (UK)
MONITOROWANIE EKG, ZABURZENIA RYTMU SERCA Zagadnienia Wskazania i techniki monitorowania elektrokardiogramu Podstawy elektrokardiografii Interpretacja elektrokardiogramu formy NZK groźne dla życia zaburzenia
Programowanie mikrokontrolerów 2.0
13.1 Programowanie mikrokontrolerów 2.0 Sterowanie fazowe Marcin Engel Marcin Peczarski Instytut Informatyki Uniwersytetu Warszawskiego 19 grudnia 2016 Triak Triak jest półprzewodnikowym elementem przełączającym
Promieniowanie elektromagnetyczne w środowisku pracy. Ocena możliwości wykonywania pracy w warunkach oddziaływania pól elektromagnetycznych
Promieniowanie elektromagnetyczne w środowisku pracy Ocena możliwości wykonywania pracy w warunkach oddziaływania pól elektromagnetycznych Charakterystyka zjawiska Promieniowanie elektromagnetyczne jest
Test ze wzrastającą energią impulsu w oznaczaniu progu defibrylacji migotania komór u pacjentów z kardiowerterem-defibrylatorem
PRACA ORYGINALNA Folia Cardiol. 2005, tom 12, nr 2, 111 116 Copyright 2005 Via Medica ISSN 1507 4145 Test ze wzrastającą energią impulsu w oznaczaniu progu defibrylacji migotania komór u pacjentów z kardiowerterem-defibrylatorem
Jarosław Blicharz, Oddział Kardiologii, Szpital Wojewódzki im. Św. Łukasza SP ZOZ w Tarnowie
Wszczepialne stymulatory i defibrylatory serca dopuszczone do badania z użyciem rezonansu magnetycznego i warunki wykonywania badania (stan na 15.07.2014) Jarosław Blicharz, Oddział Kardiologii, Szpital
Zamość, dnia 23 października 2018 r. AG.ZP Dotyczy: Wyjaśnienia, zmiana treści SIWZ
Zamość, dnia 23 października 2018 r. AG.ZP.3320.62..18 Dotyczy: Wyjaśnienia, zmiana treści SIWZ Samodzielny Publiczny Szpital Wojewódzki im. Papieża Jana Pawła II w Zamościu ul. Aleje Jana Pawła II 10,
Zamawiający Otwock, dnia 7 grudnia 2018r. Samodzielny Publiczny Szpital Kliniczny im. prof. Adama Grucy CMKP ul. Konarskiego Otwock
Zamawiający Otwock, dnia 7 grudnia 2018r. Samodzielny Publiczny Szpital Kliniczny im. prof. Adama Grucy CMKP ul. Konarskiego 13 05-400 Otwock Do Wykonawców Dotyczy: Dostawa 8 aparatów do ogrzewania pacjenta
ROZPORZĄDZENIE MINISTRA ZDROWIA 1) z dnia 2 lutego 2011 r.
Dziennik Ustaw Nr 33 2534 Poz. 167 167 ROZPORZĄDZENIE MINISTRA ZDROWIA 1) z dnia 2 lutego 2011 r. w sprawie kryteriów raportowania zdarzeń z wyrobami, sposobu zgłaszania incydentów medycznych i działań
INSTRUKCJA OBSŁUGI FRESH INTELLIVENT. Nowe funkcje 2011. 1 Ciągła praca z ustawianą prędkością i zachowaną funkcją timera.
INSTRUKCJA OBSŁUGI FRESH INTELLIVENT PL Nowe funkcje 2011 1 Ciągła praca z ustawianą prędkością i zachowaną funkcją timera. 2 Funkcja pauzy realizowana za pomocą włącznika chwilowego. 3 Wybór jednego z
ZAKRES BADAŃ BEZPIECZEŃSTWO UŻYTKOWANIA I EMC CELAMED Centralne Laboratorium Aparatury Medycznej Aspel S.A.
