BADANIA KLINICZNE. Charakterystyka pięciu różnych powierzchni implantów i ich wpływu na orientację: badanie na psach. Informacja wprowadzająca

Wielkość: px
Rozpocząć pokaz od strony:

Download "BADANIA KLINICZNE. Charakterystyka pięciu różnych powierzchni implantów i ich wpływu na orientację: badanie na psach. Informacja wprowadzająca"

Transkrypt

1 BADANIA KLINICZNE Paulo Coelho ESTEVAM BONFANTE Charakterystyka pięciu różnych powierzchni implantów i ich wpływu na orientację: badanie na psach Characterization of 5 different implant surfaces and their effect on osseointegration: study in dogs. Paulo G. Coelho PhD1, Estevam A. Bonfante PhD2, Roberto S. Pessoa PhD3, Charles Marin PhD4, Rodrigo Granato PhD4, Gabriela Giro PhD3, Łukasz Witek MS1, Marcelo Suzuki DDS5 1 Department of Biomaterials and Biomimetics, New York University, New York, NY, USA 2 Department of Prosthodontics, University of Säo Paulo - Bauru School of Dentistry, Bauru, SP, Brazylia 3 Department of Oral Diagnosis and Surgery, Universidade Estadual Paulista Jülio de Mesquita Filho, Faculdade de Odontologia de Araraquara, Araraquara, SP, Brazylia 4 Department of Dentistry, Universidade Federal de Santa Catarina, Florianópolis, SC, Brazylia 5 Department of Prosthodontics, Tufts University School of Dental Medicine, Boston, MA, USA Estevam A. Bonfante, AI. Octavia Pinheiro Brisola, Bauru - SP - Brasil , Phone: , estevamab@gmail.com Słowa kluczowe: powierzchnia implantu, in vivo, moment obrotowy, histologia, osteointegracja Key words: implant surface, in vivo, torque, histology, osseointegration Informacja wprowadzająca Chemiczna modyfikacja powierzchni implantów jest kojarzona ze zmianami w zakresie topografii powierzchni, które mogą mieć wpływ na wczesny proces osteointegracji. Celem niniejszego badania było określenie na modelu zwierzęcym wpływu kontrolowanych zmian w obrębie powierzchni implantu na wczesny proces osteointegracji. Metodologia Ocenie poddano pięć rodzajów powierzchni implantów: piaskowaną tlenkiem glinu (Alumina-Blasted), piaskowaną biologicznie (Biological Blasting), pokrytą plazmą (Plasma), mikropiaskowaną resorbowalnym środkiem (Microblasted-resorbable blasting media - microblasted RBM) oraz piaskowaną tlenkiem glinu i wytrawianą (Alumina-Blasted/Acid-Etched - AB/ AE). Topografię powierzchni scharakteryzowano za pomocą skaningowej mikroskopii elektronowej, optycznej interferometrii, a skład chemiczny poddano analizie wykorzystując fotoelektronową spektroskopię rentgenowską (XPS). Implanty wprowadzono w kość promieniową sześciu psów i pozostawiono na 2 lub 4 tygodnie in vivo. Po przeprowadzeniu eutanazji badan implanty były wykręcane momentem obrotowym powodującym zerwanie połączenia z kością, a następnie były poddawane w formie nieodwapnionej badaniom histomorfologicznym oraz oceniano kontakt kości z implantem (BIC - z ang. bone-to-implant contact) i części powierzchni zajęte przez kość (BAFO z ang. bone area fraction occupied). Analizę statystyczną przeprowadzono za pomocą jednokierunkowego testu ANOVA (p<0,05), oraz testów post-hoc Tukeya. Streszczenie Abstract 2

2 BADANIA KLINICZNE Wyniki Dla powierzchni piaskowanej tlenkiem glinu wykazano najwyższe wartości średniej szorstkości powierzchniowej (Sa) i średniej wartości pierwiastka kwadratowego z pola powierzchni (Sq), podczas gdy parametry te miały najniższe wartości dla powierzchni piaskowanej biologicznie, a średnie dla powierzchni piaskowanej i wytrawianej (AB/AE). Powierzchnie pozostałe prezentowały wartości pośrednie pomiędzy zanotowanymi dla powierzchni piaskowanej biologicznie i piaskowanej/wytrawianej. Widma XPS wykazały obecność Ca i P na powierzchniach piaskowanych biologicznie oraz mikropiaskowanych, a najwyższy poziom O zanotowano dla powierzchni pokrytej plazmą, mikropiaskowanej i piaskowanej/wytrawianej. Zaobserwowano istotnie wyższe wartości momentu obrotowego po okresie 2 tygodni dla powierzchni mikropiaskowanej RBM (p<0,04), ale po okresie 4 tygodni nie stwierdzono różnic pomiędzy badanymi powierzchniami (p>0,74). Nie zaobserwowano istotnych różnic w wartościach parametrów BIC i BAFO po upływie 2 i 4 tygodni. Wnioski Wszystkie pięć badanych rodzajów powierzchni posiada właściwości osteokondukcyjne i zapewnia wysoki stopień osteointegracji i zamocowania biomechanicznego. Na przestrzeni ostatnich czterech dziesięcioleci wykazano wysoki odsetek przetrwania implantów (często przekraczający 95% po upływie 10 lat [1,2], dzięki czemu kliniczna implantologia stomatologiczna jest uważana za jedną z najpewniejszych metod leczenia stomatologicznego. Ten wysoki odsetek powodzenia klinicznego przypisywany jest doskonałej biokompatybilności tytanu, która umożliwia zaistnienie ścisłej interakcji powierzchni implantu z kością widocznej w powiększeniu mikroskopu optycznego, nazywanej osteointegracją [3-5]. Chociaż wszczepy śródkostne cechują się wysokim odsetkiem przetrwania w wymiarze długoczasowym, to wyniki ostatnich badań wskazują na znaczenie modyfikacji ich kształtu na różnym poziomie (tzn. w zakresie nano-, mikro- i makrogeometrii) tak, aby poprawić wczesną odpowiedź tkanek gospodarza na implant. Takie potencjalne skrócenie czasu gojenia może skutkować zawężeniem ram czasowych leczenia implantologicznego, tak że okluzyjnie funkcjonalne uzupełnienia protetyczne mogłyby być zakładane na wczesnych etapach po implantacji [6-8]. Spośród możliwych modyfikacji konstrukcji implantów, które mają za zadanie poprawę odpowiedzi tkanek gospodarza na implant, najwięcej uwagi w badaniach naukowych poświęca się modyfikacji ich powierzchni [3-8]. Podstawą do działań zmierzających do modyfikacji w powierzchni implantów jest fakt, że powierzchnia implantu najwcześniej wchodzi w interakcję z płynami ustrojowymi, wpływając potencjalnie na kaskadę procesów, które doprowadzają do gojenia kości oraz jej apozycji w najbliższym otoczeniu implantu [9]. W wielu opublikowanych do tej pory artykułach o charakterze przeglądu piśmiennictwa podejmowano tematykę rozmaitych możliwości modyfikacji w zakresie powierzchni implantów i panuje ogólne przekonanie, że zarówno szorstkość powierzchni (w porównaniu do powierzchni gładkich), jak i ich chemiczna modyfikacja (dodatek bioceramiki na bazie Ca-P w różnych postaciach w porównaniu do powierzchni nieuzdatnionych) sprzyjają zaistnieniu wczesnej odpowiedzi tkanek gospodarza na implant. Z historycznego punktu widzenia, powierzchnie implantów dentystycznych ewoluowały od zupełnie gładkich do powierzchni o szorstkiej teksturze, a wyniki najnowszych badań skłaniają się ku chemicznej modyfikacji powierzchni umiarkowanie szorstkich [4-6,10,11]. Na przestrzeni czasu zainteresowanie budziły zarówno topograficzne, jak i chemiczne modyfikacje powierzchni implantów [4-6,10,11], ponieważ zarówno jedne, jak i drugie, dawały obiecujące wyniki w badaniach in vitro oraz in vivo [13-20] w porównaniu do swoich poprzedników o umiarkowanie szorstkich powierzchniach. Poprawę właściwości powierzchni uzyskiwano poprzez zmiany zwilżalności powierzchni [21], impregnację Ca i P warstwy tlenku tytanu [18], depozycję bioaktywnej ceramiki [19,20, 22], oraz poprzez inkorporację innych cząstek chemicznych, takich jak na przykład fluorki [11, 23]. Ponieważ chemiczne modyfikacje powierzchni skutkują typowo zmianami ich tekstury powstającymi w procesie obróbki, to kontrolowanie takich zmiennych, tak by osiągnąć jak najlepsze wyniki procesu gojenia, jest zadaniem trudnym, w związku z czym pożądane jest obszerne, bezpośrednie porównanie różnych rodzajów powierzchni na odpowiednim modelu in vivo. Dlatego też w niniejszym badaniu przeprowadzono porównanie biomechaniczne i histomorfometryczne efektów różnych modyfikacji powierzchni na modelu zwierzęcym. MATERIAŁY I METODOLOGIA W badaniu wykorzystano implanty śrubowe Ti-6Al-4V o średnicy 3,75 mm i długości 8 mm, dostarczone przez producenta (AB Dental). W sumie zastosowano 75 implantów, które zostały podzielona na pięć grup według sposobu modyfikacji powierzchni: piaskowane tlenkiem glinu (Alumina-Blasted), piaskowane biologicznie (Biological Blasting), pokryte plazmą (Plasma), mikropiaskowane (Micorblasted resorbable blasting media - microblasted RBM) oraz piaskowane/wytrwiane (Alumina-Blasted/ Acid-Etched - AB/AE). Trzy implanty z każdej grupy wykorzystano do charakterystyka powierzchni. Charakterystyka powierzchni Powierzchnie implantów scharakteryzowano przy pomocy trzech różnych metod (n=3 implanty o danym typie modyfikacji powierzchni). W pierwszej kolejności przeprowadzono badanie za pomocą elektronowej mikroskopii skaningowej (Phillips XL 30) (SEM) w różnych powiększeniach przy napięciu przyspieszającym 15kV, by zaobserwować topografię danych rodzajów powierzchni implantów. Kolejnym etapem było określenie parametrów szorstkości z wykorzystaniem interferometrii optycznej (Phase View 2.5 ) (IFM). Trzy implanty, po jednym z każdego rodzaju powierzchni, poddano ocenie w płaskim obszarze ich krawędzi tnących (trzy pomiary na każdy implant). Wyznaczono parametr Sa (średnia arytmetyczna wartości całkowitych wysokości powierzchni w badanym obszarze) oraz Sq (średni pierwiastek kwadratowy z wartości całkowitych różnic wysokości powierzchni w badanym obszarze) [24,25]. Aby oddzielić szorstkość od pofalowania powierzchni dla cyfrowych pomiarów 3D na skali mikrometrowej, zastosowano wysokoprzepustowy filtr Gaussa o wymiarach 250µm x 3

3 250µm. Po wykonaniu analizy zgodności z rozkładem normalnym, przeprowadzono analizę statystyczną za pomocą jednokierunkowego testu ANOVA, przyjmując poziom istotności 95%. Trzecią procedurą była specyficzna ocena składu chemicznego powierzchni przeprowadzona za pomocą fotoelektronowej spektroskopii rentgenowskiej (XPS). Implanty umieszczono w próżniowej komorze transferowej i odgazowano do 10-7 torra. Następnie przetransportowano próbki w warunkach próżni do spektrofotometru XPS. Widma XPS uzyskano przy pomocy koncentrycznego półsferycznego analizatora o średnim promieniu 165 mm, pracującym przy ciągłej wartości energii 160 ev do badania i 80 ev do skanów o wysokiej rozdzielczości. Kąt odbicia wynosił 90o, a wymiary badanego pola 150 µm x 150 µm. Powierzchnie implantów poddano ocenie w różnych miejscach (3 miejsca badania/implant). Model zwierzęcy zabieg chirurgiczny Po otrzymaniu zgody Komisji Bioetycznej ds. Badań na Zwierzętach Uniwersytetu Federalnego Santa Catarina pozyskano sześć psów mieszańców, które były następnie trzymane przez 2 tygodnie w zwierzętarni, zanim przystąpiono do pierwszej procedury chirurgicznej. Bezpośrednio przed zabiegiem, zanim uzyskano znieczulenie ogólne, podano trzy leki drogą iniekcji domięśniowej. Lekami tymi były: siarczan atropiny (0,044 mg/kg), chloran xylazyny (8 mg/kg) i chloran ketaminy (15 mg/kg). Miejscem implantacji była nasada kości promieniowej kończyny prawej, jeśli implanty wprowadzano na 4 tygodnie in vivo, lub lewej, jeśli implanty wprowadzano na 2 tygodnie in vivo. W celu wprowadzenia implantów u każdego zwierzęcia miejsce operacyjne zostało ogolone żyletką oraz odkażone roztworem jodyny. Wykonano cięcie skóry i okostnej na długości około 5 cm, po czym odpreparowano okostną, aby odsłonić kość. Implanty wkręcano ruchem ciągłym z prędkością obr./min, przy obfitym chłodzeniu roztworem soli fizjologicznej, zgodnie z zaleceniami producenta. Wszczepy zostały wprowadzone w takim rozmieszczeniu, by zminimalizować ich ukośne ustawienie w miejscach implantacji (miejsca 1 do 5 od okolicy proksymalnej do dystalnej) wzdłuż nasady kości promieniowej, tak by możliwa była ocena momentu obrotowego oraz analiza histomorfometryczna. Po wprowadzeniu implantów założono śruby gojące, a tkanki zeszyto warstwowo: okostną szwami wikrylowymi 4-0, a skórę nylonowymi 4-0. Zwierzęta pozostały w zwierzętarni; celem kontroli bólu i infekcji wprowadzono antybiotykoterapię (penicylina, jedn/kg) oraz leki przeciwzapalne (ketoprofen 1% 1ml/5kg). Po czterech tygodniach przeprowadzono eutanazję poprzez przedawkowanie leków znieczulających/usypiających, a kończyny oddzielono wykonując ostre cięcie. Celem oceny momentu obrotowego podczas wykręcania zamocowano kość promieniową w urządzeniu służącym do elektronicznego pomiaru momentu obrotowego wyposażonym w komórkę mierzącą obciążenie do 500 Ncm3. Zastosowano narzędzia indywidualnie dopasowane do wewnętrznego połączenia każdego implantu, a blok kostny umieszczono ostrożnie, tak by uniknąć przemieszczenia próbki w trakcie testu. Implanty były wykręcane w kierunku przeciwnym do ruchu wskazówek zegara z prędkością ~0,196 radianów/min, dopóki nie zanotowano spadku wartości maksymalnej o 10%, i dla każdej próbki wyznaczono krzywą zależności przemieszczenia od momentu obrotowego. Uzasadnieniem takiego postępowania była konieczność zminimalizowania ewentualnych uszkodzeń w obszarze interfejsu (poł. kości z implantem) przed przystąpieniem do badań histologicznych [13,14]. Następnie przesłano implanty pozostające w kości do analizy histomorfometrycznej. Implanty w kości zostały zredukowane do małych bloczków, po czym zanurzono je na 24 godziny w 10% roztworze zbuforowanej formaliny. Bloczki poddano następnie płukaniu w bieżącej wodzie przez kolejne 24 godziny i stopniowemu odwadnianiu w roztworach alkoholu etylowego o wzrastającym stężeniu od 70 do 100%. Po odwodnieniu bloczki zostały zatopione w żywicy na bazie metakrylanu 1 zgodnie z instrukcją producenta. Bloczki pocięto następnie na cienkie warstwy (o grubości ~300 µm) za pomocą precyzyjnej piły z nasypem diamentowymɫ, wzdłuż osi długiej implantu, w kierunku zorientowanym na jego środek. Przekroje te zostały następnie przyklejone do płytek akrylowych cementem na bazie akrylanu, a po upływie 24 godzin wymaganych do związania cementu, przystąpiono do szlifowania i polerowania. Przekrojom nadano ostateczną grubość wynoszącą ~30 µm poprzez abrazję papierem ściernym3 (o wzrastającej ziarnistości nasypu: 400, 600, 800, 1200, 2400) na szlifierce, stosując chłodzenie wodne [26]. Przekroje wybarwiono następnie błękitem toluidynowym i przesłano do badania pod mikroskopem optycznym celem oceny histomorfologicznej. Kontakt kość implant (bone-to-implant contact, BIC) oceniono w powiększeniu 50X 200X przy pomocy oprogramowania komputerowego 5. Obszary kontaktu kości z implantem wzdłuż obwodu wszczepu zostały odjęte od wartości całkowitego obwodu wszczepu, po czym obliczono wartość parametru BIC. Część powierzchni zajętej przez kość (bone area fraction occupied - BAFO) pomiędzy gwintami wszczepu a beleczkami kości gąbczastej oceniono w powiększeniu 100X przy pomocy oprogramowania komputerowego. Obszary wszczepu zajęte przez kość odjęto od całkowitego pola pomiędzy gwintami, po czym wyznaczono wartość parametru BAFO (wyrażonej w procentach części powierzchni zajętej przez kość) [27]. Wstępna analiza statystyczna nie wykazała różnic wynikających z miejsca implantacji (pozycje od 1 do 5 wzdłuż kości promieniowej) na żaden z badanych parametrów. Dlatego też pominięto ten parametr w dalszej analizie statystycznej. Do oceny zależności statystycznej pomiędzy momentem obrotowym a zerwaniem połączenia z kością (interfejsu), BIC oraz BAFO wykorzystano jednokierunkowy test ANOVA. Za istotne statystycznie przyjęto wyniki o wartości p mniejszej od 5%, a analizę post hoc przeprowadzono za pomocą testu LSD Fischera. WYNIKI Wszystkie obrazy powierzchni implantów z mikroskopu elektronowego są przedstawione na Rycinach 1 i 2, a reprezentacyjne trójwymiarowe rekonstrukcje IFM pól o wymiarach 250 µm x 250 µm na Rycinie 3. Odpowiednie wartości parametrów Sa i Sq są przedstawione na Rycinie 4A. Na poziomie średniego oraz dużego powiększenia, jak również na rekonstrukcjach IFM (Rycina 3) zaobserwowano różnice morfologiczne w teksturze powier- 4

4 zchni implantów pomiędzy badanymi grupami. Podczas gdy podobną morfologię zaobserwowano dla powierzchni piaskowanej tlenkiem glinu, mikropiaskowanej oraz piaskowanej/wytrawianej, to obrazy elektronowej mikroskopii skaningowej wykazały, że powierzchnie piaskowane biologicznie posiadają rejony szorstkie powstałe w wyniku procedury piaskowania, wraz z rejonami gładkimi z rowkami powstałymi podczas procesu produkcji. Powierzchnie plazmowe prezentowały morfologię szorstką, ale, w porównaniu do powierzchni innych grup, o zaokrąglonym zarysie (Rycina 1 E i F). Resztkowe cząsteczki środków piaskujących były widoczne jedynie na powierzchniach piaskowanych tlenkiem glinu oraz piaskowanych biologicznie (Rycina 1 C i D). Wykazano istotne statystycznie różnice w wartościach parametrów Sa i Sq w pomiarach metodą IFM (Rycina 4 A i B), gdzie najwyższą wartość zanotowano dla powierzchni piaskowanej tlenkiem glinu, najniższą dla piaskowanej biologicznie, a pośrednią dla piaskowanej/wytrawianej. Pozostałe typy powierzchni prezentowały wartości pośrednie pomiędzy wartościami powierzchni piaskowanych biologicznie i piaskowanych/wytrawianych (różnice nieistotne statystycznie pomiędzy grupami) (Ryciny 4 A i B). Za pomocą widm XPS oceniono obecność pierwiastków Al, P, Ca, N, Ti, C, V, O oraz Na we wszczepach o badanych rodzajach powierzchni (Ryciny 4A i C). Najwyższą zawartość glinu (Al) zaobserwowano dla powierzchni piaskowanej tlenkiem glinu. Najwyższą zawartość wapnia (Ca) i fosforu (P) zaobserwowano dla powierzchni piaskowanej biologicznie, następnie dla powierzchni mikropiaskowanej zawartość tych pierwiastków była już znacznie niższa, natomiast dla pozostałych badanych powierzchni nie wykazano obecności tych pierwiastków. Nie wykryto tytanu (Ti) dla powierzchni pokrytej plazmą, a drugą najniższą zawartość Ti wykazano dla powierzchni piaskowanej biologicznie. Największą zawartością węgla (C) charakteryzowały się powierzchnie pokryta plazmą oraz piaskowana tlenkiem glinu, a najwyższą zawartość tlenu (O) wykazano dla powierzchni pokrytej plazmą, mikropiaskowanej oraz piaskowanej/wytrwianej (Rycina 4A i C). W czasie zabiegu chirurgicznego oraz w okresie obserwacji pozabiegowej nie zaobserwowano żadnych powikłań związanych z samą procedurą, infekcją pozabiegową czy jakichkolwiek innych. Z badania nie wykluczono żadnego implantu z powodu niestabilności klinicznej w czasie bezpośrednio po eutanazji. Wyniki testów biomechanicznych wykazały, że wartość zerwania interfejsu pod wpływem działania momentu obrotowego była istotnie statystycznie wyższa dla implantów o powierzchni micropiaskowanej w porównaniu do innych testowanych powierzchni w okresie pierwszych 2 tygodni (p<0,04), ale dla okresu 4-tygodniowego nie wykazano różnic pomiędzy implantami o różnych typach powierzchni (p>0,74) (Rycina 5 A i B). Badania na nieodwapnionych próbkach wykonane po kontrolowanych testach działania momentem obrotowym wykazały ścisły kontakt kości z badanymi rodzajami powierzchni implantów w obrębie kości zbitej oraz beleczkowej. Obserwacja okolicy połączenia kości z powierzchnią implantu w większym powiększeniu wykazała, że przekroje nieodwapnione, które zostały poddane opisanym testom biomechanicznym, wykazały minimalne odkształcenie w zakresie morfologii przez rozerwanie kości w wyniku testów mechanicznych (Rycina 6). Ocena jakościowa cienkich skrawków wybarwionych błękitem toluidyny nie wykazała występowania różnic morfologicznych in vivo pomiędzy badanymi powierzchniami po upływie zarówno 2 tygodni (Rycina 6 A, B i C) jak i 4 tygodni (Rycina 6 D, E i F) po upływie obu czasów obserwacji widoczny był ścisły kontakt pomiędzy wszczepem a kością korową (Rycina 6 A, B, D i E) oraz beleczkową (Rycina 6 C i F). Dodatkowo, zaobserwowano różnorodne wzorce gojenia wzdłuż wszczepu, zależne od interakcji pomiędzy geometrią implantu a wymiarami zastosowanych instrumentów chirurgicznych. W okolicy implantu, gdzie wewnętrzna średnica gwintu była większa lub równa wymiarowi ostatniego użytego wiertła chirurgicznego, umożliwiając tym samym powstanie ścisłego kontaktu pomiędzy powierzchnią implantu a kością korową już bezpośrednio po implantacji, pomiędzy 2-gim (Rycina 6 A) i 4-tym (Rycina 6 D) tygodniem doszło in vivo do wyraźnego remodelingu kości otaczającej wszczep dla wszystkich badanych powierzchni. Podczas gdy po okresie 2 tygodni można było zaobserwować in vivo proces przebudowy starej kości wraz z pojawianiem się nowego utkania kostnego (Rycina 6 A), to po upływie 4 tygodni nowe utkanie kostne było wyraźnie widoczne w rejonie kości przylegającej bezpośrednio do powierzchni implantu (Rycina 6 D). W miejscach, gdzie powstała komora gojąca poprzez wypreparowanie przestrzeni pomiędzy zewnętrznym wymiarem instrumentu chirurgicznego, a wewnętrznym wymiarem nagwintowania, proces tworzenia nowego utkania kostnego zachodził na całej przestrzeni komory gojącej oraz bezpośrednio na powierzchni implantu już w okresie 2 tygodniowej obserwacji in vivo (Rycina 6B). Po upływie 4 tygodni, na całej przestrzeni komory gojącej utkanie nowej kości uległo organizacji w blaszki kostne (Rycina 6E). W obszarach kości gąbczastej obserwowano formowanie się nowego utkania kostnego po okresie 2 tygodni (Rycina 6 C), a proces organizacji tego utkania w blaszki kostne był widoczny po okresie 4 tygodni (Rycina 6 F) w otoczeniu wszczepów o wszystkich badanych rodzajach powierzchni. Wyniki badań histomorfometrycznych nie wskazały występowania istotnych statystycznie różnic pomiędzy badanymi rodzajami powierzchni dla parametrów BIC oraz BAFO zarówno po 2, jak i po 4 tygodniach in vivo (BIC odpowiednio p>0,26 i p>0,09; BAFO odpowiednio p>0,94 i p>0,09 Ryciny 5 C, D, E i F). DYSKUSJA Powierzchnie implantów ewoluowały od zupełnie gładkich w kierunku uważanych dziś za standard powierzchni szorstkich, wytwarzanych różnymi metodami, wśród których zawierają się również metody, które zostały wykorzystane do modyfikacji powierzchni implantów podlegających ocenie w niniejszym badaniu [6]. W porównaniu do wyników innych badań, uzyskane przez nas wartości szorstkości powierzchni były wyższe przez zastosowanie filtra o większym rozmiarze do pomiarów IFM (im mniejszy filtr, tym niższe wartości Sa oraz Sq [28]). Niemniej jednak względne różnice pomiędzy grupami są zgodne z wynikami wcześniej publikowanych badań [4,5]. Podczas gdy powierzchnia piaskowana tlenkiem glinu cechowała się szorstką teksturą, a zarazem zatopionymi w niej cząsteczkami środków zastosowanych do piaskow- 5