Przedstawiony formularz umożliwia wybór badań będących przedmiotem zamówienia, sporządzenia planu badań. Dla ułatwienia wyboru przedstawiono krótką charakterystykę techniczną możliwości badawczych, oraz
Załącznik nr 1 do SIWZ. Pakiet nr 1. Wysokoenergetyczny kardiowerter-defibrylator resynchronizujący /CRT-D/ z kompletem elektrod - 20 Szt.
Załącznik nr 1 do SIWZ Pakiet nr 1 Wysokoenergetyczny kardiowerter-defibrylator resynchronizujący /CRT-D/ z kompletem elektrod - 20 Szt. Rok produkcji nie wcześniej niŝ 2011 Lp Parametry graniczne Spełnienie
Czasowy zewnętrzny stymulator dwujamowy
5392 Czasowy zewnętrzny stymulator dwujamowy Instrukcja techniczna 0123 2013 Poniższa lista zawiera znaki towarowe lub zastrzeżone znaki towarowe firmy Medtronic w Stanach Zjednoczonych i prawdopodobnie
Choroby Serca i Naczyń 2012, tom 9, nr 1, E K G W P R A K T Y C E
Choroby Serca i Naczyń 2012, tom 9, nr 1, 50 54 E K G W P R A K T Y C E Redaktor działu: dr hab. n. med. Rafał Baranowski Kaszel najtańszy i łatwo dostępny lek antyarytmiczny, a czasem lek mogący uratować
Wyłącznik czasowy GAO EMT757
INSTRUKCJA OBSŁUGI Wyłącznik czasowy GAO EMT757 Produkt nr 552451 Instrukcja obsługi Strona 1 z 10 Cyfrowy programator czasowy Artykuł nr: EMT757 A. Funkcje 1. Cyfrowy programator czasowy (zwany dalej
Wytyczne resuscytacji 2010 Defibrylacja
Wytyczne resuscytacji 2010 Defibrylacja Należy minimalizowad przerwy w wysokiej jakości uciśnięciach klatki piersiowej prowadzonych podczas wszystkich interwencji ALS (w tym także defibrylacji). Znacznie
UWAGA! ELEKTRYCZNE POD NAPIĘCIEM!
tech -1- ST-360 UWAGA! URZĄDZENIE ELEKTRYCZNE POD NAPIĘCIEM! Przed dokonaniem jakichkolwiek czynności związanych z zasilaniem (podłączanie przewodów, instalacja urządzenia, itp.) należy upewnić się, że
Burza elektryczna - aktualne zasady postępowania
Burza elektryczna - aktualne zasady postępowania Część 1 definicja, występowanie, przyczyny, rozpoznanie Maciej Kempa, Szymon Budrejko Klinika Kardiologii i Elektroterapii Serca, Gdański Uniwersytet Medyczny
Aneks IV. Wnioski naukowe
Aneks IV Wnioski naukowe 1 Wnioski naukowe Od czasu dopuszczenia produktu Esmya do obrotu zgłoszono cztery przypadki poważnego uszkodzenia wątroby prowadzącego do transplantacji wątroby. Ponadto zgłoszono
Pacjent SOR w aspekcie ochrony radiologicznej - kobiety w ciąży. dr Piotr Pankowski
Pacjent SOR w aspekcie ochrony radiologicznej - kobiety w ciąży dr Piotr Pankowski Ustawa z dnia 29 listopada 2000 r. Prawo atomowe (Dz. U. 2012.264) Art. 33c 6. Dzieci, kobiety w wieku rozrodczym, kobiety
Skale i wskaźniki jakości leczenia w OIT
Skale i wskaźniki jakości leczenia w OIT Katarzyna Rutkowska Szpital Kliniczny Nr 1 w Zabrzu Wyniki leczenia (clinical outcome) śmiertelność (survival) sprawność funkcjonowania (functional outcome) jakość