5 ania, to powierzchnia piaskowana/wytrawiana była umiarkowanie szorstka oraz pozbawiona cząsteczek piaskujących, wskazując na skuteczność wytrawiania kwasem w dalszych procesach oczyszczania powierzchni po wypiaskowaniu. Chociaż zarówno powierzchnia piaskowana biologicznie, jak i mikropiaskowana zostały wypiaskowane substancjami resorbowalnymi, to zastosowanie różnej aparatury oraz innego sposobu oczyszczania powierzchni spowodowały powstanie różnic w szorstkości i składzie chemicznym, co jest typowe przy inkorporacji cząsteczek bioaktywnej ceramiki na bazie Ca i P. Po pierwsze, z powodu mniejszej twardości resorbowalnego środka piaskującego w porównaniu do tlenku glinu, zanotowano niższy stopień szorstkości tej powierzchni niż powierzchni piaskowanej Al2O3 i piaskowanej/wytrawianej. Po drugie, obserwacja obrazów powierzchni piaskowanej biologicznie oraz mikropiaskowanej z mikroskopu elektronowego wykazała bardziej jednolite rozprowadzenie ziarnistości na powierzchni mikropiaskowanej typu RBM, na której również, w przeciwieństwie do powierzchni piaskowanej biologicznie, nie zaobserwowano takich miejsc, gdzie rowki powstałe w procesie produkcji byłyby widoczne pomiędzy ziarnistościami. Po trzecie, zastosowanie różnych procedur po wypiaskowaniu przyczyniło się do obecności cząsteczek Ca-P na powierzchni piaskowanej biologicznie i wysokiej zawartości tych pierwiastków wykazanej za pomocą widm oraz chemicznej impregnacji [10] cząsteczek Ca i P na powierzchni mikropiaskowanej. Chociaż nie podano informacji dotyczących składu źródła plazmy oraz temperatury obróbki, to z obrazów mikroskopii elektronowej wynika, że powierzchnie te zostały najpierw wypiaskowane oraz że proces nanoszenia plazmy wpłynął na stopień szorstkości powierzchni. Poparciem dla tego przypuszczenia jest obraz powierzchni z zaokrąglonymi strukturami, w porównaniu do ostrych zarysów szczytów i dolin na pozostałych powierzchniach. Dodatkowo, poprzez zastosowanie tej metody uzyskano wyraźną zmianę w zakresie składu pierwiastkowego powierzchni tytan (Ti) nie został wykryty, natomiast zawartość węgla (C) oraz tlenu (O) były podwyższone. Taka zmiana składu chemicznego powierzchni może być wynikiem wyraźnego zwiększenia powierzchniowej warstwy tlenków lub wzrostu energii powierzchniowej, co mogło skutkować absorpcją związków węgla z atmosfery podczas procesu obróbki próbek. Wyniki badania zerwanego interfejsu pod wpływem działania siły obrotowej wykazały wysokie wartości zamocowania biomechanicznego dla implantów o wszystkich badanych powierzchniach po okresie 2 tygodni (istotnie statystycznie wyższe dla powierzchni mikropiaskowanej RBM w porównaniu do pozostałych) oraz ogólny wzrost wartości tego parametru dla wszystkich badanych powierzchni po okresie 4 tygodni (brak istotnych różnic pomiędzy grupami). Takie wyniki, jakie uzyskano po okresie wynoszącym 2 tygodnie, wskazują, że stopień szorstkości oraz skład chemiczny powierzchni mikropiaskowanej RBM sprzyjają wczesnej odpowiedzi tkanek gospodarza na implant, oraz że krótko po tym czasie obserwacji nie wykazano już istotnych różnic dzięki dobrym właściwościom osteokondukcyjnym oraz biokompatybilności pozostałych badanych powierzchni. Mały wymiar mechanicznego uszkodzenia ciągłości pomiędzy kością a implantem obserwowany na preparatach histologicznych wykonanych po testach wytrzymałości mechanicznej był prawdopodobnie spowodowany odpowiednim ustawieniem próbek i powolną, kontrolowaną wartością momentu obrotowego. Dlatego też mechaniczne rozerwanie połączenia między implantem a kością było widoczne tylko na kilku przekrojach histologicznych i nie wpłynęło na ocenę histomorfologiczną czy histomorfometryczną [13, 14, 17]. Generalnie rzecz biorąc, wyniki badania histologicznego przekrojów wykazały, że wszystkie badane powierzchnie były biokompatybilne i posiadały właściwości osteokondukcyjne kość znajdowała się w ścisłym kontakcie z powierzchniami implantów w rejonie kości korowej i gąbczastej. Z morfologicznego punktu widzenia nie zaobserwowano różnic pomiędzy badanymi rodzajami powierzchni. Utkanie nowej kości było widoczne wokół implantów o wszystkich typach powierzchni po okresie obserwacji wynoszącym 2 i 4 tygodnie, a początkowa organizacja nowego utkania w blaszki kostne miała miejsce dla wszystkich rodzajów powierzchni. W przypadku powierzchni piaskowanej tlenkiem glinu i mikropiaskowanej RBM, w obu okresach obserwacji, nie wykazano jakiegokolwiek szkodliwego działania, które potencjalnie mogło być spowodowane obecnością cząsteczek środków wykorzystanych do piaskowania. Zależnie od rejonu wszczepu oraz wzajemnej relacji jego wymiarów do wymiarów instrumentarium chirurgicznego, zaobserwowano różne wzory gojenia wzdłuż implantu. W miejscach, gdzie ścisły kontakt kości korowej z powierzchnią implantu występował już bezpośrednio po implantacji, zaobserwowano gojenie przez apozycję nowej kości [16, 27, 29]. Taki sposób gojenia polega na remodelingu kości w okolicy przylegającej do wszczepu z następową apozycją nowego utkania kostnego w ścisłym kontakcie z powierzchnią implantu [16, 27, 29], co zostało zaobserwowane dla wszystkich badanych grup powierzchni po 2 i 4 tygodniach. Taka sekwencja procesów przebudowy kostnej skutkuje typowo wysokim stopniem pierwotnej stabilności implantu [27]. Z drugiej jednak strony, kiedy w wyniku wzajemnej relacji geometrii implantu i wymiarów instrumentarium chirurgicznego doszło do powstania komór gojących, zaobserwowano wzór gojenia przypominający kość pochodzenia błoniastego [16, 27, 29]. Wcześniej wykazano, że komory gojące szybko wypełniają się nowym utkaniem kostnym na całej przestrzeni zajętej przez skrzep krwi, który formuje się bezpośrednio po wprowadzeniu implantu, co zapewnia osteointegrację [16, 17, 27, 29, 30]. Uzyskane przez nas wyniki wykazały, że w miejscach gdzie pomiędzy implantem a kością zbitą powstały komory gojenia, szybkie wypełnianie się tych przestrzeni nowym utkaniem kostnym miało miejsce jeszcze przed upływem 2 tygodni, a początkowa organizacja nowego utkania kostnego w blaszki kostne była obserwowana po upływie 4 tygodni, co pozostaje w zgodzie w wynikami wcześniej publikowanych badań. Taki sam trend morfologicznej ewolucji pierwotnego utkania w blaszki kostne był obserwowany w obszarze kości gąbczastej [31]. Podczas gdy nie wykazano istotnych różnic w wartościach parametrów BIC i BAFO, zaobserwowano ogólny ich wzrost w okresie od 2 do 4 tygodni in vivo, co świadczy o tym, że przyjęte w tym badaniu ramy czasowe pokrywały się z dynamicznym procesem gojenia, które ma miejsce we wczesnym etapie po implantacji. W obrębie uwzględnionych w badaniu grup, największy wzrost wartości BIC i BAFO w czasie zaobserwowano dla powierzchni piaskowanej biologicznie, dla której wartości średnie tych parametrów były po okresie 4 tygodni wyższe od pozostałych spośród badanych powierzchni. Prawdopodobną 6

6 przyczyną tego był wpływ wyższej zawartości Ca i P na tej powierzchni w porównaniu do innych powierzchni, co sugeruje też że obecność tych pierwiastków powoduje zmiany w dynamice procesu gojenia kości po implantacji [17-20, 32]. WNIOSKI Zależność pomiędzy makrogeometrią implantu i techniką chirurgiczną a pięcioma różnymi rodzajami powierzchni o właściwościach osteokondukcyjnych poskutkowało wysokim stopniem osteointegracji i mocowania biomechanicznego. Tworzenie poglądów na podstawie badania oceniającego pięć powierzchni o różnej teksturze i składzie chemicznym jest wprawdzie możliwe, to jednak jednoznaczne wskazanie, który z parametrów odpowiada za różnice wyników biomechanicznych i histomorfometrycznych, jest już niemożliwe. Konieczne są kolejne badania eksperymentalne analizujące te zmienne. Autorzy zapewniają o braku konfliktu interesów. Background: Implant surface chemical modification is typically associated with surface topographical alterations that may affect early osseointegration. This study investigated the effects of controlled surface alterations in early osseointegration in an animal model. Methods: Five implant surfaces were evaluated: Alumina-Blasted, Biological Blasting, Plasma, Microblasted-resorbable blasting media (microblasted RBM), and Alumina-Blasted/ Acid-Etched (AB/AE). Surface topography was characterized by scanning electron microscopy, and optical interferometry, and chemical assessment by x-ray photoelectron spectros copy (XPS). The implants were placed in the radius of 6 dogs, remaining 2 and 4 weeks in vivo. Following euthanization, specimens were torqued-to-interface failure, and nondecalcified processed for histomorphological, bone-to-implant con tact (BIC), and bone area fraction occupied (BAFO) evaluation. Statistical evaluation was performed by one-way ANOVA (p < 0.05), and post-hoc testing by the Tukey s test. Results: The alumina-blasted surface presented the highest average surface roughness (Sa) and mean root square of the surface (Sq) values, the biological blasting the lowest, and AB/AE an intermediate value. The remaining presented inter mediate values between the biological blasting and AB/AE. The XPS spectra revealed Ca and P for the biological blasting and microblasted RBM surface, and the highest 0 levels for the plasma, microblasted RBM, and AB/AE surfaces. Significantly higher torque was observed at 2 weeks for the microblasted RBM surface (p<0.04), but no differences existed be tween surfaces at 4 weeks (p >0.74).No significant differences in BIC and BAFO values were observed at 2 and 4 weeks. Conclusions: The five surfaces were osseoconductive and resulted in high degrees of osseointegration and biomechanical fixation. Over the last four decades, clinical oral implantology have shown high survival rates over time (often exceeding 95 /o over 10 years [1, 2] ), and has been regarded as one of the most successful treatment modalities in dentistry. Such high success rates have been attributed to the excellent biocompatibility of titanium that allow for intimate bone interaction at the optical microscopy resolution, regarded as osseointegration [3-5]. While high survival rates have been reported for endosseous devices, current research has emphasized towards implant design modification at various length scales (i.e.nano, micro, and macrogeometry) in desire to improve the early host im-plant tissue response [6]. Such potential decrease in healing time may result in reduction in treatment time frames through prosthetic restorations that could be placed in occlusal function at early implantation times [6-8]. Among implant design modifications attempting to improve the host-to-implant response, implant surface modifications have been the most investigated [3-8]. The rationale for surface modification lies upon the fact that it is the first part of the implant to interact with biofluids, potentially altering the cascade of events that leads to bone healing and intimate apposition with the device [9]. Several reviews cover the large number of possibilities included in implant surface modifications, and it is a general con sensus that both rough surfaces (over smooth turned surfaces) and surface chemistry (additions of Ca-P based bioceramics in various forms over non-coated surfaces) favor the early host-to-implant response [4-6]. From a historical perspective, dental implant surfaces evolved from the as-turned smooth surfaces towards textured rough surfaces, and recent research point towards chemistry modification of moderately rough surface [4-6,10,11]. From a temporal standpoint, both topographic and chemistry surface modifications have drawn attention [4-6,10,11] as both have been showing promising results in vitro [12] and in vivo models [13-20] relative to their moderately rough predecessors. Improvements have been achieved by alterations in surface wettability [21], impregnation of Ca and P onto the titanium oxide layer [18], deposition of discrete bioactive ceramics [19,20,22], and through minor incorporation of other chemical elements such as fluoride [11,23]. Since changes in surface chemistry typically result in surface texture during processing, controlling such variables in order to determine their relative effects in healing is a challenging task, and the largest direct comparison between various surfaces in a suitable in vivo model is desirable. Thus, the present study biomechanically and histomorphometrically evaluated the effect of various surface modifications in an animal model. MATERIALS AND METHODS The implants used in this study were Ti-6AI-4V screw type implants with 3.75 mm of diameter and 8 mm in length provided by the manufacturer (AB Dental). A total of 75 implants were used and divided in five groups according to surface treatment: Alumina-Blasted, Biological Blasting, Plasma, Microblasted resorbable blasting media (microblasted RBM), and Alumina-Blasted/Acid-Etched (AB/AE). Three implants from each group were used for surface characterization. Surface Characterization The surface characterization was accomplished with three different methods (n=3 implants per surface). First, scanning electron microscopy (SEM) was performed at various magnifications under an acceleration voltage of 15kV to observe the different groups surfaces topography. The second step was to determine the roughness parameters by optical interferometryt IFM). Three implants of each surface were evaluated at the flat region of the implant cutting edges (three measurements per implant) and Sa (arithmetic mean of the absolute values of the surface height within the sampling area) and Sq (root mean square of the surface departures within the sampling area) parameters determined [24, 25]. To separate roughness from waviness and shape for digital 30 measurements, on a micrometer scale, a high-pass Gaussian filter of 250 µm x 250 µm was utilized. Following data normality verification, sta tistical analysis at 95 % level of significance was performed by one-way ANOVA. The third procedure was the surface specific chemical assessment performed by x-ray photoelectron spectroscopy (XPS). The implants were inserted in a vacuum transfer chamber and degassed to 10-7 torr. The samples were then transferred un der vacuum to the XPS spectrometer. Survey spectra were obtained using a 165 mm mean radius concentric hemispherical analyzer operated at constant pass energy of 160 ev for survey and 80 ev for high resolution scans. The take off angle was 90o and a spot size of 150 µm x 150 µm was used. The implant surfaces were evaluated at various locations (3 per implant). Animal Model and Surgical Procedure Following approval of Ethics Committee for Animal Research at Federal University of Santa Catarina, six mongrel dogs were acquired and remained for 2 weeks in the animal facility prior to the first surgical procedure. For surgery, three drugs were administered until general anesthesia achievement by intramuscular injection. The drugs were atropine sulfate (0.044 mg/kg), xilazine chlorate (8mg/kg) and Ketamine chlorate (15mg/ kg). The implantation site was the radius epiphysis. and the right limb of each animal provided implants that remained for 4 weeks in vivo, and the left limb provided 1implants that remained 2 weeks in vivo. For implant placement, the surgical site was shaved with a razor blade and was followed by application of antiseptic iodine solution. An incision of =5cm through the skin and periosteum was performed and the periosteum was elevated for bone exposure. Sequential drills were utilized following the manufacturer s recommendation under abundant saline irrigation at 1.200rpm. The implants were placed in an interpolated distribution to minimize bias from different implantation sites (sites 1 to 5 from proximal to distal) along the radial epiphysis for torque and histomorphometric evaluation. After placement the healing caps were inserted and sutured in layers with vicryl 4-0 for periosteum and nylon 4-0 for skin was performed. The animals stayed in animal care facility and received antibiotic (Benzyl Penicilin Benzatine UI/Kg) and antiinflammatory (Ketoprofen 1% 1ml/5Kg) medication to control the pain and infection. Euthanasia was performed after 4 weeks by anesthesia overdose and the limbs were retrieved by sharp dissection. For the torque testing, the radius was adapted to an electronic torque machine equipped with a 500 Ncm torque load cell. Custom machined tooling was adapted to each implant internal connection and the bone block was carefully positioned to avoid specimen misalignment during testing. The implants were torqued in counter clockwise direction at a rate of ~0,196 radians/min until a 10% decrease in maximum value was recorded, and a torque versus displacement curve was recorded for each specimen. The rationale for this procedure was to minimize interface damage prior to histological procedures [13-14]. The implants in bone were then referred to histomorphometric analysis. The implants in bone were reduced to blocks and were then immersed in 10% buffered formalin solution for 24h. The blocks were then washed in running water for 24h, and gradually dehydrated in a series of alcohol solutions ranging from % ethanol. Following dehydration, the samples were embedded in a methacrylate-based resin according to the manufacturer s instructions. The blocks were then cut into slices (~300 µm thickness) aiming the center of the implant along its long axis with a precision diamond saw glued to acrylic plates with an acrylate-based cement, and a 24h setting time was allowed prior to grinding and polishing. The sections were reduced to a final thickness of ~300 µm by means of a series of SiC abrasive papers (400, 600, 800, 1200 and 2400) in a granding/polishing machine under water irrigation [26]. The sections were then toluidine blue stained and referred to optical microscopy for histomorphologic evaluation. The bone-to-implant contact (BIC) was determined at 50X-200X magnification by means of a computer software. The regions of bone-to-implant contact along the implant perimeter were subtracted from the total implant perimeter, and calculations were performed to determine the BIC. The bone area fraction occupied (BAFO) between threads in trabecular bone regions was determined at 1OOX magnification by means of computer software. The areas occupied by bone were subtracted from the total area between threads, and calculations were performed to determine the BAFO (reported in percentage values of bone area fraction occupied)[27]. Preliminary statistical analyses showed no effect of implant site (i.e., there were no consistent effects of positions 1 to 5 along the radius) on all measurements. Therefore. site was not considered further in the analysis. Statistical evaluation of torque to interface fracture, BIC, and BAFO were performed by one-way ANOVA. Statistical significance was indicated by p-levels less than 5%, and post-hoc testing employed the Fisher LSD test. 7

7 RESULTS All implant surfaces electron micrographs are presented in Fig.s 1 and 2 and their representative 250 µm x 250 µm m IFM three-dimensional reconstructions, in Fig. 3. Their respective Sa and Sq values are presented in Fig. 4 A. The surface texture observed at intermediate and high magnification levels (Fig. 1 and 2), as well as the IFM reconstruction (Fig. 3) revealed morphologic differences between groups. While similar morphology was observed for the alumina-blasted, microblasted RBM, and AB/AE surfaces, scanning electron micrographs showed that the biological blasting presented rough regions from the grit-blasting procedure along with flat regions with the original as-machined grooves. The plasma group surface morphology presented a rough surface with rounded morphology compared to other groups (Fig. 1 E and F). Residual blasting media particles were observed on the alumina-blasting (Fig. 1 A and B) and biological blasting surfaces only (Fig. 1 C and D). The IFM measurements presented significant differences for both Sa and Sq values (Fig. 4 A and B), where the alumina-blasted surface presented the highest, the biological blasting the lowest, and AB/AE intermediate value. The other surfaces presented intermediate values between the biological blasting and AB/AE (non-significant between groups) (Fig. 4 A and 8). The XPS spectra evaluated the presence of AI, P,Ca,N, Ti, C, V,0, and Na for the different surfaces (Fig. 4 A and C). The highest aluminum concentration was observed for the alumina-blasting surface. The highest Ca and P concentra tion was observed for the biological blasting surface, followed by the microblasted RBM at much lower concentrations and all other surfaces without the presence of these chemical elements. No Ti was detected for the plasma treated surface, and the second lowest Ti value was observed for the biological blasting surface. The highest C values were observed for the plasma and alumina blasting surfaces, and the highest 0 levels were observed for the plasma, micro blasted RBM, and AB/AE surfaces (Fig. 4 A and C). The animal surgical procedures and follow-up demonstrated no complications regarding procedural conditions, post operative infection, or other clinical concerns. No implants were excluded from the study due to clinical instability immediately after the euthanization. The biomechanical testing results showed that significantly higher torque to interface fracture occurred for the microblasted RBM surface relative to others at 2 weeks (P<0.04), but that at 4 weeks no differences were observed between surfaces (p >0.74) (Fig. 5 A and B). The nondecalcified sample processing after controlled torque testing showed intimate bone contact with all implant surfaces at regions of cortical and trabecular bone. Higher magnification of the bone-implant interface region showed that the nondecalcified sections obtained following biomechanical testing presented minimal morphologic distortion due to mechanical testing bone disruption (Fig. 6). Qualitative evaluation of the toluidine blue stained thin sections revealed no morphologic differences between surfaces at 2 weeks (Fig. 6 A, B, and C) and 4 weeks (Fig.6 D, E, and F) in vivo, where intimate contact between cortical (Fig. 6 A, B, D, and E) and trabecular (Fig. 6 C and 6 F) bone were observed. In addition, different healing patterns were observed at different regions along the implant bulk, depending on the interplay between implant geometry and surgical instrumentation dimensions. At the region of the implant where the inner thread diameter was larger or equal the final surgical drilling dimension al lowing intimate contact between implant surface and cortical bone occurred immediately after implantation, substantial bone remodeling in proximity with the implant surface occurred between 2 (Fig. 6A) and 4 (Fig. 6D) weeks in vivo for all groups. While at 2 weeks in vivo old bone remodeling was observed along with regions of newly formed woven bone (Fig. 6A), at 4 weeks substantial woven bone was observed in proximity with the implant surface (Fig. 6D). At regions where a healing chamber was formed due to the formation of a space between the outer diameter of the surgical instrumentation and the inner diameter of the implant thread, woven bone formation was observed throughout the space of the chamber and directly onto the implant surface at 2 weeks in vivo (Fig. 6B). At 4 weeks. initial woven bone replacement by lamellar bone was observed throughout the healing chamber (Fig. 6E). At trabecular bone regions, newly formed woven bone was observed at 2 weeks (Fig. 6C), and initial woven bone replacement by lamellar bone was observed at 4 weeks (Fig. 6F) at regions in proximity with all implant surfaces. The histomorphometric results demonstrated no significant differences between surfaces for both BIC and BAFO at 2 and 4 weeks in vivo (BIC p>0.26 and p>0.09, respectively: BAFO p>0.94 and P>0.09, respectively- Fig.s 5 C,D,E, and F). DISCUSSION Implant surfaces have evolved from the smooth as-machined (as-turned) surfaces towards the now considered standard rough surfaces that are fabricated by a variety of methods, which include all those employed to fabricate the different surfaces evaluated in the present study6. Compared to published data, the obtained surface roughness values were higher due to the larger filter size utilized during IFM measurements (the smaller the filter, lower Sa and Sq values obtained28). Nonetheless, the relative differences between groups are in agreement with previously published work [4,5]. While the alumina blasting group presented a textured surface along with blasting media particles embedded on the surface, the AB/AE surface presented a reduced surface roughness, however without evidence of particle embedding in the surface, demonstrating the effectiveness of acid-etching on further cleaning the surface after blasting procedures. Specific to the incorporation of Ca- and P-based bioactive ceramics, although both the biological blasting and micro blasted RBM surfaces were blasted with resorbable blasting media, the blasting machinery and subsequent surface c leaning differences resulted in different textures and chemistries. First, due to the lower hardness of RBM compared to alumina, lower degrees of roughness were observed compared to both alumina blasting and AB/AE samples. Second, observation of the electron micrographs for the biological blasting and microsblasted RBM revealed more consistent spatial distribution of texture for the microblasted RBM surface, which unlike the biological blasting group did not show regions where machining grooves were apparent between textured regions. Third. different post-blasting procedures resulted in Ca-P particles throughout the biological blasting surface and high degrees of Ca and P on its surface chemistry spectrum, and chemical impregnation10 of Ca and P elements on the microblasted RBM surface. Although no detail was provided regarding the plasma source composition and temperature for the plasma group, it is apparent from the electron micrographs that the surface was previously blasted and that the texture was affected by the plasma processing. Such statements can be rationalized since rounded structures were observed compared to sharp defined peaks and valleys observed in all other groups. In addition, a substantial chemical shift was achieved by such method, where Ti was not detected along with increased C and 0 in the surface composition. This chemical shift may have originated from a substantialincrease in the surface oxide layer, or possibly by a high surface energy characteristic that may have readily adsorbed carbon-based species from the atmosphere during sample preparation. The torque to interface fracture results showed high biomechanical fixation values for all implant surfaces at 2 weeks (significantly higher for the microblasted RBM compared to other surfaces). and a general increase for all groups at 4 weeks (no significant differences between groups). Such result at 2 weeks showed that the microblasted RBM surface roughness and chemistry combination favored the early host-to-implant response, and that shortly after this observation period, no difference was evident due to the osseoconductive and biocompatible properties of other surfaces. The low degree of mechanical disruption between bone and implant observed in the histology slides following mechanical test ing was likely due to the proper specimen alignment and the slow controlled torque rate. Thus, mechanical disruption was observed only in a few histologic sections and did not compromise the histomorphologic and histomorphometric evaluations [13,14,17]. In general, results from the histologic sections showed that all of the surfaces investigated were biocompatible and osseconductive, presenting bone in close contact with the implant surface at regions of cortical and trabecular bone. From a morphologic standpoint no differences were observed between all surface investigated. Woven bone was observed around all surfaces at 2 weeks, and at 4 weeks, initial replacement of woven bone by lamellar bone was observed for all surfaces. No detrimental effect due to blasting media particle presence on the surface was observed for the alumina blasting and microblasted RBM groups at both implantation times. Specific to different regions of the implant and its relationship with the surgical drilling dimensions, different healing patterns were observed throughout the implant length. At regions where intimate contact between cortical bone and implant surface existed immediately after placement, an appositional bone healing was observed [16, 27, 29]. Such healing pattern comprises in interfacial remodeling with subsequent woven bone apposition in close contact with the implant surface [16, 27, 29], as observed sequentially for all groups at 2 and 4 weeks. This type of configuration typically results in high degrees of implant primary stability [27]. On the other hand, when the interplay between implant geometry and surgical drilling dimensions resulted in healing chamber formation. an intramembranous-like healing pattern was observed[16,27,29]. Healing chambers have been previously shown to be rapidly filled with woven bone throughout the volume occupied by the blood clot immediately after placement for osseointegration achievement [16,17,27,29,30]. In agreement with previous studies, the present results showed that in regions where healing chambers formed between implant and cortical bone. rapid woven bone filling occurred before or at 2 weeks, and initial remodeling comprising initial woven bone replacement by lamellar bone was observed by 4 weeks implantation time. The same morphologic evolution trend was observed at regions of trabecular bone [31]. While no significant differences were observed for both BIC and BAFO, a general increase was observed from 2 to 4 weeks in vivo, revealing that the time frames investigated in the present study was within the dynamic healing phases that occur at early implantation times. Within groups, the highest increase in BIC and BAFO values over time was ob served for the biological blasting group. which at 4 weeks presented the highest mean values for BIC and BAFO among all surfaces evaluated. Such observation was likely due to the effect of higher amounts of Ca and P elements on the surface relative to other groups, suggesting that their presence resulted in alteration in bone healing dynamics after implantation [17-20,32]. CONCLUSION The association of the implant macrogeometry and associated surgical technique with five different osseoconductive surfaces resulted in high degrees of osseointegration and biomechanical fixation. While insight can be provided by evaluating five surfaces with distinct texture and chemistry, pinpointing which of the surface parameters accounted for the differences in biomechanical and histomorphometric results,is not possible, experimental studies controlling these variables are warranted. The authors declare no conflict of interest. 8

8 Piśmiennictwo 1. Chuang SK, Tian L, Wei LJ, Dodson TB. Predicting dental implant survival by use of the marginal approach of the semi-parametric survival methods for clustered observations. J Dent Res;81: , Chuang SK, Wei LJ, Douglass CW, Dodson TB. Risk factors for dental implant failure: a strategy for the analysis of clustered failure-time observations. J Dent Res;81: , Albrektsson T, Gottlow J, Meirelles L,Ostman PO. Rocci A, Sennerby L. Survival of NobelDirect implants: an analysis of 550 consecutively placed implants at 18 different clinical centers. Clin Implant Dent Relat Res;9:65-70, Albrektsson T, Wennerberg A. Oral implant surfaces:part 1--review focusing on topographic and chemical properties of different surfaces and in vivo responses to them. lnt J Prosthodont;17: , Albrektsson T, Wennerberg A. Oral implant surfaces: Part 2--review focusing on clinical knowledge of different surfaces. lnt J Prosthodont;17: , Coelho PG,Granjeiro JM, Romanos GE, et al. Basic research methods and current trends of dental implant surfaces. J Biomed Mater Res B Appl Biomater;88: , Jimbo A, Ono D. Hirakawa Y, Odatsu T, Tanaka T, Sawase T. Accelerated Photo-Induced Hydrophilicity Promotes Osseointegration: An Animal Study. Clin Implant Dent Relat Res 2009 in press. 8. Jimbo A, Sawase T, Baba K. Kurogi T, Shibata Y, Atsuta M. Enhanced initial cell responses to chemically modified anodized titanium.clin Implant Dent Relat Res;10:55-61, Jimbo A, Sawase T, Shibata Y, et at. Enhanced osseointegration by the chemotactic activity of plasma fibronectin for cellular fibronectin positive cells. Biomaterials;28: n, Dohan Ehrenfest OM, Coelho PG, Kang BS, Sui YT, Albrektsson T. Classification of osseointegrated implant surfaces: materials, chemistry and topography. Trends Biotechnol;28: , Kang BS Sui YT, Oh SJ. Lee HJ, Albrektsson T. XPS, AES and SEM analysis of recent dental implants. Acta Biomater;5: , Moura CC, Souza MA,DechichiP, Zanetta-Barbosa 0, Teixeira CC, Coelho PG. The effect of a nanothickness coating on rough titanium substrate in the osteogenic properties of human bone cells. J Biomed Mater Res A;94: , Coelho PG, Cardaropoli G, Suzuki M, Lemons JE. Early healing of nanothickness bioceramic coatings on dental implants. An experimental study in dogs. J Biomed Mater Res B Appl Biomater;88: , Coelho PG, Lemons JE. Physico/chemical characterization and in vivo evaluation of nanothickness bioceramic depositions on alumina-blasted/acid-etched Ti-6AI-4V implant surfaces. J Biomed Mater Res A;90: , Coelho PG, Marin C, Granato A, Suzuki M. Histomorphologic analysis of 30 plateau root form implants retrieved after 8 to 13 years in function. A human retrieval study. J Biomed Mater Res B Appl Biomater;91: , Coelho PGt Suzuki M, Guimaraes MV, et al. Early bone healing around different implant bulk designs and surgical techniques: A study in dogs.clin Implant Dent Relat Res;12: , Granato A, Marin C, Suzuki M, Gil JN, JanalMN, Coelho PG. Biomechanical and histomorphometric evaluation of a thin ion beam bioceramic deposition on plateau root form implants: an experimental study in dogs. J Biomed Mater Res B Appl Biomater;90: , Marin C, Granato R, Suzuki M, Gil JN, Piattelli A, Coelho PG. RemovalTorque and Histomorphometric Evaluation of Bioceramic Grit-Blasted/Acid-Etched and Dual Acid-Etched Implant Surfaces:An Experimental Study in Dogs. J Periodontol:79: , Mendes VC, Moineddin R, Davies JE. The effect of discrete calcium phosphate nanocrystals on bone-bonding to titanium surfaces. Biomaterials;28: , Mendes VC, Moineddin A, Davies JE. Discrete calcium phosphate nanocrystalline deposition enhances osteoconduction on titanium-based implant surfaces. J Biomed Mater Res A;90: , Buser D, Broggini N, Wieland M, et al. Enhanced bone apposition to a chemically modified SLA titanium surface. J Dent Res; 83: , Orsini G, Piattelli M, Scarano A, et al. Randomized, controlled histologic and histomorphometric evaluation of im plants with nanometer-scale calcium phosphate added to the dual acid-etched surface in the human posterior maxilla. J Periodontal; 78: , Mendonca G. Mendonca DB. Aragao FJ. Cooper LF. Advancing dental implant surface technologyfrom micron- to nanotopography. Biomaterials; 29: , Leach A. Surface Topography Characterisation. In: Fundamental principles of engineering nanometrology: William Andrew Pub,: , Stout K, Sullivan P. Dong W. et al. Development of methods for the characterisation of roughness in three dimensions: Penton Pr; Donath K, Breuner G. A method for the study of undecalcified bones and teeth with attached soft tissues. The Sage Schliff {sawing and grinding) technique. J Oral Pathol;11: , Leonard G, Coelho P, Polyzois I, Stassen L, Claffey N. A study of the bone healing kinetics of plateau versus screw root design titanium dental implants. Clin Oral Implants Res; 20: , Wennerberg A, Albrektsson T. Suggested guidelines for the topographic evaluation of implant surfaces. lnt J Oral Maxillofac Implants;15: , Berglundh T,Abrahamsson I, Lang NP, Lindhe J. De novo alveolar bone formation adjacent to endosseous implants. Clin Oral Implants Res;14: , Coelho PG, Granato A, Marin C, Bonfante EA, Janal MN, Suzuki M. Biomechanical and bone histomorphologic evaluation of four surfaces on plateau root form implants: an experimental study in dogs. Oral Surg Oral Med Oral Pathol Oral Radial Endod; 109:e39-45, Quaranta A, Iezzi G, Scarano A, et al. A histomorphometric study of nanothickness and plasmasprayed calcium phosphorous-coated implant surfaces in rabbit bone. J Periodontal; 81: , Meirelles L, Melin L, Peltola T, et al. Effect of hydroxyapatite and titania nanostructures on early in vivo bone response. Clin Implant Dent Aelat Res; 10: ,

9 10

10 11

Ocena skuteczności preparatów miejscowo znieczulających skórę w redukcji bólu w trakcie pobierania krwi u dzieci badanie z randomizacją

Ocena skuteczności preparatów miejscowo znieczulających skórę w redukcji bólu w trakcie pobierania krwi u dzieci badanie z randomizacją 234 Ocena skuteczności preparatów miejscowo znieczulających skórę w redukcji bólu w trakcie pobierania krwi u dzieci badanie z randomizacją The effectiveness of local anesthetics in the reduction of needle

Bardziej szczegółowo

Akademia Morska w Szczecinie. Wydział Mechaniczny

Akademia Morska w Szczecinie. Wydział Mechaniczny Akademia Morska w Szczecinie Wydział Mechaniczny ROZPRAWA DOKTORSKA mgr inż. Marcin Kołodziejski Analiza metody obsługiwania zarządzanego niezawodnością pędników azymutalnych platformy pływającej Promotor:

Bardziej szczegółowo

IDENTYFIKACJA I ANALIZA PARAMETRÓW GEOMETRYCZNYCH I MECHANICZNYCH KOŚCI MIEDNICZNEJ CZŁOWIEKA

IDENTYFIKACJA I ANALIZA PARAMETRÓW GEOMETRYCZNYCH I MECHANICZNYCH KOŚCI MIEDNICZNEJ CZŁOWIEKA POLITECHNIKA ŚLĄSKA ZESZYTY NAUKOWE Nr 1651 Antoni JOHN SUB Gottingen 7 217 780 458 2005 A 3012 IDENTYFIKACJA I ANALIZA PARAMETRÓW GEOMETRYCZNYCH I MECHANICZNYCH KOŚCI MIEDNICZNEJ CZŁOWIEKA Gliwice 2004

Bardziej szczegółowo

Radiologiczna ocena progresji zmian próchnicowych po zastosowaniu infiltracji. żywicą o niskiej lepkości (Icon). Badania in vivo.

Radiologiczna ocena progresji zmian próchnicowych po zastosowaniu infiltracji. żywicą o niskiej lepkości (Icon). Badania in vivo. Renata Zielińska Radiologiczna ocena progresji zmian próchnicowych po zastosowaniu infiltracji żywicą o niskiej lepkości (Icon). Badania in vivo. Rozprawa na stopień doktora nauk medycznych Zakład Stomatologii

Bardziej szczegółowo

Rozpoznawanie twarzy metodą PCA Michał Bereta 1. Testowanie statystycznej istotności różnic między jakością klasyfikatorów

Rozpoznawanie twarzy metodą PCA Michał Bereta   1. Testowanie statystycznej istotności różnic między jakością klasyfikatorów Rozpoznawanie twarzy metodą PCA Michał Bereta www.michalbereta.pl 1. Testowanie statystycznej istotności różnic między jakością klasyfikatorów Wiemy, że możemy porównywad klasyfikatory np. za pomocą kroswalidacji.

Bardziej szczegółowo

Pro-tumoral immune cell alterations in wild type and Shbdeficient mice in response to 4T1 breast carcinomas

Pro-tumoral immune cell alterations in wild type and Shbdeficient mice in response to 4T1 breast carcinomas www.oncotarget.com Oncotarget, Supplementary Materials Pro-tumoral immune cell alterations in wild type and Shbdeficient mice in response to 4T1 breast carcinomas SUPPLEMENTARY MATERIALS Supplementary

Bardziej szczegółowo

4. EKSPLOATACJA UKŁADU NAPĘD ZWROTNICOWY ROZJAZD. DEFINICJA SIŁ W UKŁADZIE Siła nastawcza Siła trzymania

4. EKSPLOATACJA UKŁADU NAPĘD ZWROTNICOWY ROZJAZD. DEFINICJA SIŁ W UKŁADZIE Siła nastawcza Siła trzymania 3 SPIS TREŚCI Przedmowa... 11 1. WPROWADZENIE... 13 1.1. Budowa rozjazdów kolejowych... 14 1.2. Napędy zwrotnicowe... 15 1.2.1. Napęd zwrotnicowy EEA-4... 18 1.2.2. Napęd zwrotnicowy EEA-5... 20 1.3. Współpraca

Bardziej szczegółowo

Uniwersytet Medyczny w Łodzi. Wydział Lekarski. Jarosław Woźniak. Rozprawa doktorska

Uniwersytet Medyczny w Łodzi. Wydział Lekarski. Jarosław Woźniak. Rozprawa doktorska Uniwersytet Medyczny w Łodzi Wydział Lekarski Jarosław Woźniak Rozprawa doktorska Ocena funkcji stawu skokowego po leczeniu operacyjnym złamań kostek goleni z uszkodzeniem więzozrostu piszczelowo-strzałkowego

Bardziej szczegółowo

Towards Stability Analysis of Data Transport Mechanisms: a Fluid Model and an Application

Towards Stability Analysis of Data Transport Mechanisms: a Fluid Model and an Application Towards Stability Analysis of Data Transport Mechanisms: a Fluid Model and an Application Gayane Vardoyan *, C. V. Hollot, Don Towsley* * College of Information and Computer Sciences, Department of Electrical

Bardziej szczegółowo

Cracow University of Economics Poland. Overview. Sources of Real GDP per Capita Growth: Polish Regional-Macroeconomic Dimensions 2000-2005

Cracow University of Economics Poland. Overview. Sources of Real GDP per Capita Growth: Polish Regional-Macroeconomic Dimensions 2000-2005 Cracow University of Economics Sources of Real GDP per Capita Growth: Polish Regional-Macroeconomic Dimensions 2000-2005 - Key Note Speech - Presented by: Dr. David Clowes The Growth Research Unit CE Europe

Bardziej szczegółowo

ESCA+AES Electron Spectroscopy for Chemical Analysis + Auger Electron Spectroscopy

ESCA+AES Electron Spectroscopy for Chemical Analysis + Auger Electron Spectroscopy ESCA+AES Electron Spectroscopy for Chemical Analysis + Auger Electron Spectroscopy Badanie składu chemicznego powierzchni z wykorzystaniem elektronów ESCA zasada metody Electron Spectroscopy for Chemical

Bardziej szczegółowo

aforementioned device she also has to estimate the time when the patients need the infusion to be replaced and/or disconnected. Meanwhile, however, she must cope with many other tasks. If the department

Bardziej szczegółowo

Has the heat wave frequency or intensity changed in Poland since 1950?

Has the heat wave frequency or intensity changed in Poland since 1950? Has the heat wave frequency or intensity changed in Poland since 1950? Joanna Wibig Department of Meteorology and Climatology, University of Lodz, Poland OUTLINE: Motivation Data Heat wave frequency measures

Bardziej szczegółowo

ROZPRAWY NR 128. Stanis³aw Mroziñski

ROZPRAWY NR 128. Stanis³aw Mroziñski UNIWERSYTET TECHNOLOGICZNO-PRZYRODNICZY IM. JANA I JÊDRZEJA ŒNIADECKICH W BYDGOSZCZY ROZPRAWY NR 128 Stanis³aw Mroziñski STABILIZACJA W ASNOŒCI CYKLICZNYCH METALI I JEJ WP YW NA TRWA OŒÆ ZMÊCZENIOW BYDGOSZCZ

Bardziej szczegółowo

Zarządzanie sieciami telekomunikacyjnymi

Zarządzanie sieciami telekomunikacyjnymi SNMP Protocol The Simple Network Management Protocol (SNMP) is an application layer protocol that facilitates the exchange of management information between network devices. It is part of the Transmission

Bardziej szczegółowo

Formularz recenzji magazynu. Journal of Corporate Responsibility and Leadership Review Form

Formularz recenzji magazynu. Journal of Corporate Responsibility and Leadership Review Form Formularz recenzji magazynu Review Form Identyfikator magazynu/ Journal identification number: Tytuł artykułu/ Paper title: Recenzent/ Reviewer: (imię i nazwisko, stopień naukowy/name and surname, academic

Bardziej szczegółowo

INSPECTION METHODS FOR QUALITY CONTROL OF FIBRE METAL LAMINATES IN AEROSPACE COMPONENTS

INSPECTION METHODS FOR QUALITY CONTROL OF FIBRE METAL LAMINATES IN AEROSPACE COMPONENTS Kompozyty 11: 2 (2011) 130-135 Krzysztof Dragan 1 * Jarosław Bieniaś 2, Michał Sałaciński 1, Piotr Synaszko 1 1 Air Force Institute of Technology, Non Destructive Testing Lab., ul. ks. Bolesława 6, 01-494

Bardziej szczegółowo

Weronika Mysliwiec, klasa 8W, rok szkolny 2018/2019

Weronika Mysliwiec, klasa 8W, rok szkolny 2018/2019 Poniższy zbiór zadań został wykonany w ramach projektu Mazowiecki program stypendialny dla uczniów szczególnie uzdolnionych - najlepsza inwestycja w człowieka w roku szkolnym 2018/2019. Tresci zadań rozwiązanych

Bardziej szczegółowo

Maciej Pawlak. Zastosowanie nowoczesnych implantów i technik operacyjnych w leczeniu przepuklin brzusznych.

Maciej Pawlak. Zastosowanie nowoczesnych implantów i technik operacyjnych w leczeniu przepuklin brzusznych. Maciej Pawlak Zastosowanie nowoczesnych implantów i technik operacyjnych w leczeniu przepuklin brzusznych. Prospektywne randomizowane badanie porównujące zastosowanie dwóch różnych koncepcji siatki i staplera

Bardziej szczegółowo

deep learning for NLP (5 lectures)

deep learning for NLP (5 lectures) TTIC 31210: Advanced Natural Language Processing Kevin Gimpel Spring 2019 Lecture 6: Finish Transformers; Sequence- to- Sequence Modeling and AJenKon 1 Roadmap intro (1 lecture) deep learning for NLP (5

Bardziej szczegółowo

WPŁYW ODKSZTAŁCENIA WZGLĘDNEGO NA WSKAŹNIK ZMNIEJSZENIA CHROPOWATOŚCI I STOPIEŃ UMOCNIENIA WARSTWY POWIERZCHNIOWEJ PO OBRÓBCE NAGNIATANEM

WPŁYW ODKSZTAŁCENIA WZGLĘDNEGO NA WSKAŹNIK ZMNIEJSZENIA CHROPOWATOŚCI I STOPIEŃ UMOCNIENIA WARSTWY POWIERZCHNIOWEJ PO OBRÓBCE NAGNIATANEM Tomasz Dyl Akademia Morska w Gdyni WPŁYW ODKSZTAŁCENIA WZGLĘDNEGO NA WSKAŹNIK ZMNIEJSZENIA CHROPOWATOŚCI I STOPIEŃ UMOCNIENIA WARSTWY POWIERZCHNIOWEJ PO OBRÓBCE NAGNIATANEM W artykule określono wpływ odkształcenia

Bardziej szczegółowo

Metodyki projektowania i modelowania systemów Cyganek & Kasperek & Rajda 2013 Katedra Elektroniki AGH

Metodyki projektowania i modelowania systemów Cyganek & Kasperek & Rajda 2013 Katedra Elektroniki AGH Kierunek Elektronika i Telekomunikacja, Studia II stopnia Specjalność: Systemy wbudowane Metodyki projektowania i modelowania systemów Cyganek & Kasperek & Rajda 2013 Katedra Elektroniki AGH Zagadnienia

Bardziej szczegółowo

PARAMETRY TECHNICZNE DEKLAROWANE PRZEZ PRODUCENTA POTWIERDZONE BADANIAMI / RATINGS ASSIGNED BY THE MANUFACTURER AND PROVED BY TESTS

PARAMETRY TECHNICZNE DEKLAROWANE PRZEZ PRODUCENTA POTWIERDZONE BADANIAMI / RATINGS ASSIGNED BY THE MANUFACTURER AND PROVED BY TESTS Sprawozdanie z Strona/Page 2/24 PARAMETRY TECHNICZNE DEKLAROWANE PRZEZ PRODUCENTA POTWIERDZONE BADANIAMI / RATINGS ASSIGNED BY THE MANUFACTURER AND PROVED BY TESTS Typ Type Napięcie trwałej pracy Continuous

Bardziej szczegółowo

WYKAZ PRÓB / SUMMARY OF TESTS. mgr ing. Janusz Bandel

WYKAZ PRÓB / SUMMARY OF TESTS. mgr ing. Janusz Bandel Sprawozdanie z Badań Nr Strona/Page 2/24 WYKAZ PRÓB / SUMMARY OF TESTS STRONA PAGE Próba uszkodzenia przy przepięciach dorywczych TOV failure test 5 Próby wykonał / The tests were carried out by: mgr ing.

Bardziej szczegółowo

SPITSBERGEN HORNSUND

SPITSBERGEN HORNSUND Polska Stacja Polarna Instytut Geofizyki Polska Akademia Nauk Polish Polar Station Institute of Geophysics Polish Academy of Sciences BIULETYN METEOROLOGICZNY METEOROLOGICAL BULLETIN SPITSBERGEN HORNSUND

Bardziej szczegółowo

Fizyka Procesów Klimatycznych Wykład 11 Aktualne zmiany klimatu: atmosfera, hydrosfera, kriosfera

Fizyka Procesów Klimatycznych Wykład 11 Aktualne zmiany klimatu: atmosfera, hydrosfera, kriosfera Fizyka Procesów Klimatycznych Wykład 11 Aktualne zmiany klimatu: atmosfera, hydrosfera, kriosfera prof. dr hab. Szymon Malinowski Instytut Geofizyki, Wydział Fizyki Uniwersytet Warszawski malina@igf.fuw.edu.pl

Bardziej szczegółowo

Patients price acceptance SELECTED FINDINGS

Patients price acceptance SELECTED FINDINGS Patients price acceptance SELECTED FINDINGS October 2015 Summary With growing economy and Poles benefiting from this growth, perception of prices changes - this is also true for pharmaceuticals It may

Bardziej szczegółowo

Helena Boguta, klasa 8W, rok szkolny 2018/2019

Helena Boguta, klasa 8W, rok szkolny 2018/2019 Poniższy zbiór zadań został wykonany w ramach projektu Mazowiecki program stypendialny dla uczniów szczególnie uzdolnionych - najlepsza inwestycja w człowieka w roku szkolnym 2018/2019. Składają się na

Bardziej szczegółowo

W trzech niezależnych testach frezy z powłoką X tremeblue typu V803 był w każdym przypadku prawie 2 razy bardziej wydajne niż wersja niepowlekana.

W trzech niezależnych testach frezy z powłoką X tremeblue typu V803 był w każdym przypadku prawie 2 razy bardziej wydajne niż wersja niepowlekana. To nowa powłoka ochronna i jest znacznie lepsza jak DLC, - X-TremeBLUE jest nową aplikacją powlekania oparta na najnowszych technologiach NANO struktury. - X-TremeBLUE to powłoka o mikronowej grubości,

Bardziej szczegółowo

WPŁYW CHROPOWATOŚCI POWIERZCHNI MATERIAŁU NA GRUBOŚĆ POWŁOKI PO ALFINOWANIU

WPŁYW CHROPOWATOŚCI POWIERZCHNI MATERIAŁU NA GRUBOŚĆ POWŁOKI PO ALFINOWANIU 51/17 ARCHIWUM ODLEWNICTWA Rok 2005, Rocznik 5, Nr 17 Archives of Foundry Year 2005, Volume 5, Book 17 PAN - Katowice PL ISSN 1642-5308 WPŁYW CHROPOWATOŚCI POWIERZCHNI MATERIAŁU NA GRUBOŚĆ POWŁOKI PO ALFINOWANIU

Bardziej szczegółowo

SPITSBERGEN HORNSUND

SPITSBERGEN HORNSUND Polska Stacja Polarna Instytut Geofizyki Polska Akademia Nauk Polish Polar Station Institute of Geophysics Polish Academy of Sciences BIULETYN METEOROLOGICZNY METEOROLOGICAL BULLETIN SPITSBERGEN HORNSUND

Bardziej szczegółowo

SPITSBERGEN HORNSUND

SPITSBERGEN HORNSUND Polska Stacja Polarna Instytut Geofizyki Polska Akademia Nauk Polish Polar Station Institute of Geophysics Polish Academy of Sciences BIULETYN METEOROLOGICZNY METEOROLOGICAL BULLETIN SPITSBERGEN HORNSUND

Bardziej szczegółowo

Krytyczne czynniki sukcesu w zarządzaniu projektami

Krytyczne czynniki sukcesu w zarządzaniu projektami Seweryn SPAŁEK Krytyczne czynniki sukcesu w zarządzaniu projektami MONOGRAFIA Wydawnictwo Politechniki Śląskiej Gliwice 2004 SPIS TREŚCI WPROWADZENIE 5 1. ZARZĄDZANIE PROJEKTAMI W ORGANIZACJI 13 1.1. Zarządzanie

Bardziej szczegółowo

www.irs.gov/form990. If "Yes," complete Schedule A Schedule B, Schedule of Contributors If "Yes," complete Schedule C, Part I If "Yes," complete Schedule C, Part II If "Yes," complete Schedule C, Part

Bardziej szczegółowo

SPITSBERGEN HORNSUND

SPITSBERGEN HORNSUND Polska Stacja Polarna Instytut Geofizyki Polska Akademia Nauk Polish Polar Station Institute of Geophysics Polish Academy of Sciences BIULETYN METEOROLOGICZNY METEOROLOGICAL BULLETIN SPITSBERGEN HORNSUND

Bardziej szczegółowo

EVALUATION OF SELECTED MECHANICAL AND CHEMICAL METHODS OF MODIFICATIONS OF TITANIUM

EVALUATION OF SELECTED MECHANICAL AND CHEMICAL METHODS OF MODIFICATIONS OF TITANIUM Physicochemical Problems of Mineral Processing, 39 (2005), 149-154 Fizykochemiczne Problemy Mineralurgii, 39 (2005), 149-154 Wiesław HEDZELEK, Barbara SIKORSKA *#, Ludwik DOMKA ** EVALUATION OF SELECTED

Bardziej szczegółowo

Lek. Ewelina Anna Dziedzic. Wpływ niedoboru witaminy D3 na stopień zaawansowania miażdżycy tętnic wieńcowych.

Lek. Ewelina Anna Dziedzic. Wpływ niedoboru witaminy D3 na stopień zaawansowania miażdżycy tętnic wieńcowych. Lek. Ewelina Anna Dziedzic Wpływ niedoboru witaminy D3 na stopień zaawansowania miażdżycy tętnic wieńcowych. Rozprawa na stopień naukowy doktora nauk medycznych Promotor: Prof. dr hab. n. med. Marek Dąbrowski

Bardziej szczegółowo

DM-ML, DM-FL. Auxiliary Equipment and Accessories. Damper Drives. Dimensions. Descritpion

DM-ML, DM-FL. Auxiliary Equipment and Accessories. Damper Drives. Dimensions. Descritpion DM-ML, DM-FL Descritpion DM-ML and DM-FL actuators are designed for driving round dampers and square multi-blade dampers. Example identification Product code: DM-FL-5-2 voltage Dimensions DM-ML-6 DM-ML-8

Bardziej szczegółowo

Rozprawa na stopień naukowy doktora nauk medycznych w zakresie medycyny

Rozprawa na stopień naukowy doktora nauk medycznych w zakresie medycyny Lek. Maciej Jesionowski Efektywność stosowania budezonidu MMX u pacjentów z aktywną postacią łagodnego do umiarkowanego wrzodziejącego zapalenia jelita grubego w populacji polskiej. Rozprawa na stopień

Bardziej szczegółowo

Przykłady zastosowania wszczepów dentystycznych w różnych przypadkach braków zębowych. Część pierwsza etap chirurgiczny

Przykłady zastosowania wszczepów dentystycznych w różnych przypadkach braków zębowych. Część pierwsza etap chirurgiczny Maciej Stupka Przykłady zastosowania wszczepów dentystycznych w różnych przypadkach braków zębowych. Część pierwsza etap chirurgiczny Examples of dental implants in various clinical cases. Part one surgical

Bardziej szczegółowo

STRESZCZENIE Słowa kluczowe: Wstęp Cel pracy

STRESZCZENIE Słowa kluczowe: Wstęp Cel pracy STRESZCZENIE Słowa kluczowe: Noworodek urodzony przedwcześnie, granulocyt obojętnochłonny, molekuły adhezji komórkowej CD11a, CD11b, CD11c, CD18, CD54, CD62L, wczesne zakażenie, posocznica. Wstęp W ostatnich

Bardziej szczegółowo

ZASTOSOWANIE TECHNOLOGII REP-RAP DO WYTWARZANIA FUNKCJONALNYCH STRUKTUR Z PLA

ZASTOSOWANIE TECHNOLOGII REP-RAP DO WYTWARZANIA FUNKCJONALNYCH STRUKTUR Z PLA Aktualne Problemy Biomechaniki, nr 8/2014 109 Emilia MAZGAJCZYK, Patrycja SZYMCZYK, Edward CHLEBUS, Katedra Technologii Laserowych, Automa ZASTOSOWANIE TECHNOLOGII REP-RAP DO WYTWARZANIA FUNKCJONALNYCH

Bardziej szczegółowo

TECHNICAL CATALOGUE WHITEHEART MALLEABLE CAST IRON FITTINGS EE

TECHNICAL CATALOGUE WHITEHEART MALLEABLE CAST IRON FITTINGS EE TECHNICAL CATALOGUE WHITEHEART MALLEABLE CAST IRON FITTINGS EE Poland GENERAL INFORMATION USE Whiteheart malleable cast iron fittings brand EE are used in threaded pipe joints, particularly in water, gas,

Bardziej szczegółowo

Knovel Math: Jakość produktu

Knovel Math: Jakość produktu Knovel Math: Jakość produktu Knovel jest agregatorem materiałów pełnotekstowych dostępnych w formacie PDF i interaktywnym. Narzędzia interaktywne Knovel nie są stworzone wokół specjalnych algorytmów wymagających

Bardziej szczegółowo

Machine Learning for Data Science (CS4786) Lecture 11. Spectral Embedding + Clustering

Machine Learning for Data Science (CS4786) Lecture 11. Spectral Embedding + Clustering Machine Learning for Data Science (CS4786) Lecture 11 Spectral Embedding + Clustering MOTIVATING EXAMPLE What can you say from this network? MOTIVATING EXAMPLE How about now? THOUGHT EXPERIMENT For each

Bardziej szczegółowo

STRESZCZENIE. Wstęp: Cel pracy:

STRESZCZENIE. Wstęp: Cel pracy: STRESZCZENIE Wstęp: Rak piersi jest najczęstszym nowotworem złośliwym występującym u kobiet. Pomimo że tradycyjna mastektomia zastępowana jest coraz częściej przez leczenie oszczędzające, zaburzenia funkcjonalne

Bardziej szczegółowo

POLITECHNIKA WARSZAWSKA. Wydział Zarządzania ROZPRAWA DOKTORSKA. mgr Marcin Chrząścik

POLITECHNIKA WARSZAWSKA. Wydział Zarządzania ROZPRAWA DOKTORSKA. mgr Marcin Chrząścik POLITECHNIKA WARSZAWSKA Wydział Zarządzania ROZPRAWA DOKTORSKA mgr Marcin Chrząścik Model strategii promocji w zarządzaniu wizerunkiem regionu Warmii i Mazur Promotor dr hab. Jarosław S. Kardas, prof.

Bardziej szczegółowo

EDYTA KATARZYNA GŁAŻEWSKA METALOPROTEINAZY ORAZ ICH TKANKOWE INHIBITORY W OSOCZU OSÓB CHORYCH NA ŁUSZCZYCĘ LECZONYCH METODĄ FOTOTERAPII UVB.

EDYTA KATARZYNA GŁAŻEWSKA METALOPROTEINAZY ORAZ ICH TKANKOWE INHIBITORY W OSOCZU OSÓB CHORYCH NA ŁUSZCZYCĘ LECZONYCH METODĄ FOTOTERAPII UVB. Wydział Farmaceutyczny z Oddziałem Medycyny Laboratoryjnej Uniwersytet Medyczny w Białymstoku EDYTA KATARZYNA GŁAŻEWSKA METALOPROTEINAZY ORAZ ICH TKANKOWE INHIBITORY W OSOCZU OSÓB CHORYCH NA ŁUSZCZYCĘ

Bardziej szczegółowo

Dorota Kunkel. WyŜsza Szkoła InŜynierii Dentystycznej

Dorota Kunkel. WyŜsza Szkoła InŜynierii Dentystycznej Dorota Kunkel Implant wszystkie przyrządy medyczne wykonywane z jednego lub więcej biomateriałów, które mogą być umiejscowione wewnątrz organizmu, jak też częściowo lub całkowicie pod powierzchnią nabłonka

Bardziej szczegółowo

EXAMPLES OF CABRI GEOMETRE II APPLICATION IN GEOMETRIC SCIENTIFIC RESEARCH

EXAMPLES OF CABRI GEOMETRE II APPLICATION IN GEOMETRIC SCIENTIFIC RESEARCH Anna BŁACH Centre of Geometry and Engineering Graphics Silesian University of Technology in Gliwice EXAMPLES OF CABRI GEOMETRE II APPLICATION IN GEOMETRIC SCIENTIFIC RESEARCH Introduction Computer techniques

Bardziej szczegółowo

Tytuł pracy w języku angielskim: Microstructural characterization of Ag/X/Ag (X = Sn, In) joints obtained as the effect of diffusion soledering.

Tytuł pracy w języku angielskim: Microstructural characterization of Ag/X/Ag (X = Sn, In) joints obtained as the effect of diffusion soledering. Dr inż. Przemysław Skrzyniarz Kierownik pracy: Prof. dr hab. inż. Paweł Zięba Tytuł pracy w języku polskim: Charakterystyka mikrostruktury spoin Ag/X/Ag (X = Sn, In) uzyskanych w wyniku niskotemperaturowego

Bardziej szczegółowo

Lek. med. Arkadiusz Soski. Streszczenie

Lek. med. Arkadiusz Soski. Streszczenie Lek. med. Arkadiusz Soski Streszczenie Wstęp Rozwojową dysplazję stawu biodrowego (DDH developmental dysplasia of the hip) uznaje się za najczęstszą wadę u noworodków i niemowląt. Wada ta powstać może

Bardziej szczegółowo

www.irs.gov/form990. If "Yes," complete Schedule A Schedule B, Schedule of Contributors If "Yes," complete Schedule C, Part I If "Yes," complete Schedule C, Part II If "Yes," complete Schedule C, Part

Bardziej szczegółowo

Ocena potrzeb pacjentów z zaburzeniami psychicznymi

Ocena potrzeb pacjentów z zaburzeniami psychicznymi Mikołaj Trizna Ocena potrzeb pacjentów z zaburzeniami psychicznymi przebywających na oddziałach psychiatrii sądowej Rozprawa na stopień doktora nauk medycznych Promotor: dr hab.n.med. Tomasz Adamowski,

Bardziej szczegółowo

Latent Dirichlet Allocation Models and their Evaluation IT for Practice 2016

Latent Dirichlet Allocation Models and their Evaluation IT for Practice 2016 Latent Dirichlet Allocation Models and their Evaluation IT for Practice 2016 Paweł Lula Cracow University of Economics, Poland pawel.lula@uek.krakow.pl Latent Dirichlet Allocation (LDA) Documents Latent

Bardziej szczegółowo

NAPRĘŻENIA ŚCISKAJĄCE PRZY 10% ODKSZTAŁCENIU WZGLĘDNYM PRÓBEK NORMOWYCH POBRANYCH Z PŁYT EPS O RÓŻNEJ GRUBOŚCI

NAPRĘŻENIA ŚCISKAJĄCE PRZY 10% ODKSZTAŁCENIU WZGLĘDNYM PRÓBEK NORMOWYCH POBRANYCH Z PŁYT EPS O RÓŻNEJ GRUBOŚCI PRACE INSTYTUTU TECHNIKI BUDOWLANEJ - KWARTALNIK 1 (145) 2008 BUILDING RESEARCH INSTITUTE - QUARTERLY No 1 (145) 2008 Zbigniew Owczarek* NAPRĘŻENIA ŚCISKAJĄCE PRZY 10% ODKSZTAŁCENIU WZGLĘDNYM PRÓBEK NORMOWYCH

Bardziej szczegółowo

BLACKLIGHT SPOT 400W F

BLACKLIGHT SPOT 400W F BLACKLIGHT SPOT 400W F2000339 USER MANUAL / INSTRUKCJA OBSŁUGI BLACKLIGHT SPOT 400W F2000339 Table of Contents 1 Introduction... 2 2 Safety information... 2 3 Product information... 2 3.1 Specification...

Bardziej szczegółowo

Fig 5 Spectrograms of the original signal (top) extracted shaft-related GAD components (middle) and

Fig 5 Spectrograms of the original signal (top) extracted shaft-related GAD components (middle) and Fig 4 Measured vibration signal (top). Blue original signal. Red component related to periodic excitation of resonances and noise. Green component related. Rotational speed profile used for experiment

Bardziej szczegółowo

SPITSBERGEN HORNSUND

SPITSBERGEN HORNSUND Polska Stacja Polarna Instytut Geofizyki Polska Akademia Nauk Polish Polar Station Institute of Geophysics Polish Academy of Sciences BIULETYN METEOROLOGICZNY METEOROLOGICAL BULLETIN SPITSBERGEN HORNSUND

Bardziej szczegółowo

METODYKA OCENY TOPOGRAFII FOLII ŚCIERNYCH ZE SZCZEGÓLNYM UWZGLĘDNIENIEM ROZMIESZCZENIA ZIAREN ŚCIERNYCH

METODYKA OCENY TOPOGRAFII FOLII ŚCIERNYCH ZE SZCZEGÓLNYM UWZGLĘDNIENIEM ROZMIESZCZENIA ZIAREN ŚCIERNYCH XXXIII NAUKOWA SZKOŁA OBRÓBKI ŚCIERNEJ Łódź, -1 września 1 r. METODYKA OCENY TOPOGRAFII FOLII ŚCIERNYCH ZE SZCZEGÓLNYM UWZGLĘDNIENIEM ROZMIESZCZENIA ZIAREN ŚCIERNYCH Wojciech Kacalak *), Katarzyna Tandecka

Bardziej szczegółowo

ROZPRAWA DOKTORSKA. Mateusz Romanowski

ROZPRAWA DOKTORSKA. Mateusz Romanowski Mateusz Romanowski Wpływ krioterapii ogólnoustrojowej na aktywność choroby i sprawność chorych na zesztywniające zapalenie stawów kręgosłupa ROZPRAWA DOKTORSKA Promotor: Dr hab., prof. AWF Anna Straburzyńska-Lupa

Bardziej szczegółowo

BŁĘDY OKREŚLANIA MASY KOŃCOWEJ W ZAKŁADACH SUSZARNICZYCH WYKORZYSTUJĄC METODY LABORATORYJNE

BŁĘDY OKREŚLANIA MASY KOŃCOWEJ W ZAKŁADACH SUSZARNICZYCH WYKORZYSTUJĄC METODY LABORATORYJNE Inżynieria Rolnicza 5(103)/2008 BŁĘDY OKREŚLANIA MASY KOŃCOWEJ W ZAKŁADACH SUSZARNICZYCH WYKORZYSTUJĄC METODY LABORATORYJNE Zbigniew Zdrojewski, Stanisław Peroń, Mariusz Surma Instytut Inżynierii Rolniczej,

Bardziej szczegółowo

Warsztaty Ocena wiarygodności badania z randomizacją

Warsztaty Ocena wiarygodności badania z randomizacją Warsztaty Ocena wiarygodności badania z randomizacją Ocena wiarygodności badania z randomizacją Każda grupa Wspólnie omawia odpowiedź na zadane pytanie Wybiera przedstawiciela, który w imieniu grupy przedstawia

Bardziej szczegółowo

SPITSBERGEN HORNSUND

SPITSBERGEN HORNSUND Polska Stacja Polarna Instytut Geofizyki Polska Akademia Nauk Polish Polar Station Institute of Geophysics Polish Academy of Sciences BIULETYN METEOROLOGICZNY METEOROLOGICAL BULLETIN SPITSBERGEN HORNSUND

Bardziej szczegółowo

Proposal of thesis topic for mgr in. (MSE) programme in Telecommunications and Computer Science

Proposal of thesis topic for mgr in. (MSE) programme in Telecommunications and Computer Science Proposal of thesis topic for mgr in (MSE) programme 1 Topic: Monte Carlo Method used for a prognosis of a selected technological process 2 Supervisor: Dr in Małgorzata Langer 3 Auxiliary supervisor: 4

Bardziej szczegółowo

SPITSBERGEN HORNSUND

SPITSBERGEN HORNSUND Polska Stacja Polarna Instytut Geofizyki Polska Akademia Nauk Polish Polar Station Institute of Geophysics Polish Academy of Sciences BIULETYN METEOROLOGICZNY METEOROLOGICAL BULLETIN SPITSBERGEN HORNSUND

Bardziej szczegółowo

Conception of reuse of the waste from onshore and offshore in the aspect of

Conception of reuse of the waste from onshore and offshore in the aspect of Conception of reuse of the waste from onshore and offshore in the aspect of environmental protection" Koncepcja zagospodarowania odpadów wiertniczych powstających podczas wierceń lądowych i morskich w

Bardziej szczegółowo

Cystatin C as potential marker of Acute Kidney Injury in patients after Abdominal Aortic Aneurysms Surgery preliminary study

Cystatin C as potential marker of Acute Kidney Injury in patients after Abdominal Aortic Aneurysms Surgery preliminary study Cystatin C as potential marker of Acute Kidney Injury in patients after Abdominal Aortic Aneurysms Surgery preliminary study Anna Bekier-Żelawska 1, Michał Kokot 1, Grzegorz Biolik 2, Damian Ziaja 2, Krzysztof

Bardziej szczegółowo

Is there a relationship between age and side dominance of tubal ectopic pregnancies? A preliminary report

Is there a relationship between age and side dominance of tubal ectopic pregnancies? A preliminary report Is there a relationship between age and side dominance of tubal ectopic pregnancies? A preliminary report Czy istnieje zależność pomiędzy wiekiem i stroną, po której umiejscawia się ciąża ektopowa jajowodowa?

Bardziej szczegółowo

Charakterystyka kliniczna chorych na raka jelita grubego

Charakterystyka kliniczna chorych na raka jelita grubego Lek. Łukasz Głogowski Charakterystyka kliniczna chorych na raka jelita grubego Rozprawa na stopień doktora nauk medycznych Opiekun naukowy: Dr hab. n. med. Ewa Nowakowska-Zajdel Zakład Profilaktyki Chorób

Bardziej szczegółowo

PROCEEDINGS OF THE INSTITUTE OF VEHICLES 2(106)/2016 (12 pt)

PROCEEDINGS OF THE INSTITUTE OF VEHICLES 2(106)/2016 (12 pt) PROCEEDINGS OF THE INSTITUTE OF VEHICLES 2(106)/2016 Maciej Foremny 1, Szymon Gudowski 2, Michał Malesza 3, Henryk Bąkowski 4 TRIBOLOGICAL WEAR ESTIMATION OF THE ENGINE OILS USED IN DRIFTING 1. Introduction

Bardziej szczegółowo

WENTYLATORY PROMIENIOWE SINGLE-INLET DRUM BĘBNOWE JEDNOSTRUMIENIOWE CENTRIFUGAL FAN

WENTYLATORY PROMIENIOWE SINGLE-INLET DRUM BĘBNOWE JEDNOSTRUMIENIOWE CENTRIFUGAL FAN WENTYLATORY PROMIENIOWE SINGLE-INLET DRUM BĘBNOWE JEDNOSTRUMIENIOWE CENTRIFUGAL FAN TYP WPB TYPE WPB Wentylatory promieniowe jednostrumieniowe bębnowe (z wirnikiem typu Single-inlet centrifugal fans (with

Bardziej szczegółowo

Updated Action Plan received from the competent authority on 4 May 2017

Updated Action Plan received from the competent authority on 4 May 2017 1 To ensure that the internal audits are subject to Response from the GVI: independent scrutiny as required by Article 4(6) of Regulation (EC) No 882/2004. We plan to have independent scrutiny of the Recommendation

Bardziej szczegółowo

Walyeldeen Godah Małgorzata Szelachowska Jan Kryński. Instytut Geodezji i Kartografii (IGiK), Warszawa Centrum Geodezji i Geodynamiki

Walyeldeen Godah Małgorzata Szelachowska Jan Kryński. Instytut Geodezji i Kartografii (IGiK), Warszawa Centrum Geodezji i Geodynamiki Porównanie globalnych modeli geopotencjału opracowanych na podstawie danych z misji GRACE oraz GOCE/GRACE z przyspieszeniem siły ciężkości pomierzonym w Obserwatorium Geodezyjno-Geofizycznym Borowa Góra

Bardziej szczegółowo

LECZENIE ZŁAMAŃ KOŃCA DALSZEGO KOŚCI UDOWEJ ZESPOLENIEM PŁYTĄ BLOKOWANĄ OCENA KLINICZNA I BADANIE BIOMECHANICZNE

LECZENIE ZŁAMAŃ KOŃCA DALSZEGO KOŚCI UDOWEJ ZESPOLENIEM PŁYTĄ BLOKOWANĄ OCENA KLINICZNA I BADANIE BIOMECHANICZNE Grzegorz Pakuła LECZENIE ZŁAMAŃ KOŃCA DALSZEGO KOŚCI UDOWEJ ZESPOLENIEM PŁYTĄ BLOKOWANĄ OCENA KLINICZNA I BADANIE BIOMECHANICZNE Streszczenie Wstęp Płyty blokowane zastosowane do leczenia złamań końca

Bardziej szczegółowo

Machine Learning for Data Science (CS4786) Lecture11. Random Projections & Canonical Correlation Analysis

Machine Learning for Data Science (CS4786) Lecture11. Random Projections & Canonical Correlation Analysis Machine Learning for Data Science (CS4786) Lecture11 5 Random Projections & Canonical Correlation Analysis The Tall, THE FAT AND THE UGLY n X d The Tall, THE FAT AND THE UGLY d X > n X d n = n d d The

Bardziej szczegółowo

Podkłady pod stopy podpór i pojemniki na podkłady Stabilizer blocks and box for blocks storage

Podkłady pod stopy podpór i pojemniki na podkłady Stabilizer blocks and box for blocks storage Podkłady pod stopy podpór i pojemniki na podkłady Stabilizer blocks and box for blocks storage PA PODKŁADY POD STOPY PODPÓR STABILIZER PADS 102 A B H D Dop. obciążenie Load capacity Żurawie - Cranes od/from

Bardziej szczegółowo

Unit of Social Gerontology, Institute of Labour and Social Studies ageing and its consequences for society

Unit of Social Gerontology, Institute of Labour and Social Studies ageing and its consequences for society Prof. Piotr Bledowski, Ph.D. Institute of Social Economy, Warsaw School of Economics local policy, social security, labour market Unit of Social Gerontology, Institute of Labour and Social Studies ageing

Bardziej szczegółowo

ZGŁOSZENIE WSPÓLNEGO POLSKO -. PROJEKTU NA LATA: APPLICATION FOR A JOINT POLISH -... PROJECT FOR THE YEARS:.

ZGŁOSZENIE WSPÓLNEGO POLSKO -. PROJEKTU NA LATA: APPLICATION FOR A JOINT POLISH -... PROJECT FOR THE YEARS:. ZGŁOSZENIE WSPÓLNEGO POLSKO -. PROJEKTU NA LATA: APPLICATION FOR A JOINT POLISH -... PROJECT FOR THE YEARS:. W RAMACH POROZUMIENIA O WSPÓŁPRACY NAUKOWEJ MIĘDZY POLSKĄ AKADEMIĄ NAUK I... UNDER THE AGREEMENT

Bardziej szczegółowo

Cracow University of Economics Poland

Cracow University of Economics Poland Cracow University of Economics Poland Sources of Real GDP per Capita Growth: Polish Regional-Macroeconomic Dimensions 2000-2005 - Keynote Speech - Presented by: Dr. David Clowes The Growth Research Unit,

Bardziej szczegółowo

Tytuł pracy w języku angielskim: Physical properties of liquid crystal mixtures of chiral and achiral compounds for use in LCDs

Tytuł pracy w języku angielskim: Physical properties of liquid crystal mixtures of chiral and achiral compounds for use in LCDs Dr inż. Jan Czerwiec Kierownik pracy: dr hab. Monika Marzec Tytuł pracy w języku polskim: Właściwości fizyczne mieszanin ciekłokrystalicznych związków chiralnych i achiralnych w odniesieniu do zastosowań

Bardziej szczegółowo

Country fact sheet. Noise in Europe overview of policy-related data. Poland

Country fact sheet. Noise in Europe overview of policy-related data. Poland Country fact sheet Noise in Europe 2015 overview of policy-related data Poland April 2016 The Environmental Noise Directive (END) requires EU Member States to assess exposure to noise from key transport

Bardziej szczegółowo

TOPOGRAFIA WSPÓŁPRACUJĄCYCH POWIERZCHNI ŁOŻYSK TOCZNYCH POMIERZONA NA MIKROSKOPIE SIŁ ATOMOWYCH

TOPOGRAFIA WSPÓŁPRACUJĄCYCH POWIERZCHNI ŁOŻYSK TOCZNYCH POMIERZONA NA MIKROSKOPIE SIŁ ATOMOWYCH 5-2011 T R I B O L O G I A 31 Adam CZABAN *, Andrzej MISZCZAK * TOPOGRAFIA WSPÓŁPRACUJĄCYCH POWIERZCHNI ŁOŻYSK TOCZNYCH POMIERZONA NA MIKROSKOPIE SIŁ ATOMOWYCH TOPOGRAPHY OF ROLLING BEARINGS COOPERATING

Bardziej szczegółowo

Few-fermion thermometry

Few-fermion thermometry Few-fermion thermometry Phys. Rev. A 97, 063619 (2018) Tomasz Sowiński Institute of Physics of the Polish Academy of Sciences Co-authors: Marcin Płodzień Rafał Demkowicz-Dobrzański FEW-BODY PROBLEMS FewBody.ifpan.edu.pl

Bardziej szczegółowo

INSTYTUT GENETYKI I HODOWLI ZWIERZĄT POLSKIEJ AKADEMII NAUK W JASTRZĘBCU. mgr inż. Ewa Metera-Zarzycka

INSTYTUT GENETYKI I HODOWLI ZWIERZĄT POLSKIEJ AKADEMII NAUK W JASTRZĘBCU. mgr inż. Ewa Metera-Zarzycka INSTYTUT GENETYKI I HODOWLI ZWIERZĄT POLSKIEJ AKADEMII NAUK W JASTRZĘBCU mgr inż. Ewa Metera-Zarzycka Profil metaboliczny osocza krwi i wartość biologiczna mleka krów w gospodarstwach ekologicznych Praca

Bardziej szczegółowo

The impact of the global gravity field models on the orbit determination of LAGEOS satellites

The impact of the global gravity field models on the orbit determination of LAGEOS satellites models on the Satelitarne metody wyznaczania pozycji we współczesnej geodezji i nawigacji, Poland 2-4.06.2011 Krzysztof Sośnica, Daniela Thaller, Adrian Jäggi, Rolf Dach and Gerhard Beutler Astronomical

Bardziej szczegółowo

Próbnik kolorów Wood veneer

Próbnik kolorów Wood veneer Fornir jest produktem naturalnym, jego wzór, odcień oraz rysunek mogą się nieznacznie różnić, tak jak osobniczo różnią się pomiędzy sobą poszczególne drzewa tego samego gatunku. Fornir podlega naturalnemu

Bardziej szczegółowo

WPŁYW WARUNKÓW UTWARDZANIA I GRUBOŚCI UTWARDZONEJ WARSTEWKI NA WYTRZYMAŁOŚĆ NA ROZCIĄGANIE ŻYWICY SYNTETYCZNEJ

WPŁYW WARUNKÓW UTWARDZANIA I GRUBOŚCI UTWARDZONEJ WARSTEWKI NA WYTRZYMAŁOŚĆ NA ROZCIĄGANIE ŻYWICY SYNTETYCZNEJ 61/2 Archives of Foundry, Year 21, Volume 1, 1 (2/2) Archiwum Odlewnictwa, Rok 21, Rocznik 1, Nr 1 (2/2) PAN Katowice PL ISSN 1642-58 WPŁYW WARUNKÓW UTWARDZANIA I GRUBOŚCI UTWARDZONEJ WARSTEWKI NA WYTRZYMAŁOŚĆ

Bardziej szczegółowo

O NIEKTÓRYCH SKUTKACH ODDZIAŁYWANIA PROMIENIOWANIA LASERA RUBINOWEGO Z UKŁADEM CIENKA WARSTWA WĘGLIKÓW METALI NA KAPILARNO-POROWATYM PODŁOŻU

O NIEKTÓRYCH SKUTKACH ODDZIAŁYWANIA PROMIENIOWANIA LASERA RUBINOWEGO Z UKŁADEM CIENKA WARSTWA WĘGLIKÓW METALI NA KAPILARNO-POROWATYM PODŁOŻU FIZYA BUDWLI W TERII I PRATYCE TM IV, 29 Sekcja Fizyki Budowli ILiW PAN NIETÓRYCH SUTACH DDZIAŁYWANIA PRMIENIWANIA LASERA RUBINWEG Z UŁADEM CIENA WARSTWA WĘGLIÓW METALI NA APILARN-PRWATYM PDŁŻU Piotr LEMM

Bardziej szczegółowo

WPŁYW SZYBKOŚCI STYGNIĘCIA NA WŁASNOŚCI TERMOFIZYCZNE STALIWA W STANIE STAŁYM

WPŁYW SZYBKOŚCI STYGNIĘCIA NA WŁASNOŚCI TERMOFIZYCZNE STALIWA W STANIE STAŁYM 2/1 Archives of Foundry, Year 200, Volume, 1 Archiwum Odlewnictwa, Rok 200, Rocznik, Nr 1 PAN Katowice PL ISSN 1642-308 WPŁYW SZYBKOŚCI STYGNIĘCIA NA WŁASNOŚCI TERMOFIZYCZNE STALIWA W STANIE STAŁYM D.

Bardziej szczegółowo

DUAL SIMILARITY OF VOLTAGE TO CURRENT AND CURRENT TO VOLTAGE TRANSFER FUNCTION OF HYBRID ACTIVE TWO- PORTS WITH CONVERSION

DUAL SIMILARITY OF VOLTAGE TO CURRENT AND CURRENT TO VOLTAGE TRANSFER FUNCTION OF HYBRID ACTIVE TWO- PORTS WITH CONVERSION ELEKTRYKA 0 Zeszyt (9) Rok LX Andrzej KUKIEŁKA Politechnika Śląska w Gliwicach DUAL SIMILARITY OF VOLTAGE TO CURRENT AND CURRENT TO VOLTAGE TRANSFER FUNCTION OF HYBRID ACTIVE TWO- PORTS WITH CONVERSION

Bardziej szczegółowo

METODY BADAŃ BIOMATERIAŁÓW

METODY BADAŃ BIOMATERIAŁÓW METODY BADAŃ BIOMATERIAŁÓW 1 Cel badań: ograniczenie ryzyka związanego ze stosowaniem biomateriałów w medycynie Rodzaje badań: 1. Badania biofunkcyjności implantów, 2. Badania degradacji implantów w środowisku

Bardziej szczegółowo

Wpływ zanieczyszczenia torowiska na drogę hamowania tramwaju

Wpływ zanieczyszczenia torowiska na drogę hamowania tramwaju DYCHTO Rafał 1 PIETRUSZEWSKI Robert 2 Wpływ zanieczyszczenia torowiska na drogę hamowania tramwaju WSTĘP W Katedrze Pojazdów i Podstaw Budowy Maszyn Politechniki Łódzkiej prowadzone są badania, których

Bardziej szczegółowo

Obrazowanie kręgosłupa w badaniu TK i MR w różnych grupach wiekowych

Obrazowanie kręgosłupa w badaniu TK i MR w różnych grupach wiekowych Lekarz Daria Dziechcińska-Połetek Obrazowanie kręgosłupa w badaniu TK i MR w różnych grupach wiekowych Rozprawa na stopień doktora nauk medycznych Promotor: Dr hab. n. med. Ewa Kluczewska, prof. nadzw.

Bardziej szczegółowo

Installation of EuroCert software for qualified electronic signature

Installation of EuroCert software for qualified electronic signature Installation of EuroCert software for qualified electronic signature for Microsoft Windows systems Warsaw 28.08.2019 Content 1. Downloading and running the software for the e-signature... 3 a) Installer

Bardziej szczegółowo

Raport bieżący: 44/2018 Data: g. 21:03 Skrócona nazwa emitenta: SERINUS ENERGY plc

Raport bieżący: 44/2018 Data: g. 21:03 Skrócona nazwa emitenta: SERINUS ENERGY plc Raport bieżący: 44/2018 Data: 2018-05-23 g. 21:03 Skrócona nazwa emitenta: SERINUS ENERGY plc Temat: Zawiadomienie o zmianie udziału w ogólnej liczbie głosów w Serinus Energy plc Podstawa prawna: Inne

Bardziej szczegółowo

Veles started in Our main goal is quality. Thanks to the methods and experience, we are making jobs as fast as it is possible.

Veles started in Our main goal is quality. Thanks to the methods and experience, we are making jobs as fast as it is possible. Veles started in 2014. Our main goal is quality. Thanks to the methods and experience, we are making jobs as fast as it is possible. Our services: 1. CO2 Laser cutting... 2 2. Laser engraving... 4 3. Thermoplastic

Bardziej szczegółowo

SPITSBERGEN HORNSUND

SPITSBERGEN HORNSUND Polska Stacja Polarna Instytut Geofizyki Polska Akademia Nauk Polish Polar Station Institute of Geophysics Polish Academy of Sciences BIULETYN METEOROLOGICZNY METEOROLOGICAL BULLETIN SPITSBERGEN HORNSUND

Bardziej szczegółowo

OCENA MOśLIWOŚCI WYKORZYSTANIA HODOWLI ŚWIŃ RASY ZŁOTNICKIEJ

OCENA MOśLIWOŚCI WYKORZYSTANIA HODOWLI ŚWIŃ RASY ZŁOTNICKIEJ ASSESSMENT OF POTENTIAL FOR ZŁOTNICKA SPOTTED PIG BREEDING IN ORGANIC FARMS OCENA MOśLIWOŚCI WYKORZYSTANIA HODOWLI ŚWIŃ RASY ZŁOTNICKIEJ PSTREJ W GOSPODARSTWACH EKOLOGICZNYCH Janusz Tomasz Buczyński (1),

Bardziej szczegółowo

Classic. Polski English

Classic. Polski English Classic Polski English Modernistyczna bryła mebli Classic wyraźnie nawiązuje do projektów mistrzów szkoły Bauhausu. Ponadczasowe wzornictwo połączone z funkcjonalną prostotą, oddaje reprezentacyjny charakter

Bardziej szczegółowo